Interlaboratory Validation for a Real-Time PCR Salmonella Detection Method Using the ABI 7500 FAST Real-Time PCR System

被引:17
|
作者
Cheng, Chorng-Ming [1 ]
Doran, Tara [2 ]
Lin, Wen [3 ]
Chen, Kai-Shun [1 ]
Williams-Hill, Donna [1 ]
Pamboukian, Ruiqing [3 ]
机构
[1] US FDA, Off Regulatory Affairs, Pacific Reg Lab Southwest, Irvine, CA 92612 USA
[2] US FDA, Ctr Food Safety & Appl Nutr, College Pk, MD 20740 USA
[3] US FDA, Off Regulatory Affairs, Off Regulatory Sci, Rockville, MD 20857 USA
关键词
POLYMERASE-CHAIN-REACTION; GENE; AMPLIFICATION; TYPHIMURIUM; FOODS;
D O I
10.4315/0362-028X.JFP-14-244
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Sixteen FERN (Food Emergency Response Network) member laboratories collaborated in this study to verify extension of the real-time PCR Salmonella detection method originally designed for the single-tube Cepheid SmartCycler II and validated against the Salmonella method of the U.S. Food and Drug Administration Bacteriological Analytical Manual to the Applied Biosystems (ABI) 7500 FAST Real-Time PCR system multiwell plate platform. Four foods were selected for this study: chili powder, soft cheese, fish, and tomatoes; these foods represent products that are commonly analyzed for the presence of Salmonella for regulatory purposes. Each food consisted of six uninoculated control samples, six samples inoculated with low Salmonella levels (target 1 to 5 CFU/25 g), and six samples inoculated with high levels (target 10 to 50 CFU/25 g). All samples were tested for Salmonella using the 24-h quantitative PCR (qPCR) method for detecting Salmonella, which utilizes modified buffered peptone water as the sole enrichment medium and an internal control for the qPCR. Each of these 18 samples was individually analyzed for Salmonella by the collaborating laboratories using both the ABI 7500 FAST system (alternative method) and the SmartCycler LE system (reference method). Statistical analysis of the data revealed no significant difference (P >= 0.05) between these two qPCR platforms except for the chili powder samples. The differences noted with chili powder (P = 0.0455) were attributed to the enhanced sensitivity of the ABI 7500 FAST system compared with the SmartCycler II system. The detection limit of both qPCR methods was 0.02 to 0.15 CFU/g. These results provide a solid basis for extending the 24-h qPCR Salmonella method to the ABI 7500 FAST system for high-throughput detection of Salmonella in foods:
引用
收藏
页码:1119 / 1124
页数:6
相关论文
共 50 条
  • [21] Real-Time PCR Method for Detection of Salmonella spp. in Environmental Samples
    Kasturi, Kuppuswamy N.
    Drgon, Tomas
    APPLIED AND ENVIRONMENTAL MICROBIOLOGY, 2017, 83 (14)
  • [22] Real-time PCR detection chemistry
    Navarro, E.
    Serrano-Heras, G.
    Castano, M. J.
    Solera, J.
    CLINICA CHIMICA ACTA, 2015, 439 : 231 - 250
  • [23] Real-Time PCR
    Ulrich Busch
    Journal für Verbraucherschutz und Lebensmittelsicherheit, 2007, 2 : 111 - 112
  • [24] Real-Time PCR
    Jia, Yibing
    LABORATORY METHODS IN CELL BIOLOGY: BIOCHEMISTRY AND CELL CULTURE, 2012, 112 : 55 - 68
  • [25] Real-time PCR
    不详
    BIOTECHNIQUES, 2008, 44 (02) : 179 - +
  • [26] Real-time PCR
    Salmon, M
    BIOFUTUR, 2002, (219) : A1 - A8
  • [27] Real-time PCR
    Bowyer V.L.
    Forensic Science, Medicine, and Pathology, 2007, 3 (1) : 61 - 63
  • [28] A new real-time PCR method for the identification of Salmonella Dublin
    Persson, S.
    Jacobsen, T.
    Olsen, J. E.
    Olsen, K. E. P.
    Hansen, F.
    JOURNAL OF APPLIED MICROBIOLOGY, 2012, 113 (03) : 615 - 621
  • [29] Comparison of primers for the detection of Salmonella enterica serovars using real-time PCR
    Csordas, AT
    Barak, JD
    Delwiche, MJ
    LETTERS IN APPLIED MICROBIOLOGY, 2004, 39 (02) : 187 - 193
  • [30] Quantitative detection of Salmonella typhimurium contamination in milk, using real-time PCR
    Jung, SJ
    Kim, HJ
    Kim, HY
    JOURNAL OF MICROBIOLOGY AND BIOTECHNOLOGY, 2005, 15 (06) : 1353 - 1358