Calibration test bench for three-axis accelerometers

被引:0
|
作者
D'Emilia, Giulio [1 ]
Gaspari, Antonella [1 ]
Natale, Emanuela [1 ]
机构
[1] Univ Aquila, Dept Ind & Informat Engn & Econ, Laquila, Italy
关键词
uncertainty; three-axis accelerometer; calibration; sensitivity matrix; transverse sensitivity; MICROACCELEROMETER; IDENTIFICATION; DESIGN;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper describes and analyzes critically an experimental methodology for the calibration of three-axis accelerometers with reference to sinusoidal vibrations in the range up to 5 Hz. The approach is based on a rotary disk, moved by a high performance, high accuracy servomotor, for the realization of the reference accelerations along different axes and on the sensitivity matrix for the evaluation of main and transversal sensitivities. Therefore, the method is expected to supply sensitivity values and bias values for the all three axes. Particular attention has been paid to the evaluation of the effect of the procedural aspects and of the main interfering quantities, in order to set low the whole calibration uncertainty, even though simple and economical solution are considered for the test bench. The results of a sample calibration of a three- axis MEMS accelerometer are discussed for evaluation purposed of the proposed methodology.
引用
收藏
页码:1574 / 1579
页数:6
相关论文
共 50 条
  • [21] Design Consideration for Three-Axis MEMS Accelerometers using an Asymmetric Proof Mass
    Darmanin, Jean Marie
    Grech, Ivan
    Micallef, Joseph
    Gatt, Edward
    Casha, Owen
    Briffa, Ansel
    2013 IEEE EUROCON, 2013, : 2100 - 2105
  • [22] Calibration Method of Three-Axis Magnetometer Based on Sine Fitting
    Jia W.
    Lin C.
    Lin P.
    Zhai G.
    Wuhan Daxue Xuebao (Xinxi Kexue Ban)/Geomatics and Information Science of Wuhan University, 2019, 44 (09): : 1328 - 1333and1406
  • [23] Calibration for installation error of MIMU based on three-axis turntable
    Han, Xu
    Zhu, Xinhua
    Wang, Yu
    Wu, Zhiqiang
    Yi Qi Yi Biao Xue Bao/Chinese Journal of Scientific Instrument, 2013, 34 (6 SUPPL.): : 211 - 214
  • [24] Complete Calibration of Three-Axis Strapdown Magnetometer in Mounting Frame
    Ousaloo, H. S.
    Sharifi, G.
    Mahdian, J.
    Nodeh, M. T.
    IEEE SENSORS JOURNAL, 2017, 17 (23) : 7886 - 7893
  • [25] Three-Axis Magnetometer Calibration Using Total Least Squares
    Crassidis, John L.
    Cheng, Yang
    JOURNAL OF GUIDANCE CONTROL AND DYNAMICS, 2021, 44 (08) : 1410 - 1424
  • [26] Integrated calibration and magnetic disturbance compensation of three-axis magnetometers
    Pang, Hongfeng
    Pan, Mengchun
    Chen, Jinfei
    Li, Ji
    Zhang, Qi
    Luo, Shitu
    MEASUREMENT, 2016, 93 : 409 - 413
  • [27] Calibration of QM-MOURA three-axis magnetometer and gradiometer
    Diaz-Michelena, M.
    Sanz, R.
    Cerdan, M. F.
    Fernandez, A. B.
    GEOSCIENTIFIC INSTRUMENTATION METHODS AND DATA SYSTEMS, 2015, 4 (01) : 1 - 18
  • [28] Study of calibration of three-axis fluxgate magnetometers on unmanned helicopter
    Wu Peilin
    Zhang Qunying
    Chen Luzhao
    Fang Guangyou
    Liu Wei
    PROCEEDINGS OF 2017 13TH IEEE INTERNATIONAL CONFERENCE ON ELECTRONIC MEASUREMENT & INSTRUMENTS (ICEMI), VOL 1, 2017, : 453 - 457
  • [29] Computationally Efficient Calibration Algorithm for Three-Axis Accelerometer and Magnetometer
    Papafotis, Konstantinos
    Sotiriadis, Paul P.
    2019 8TH INTERNATIONAL CONFERENCE ON MODERN CIRCUITS AND SYSTEMS TECHNOLOGIES (MOCAST), 2019,
  • [30] A Novel Linear Calibration Model for Three-Axis Fluxgate Magnetometer
    Liu, Jianguo
    Li, Xiangang
    Yan, Shenggang
    Yan, Youyu
    Jia, Wei
    Zhang, Qingguo
    IEEE SENSORS JOURNAL, 2021, 21 (21) : 23917 - 23925