The evaluation of a 2D diode array in "magic phantom" for use in high dose rate brachytherapy pretreatment quality assurance

被引:22
|
作者
Espinoza, A. [1 ]
Petasecca, M. [1 ]
Fuduli, I. [1 ]
Howie, A. [2 ]
Bucci, J. [2 ]
Corde, S. [3 ]
Jackson, M. [3 ]
Lerch, M. L. F. [1 ]
Rosenfeld, A. B. [1 ]
机构
[1] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW 2522, Australia
[2] St George Hosp, Canc Care Ctr, Kogarah, NSW 2217, Australia
[3] Prince Wales Hosp, Dept Radiat Oncol, Randwick, NSW 2031, Australia
基金
澳大利亚研究理事会;
关键词
brachytherapy; quality assurance; magic plate; gamma index; diode array; VERIFICATION; DOSIMETRY; DETECTOR;
D O I
10.1118/1.4905233
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: High dose rate (HDR) brachytherapy is a treatment method that is used increasingly worldwide. The development of a sound quality assurance program for the verification of treatment deliveries can be challenging due to the high source activity utilized and the need for precise measurements of dwell positions and times. This paper describes the application of a novel phantom, based on a 2D 11x11 diode array detection system, named "magic phantom" (MPh), to accurately measure plan dwell positions and times, compare them directly to the treatment plan, determine errors in treatment delivery, and calculate absorbed dose. Methods: The magic phantom system was CT scanned and a 20 catheter plan was generated to simulate a nonspecific treatment scenario. This plan was delivered to the MPh and, using a custom developed software suite, the dwell positions and times were measured and compared to the plan. The original plan was also modified, with changes not disclosed to the primary authors, and measured again using the device and software to determine the modifications. A new metric, the "position-time gamma index," was developed to quantify the quality of a treatment delivery when compared to the treatment plan. The MPh was evaluated to determine the minimum measurable dwell time and step size. The incorporation of the TG-43U1 formalism directly into the software allows for dose calculations to be made based on the measured plan. The estimated dose distributions calculated by the software were compared to the treatment plan and to calibrated EBT3 film, using the 2D gamma analysis method. Results: For the original plan, the magic phantom system was capable of measuring all dwell points and dwell times and the majority were found to be within 0.93 mm and 0.25 s, respectively, from the plan. By measuring the altered plan and comparing it to the unmodified treatment plan, the use of the position-time gamma index showed that all modifications made could be readily detected. The MPh was able to measure dwell times down to 0.067 +/- 0.001 s and planned dwell positions separated by 1 mm. The dose calculation carried out by the MPh software was found to be in agreement with values calculated by the treatment planning system within 0.75%. Using the 2D gamma index, the dose map of the MPh plane and measured EBT3 were found to have a pass rate of over 95% when compared to the original plan. Conclusions: The application of this magic phantom quality assurance system to HDR brachytherapy has demonstrated promising ability to perform the verification of treatment plans, based upon the measured dwell positions and times. The introduction of the quantitative position-time gamma index allows for direct comparison of measured parameters against the plan and could be used prior to patient treatment to ensure accurate delivery. (C) 2015 American Association of Physicists in Medicine.
引用
收藏
页码:663 / 673
页数:11
相关论文
共 50 条
  • [1] "Magic phantom" - a quality assurance system for high dose rate brachytherapy
    Espinoza, A.
    Petasecca, M.
    Fuduli, I.
    Howie, A.
    Corde, S.
    Bucci, J.
    Jackson, M.
    Lerch, M. L. F.
    Rosenfeld, A. B.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S139 - S140
  • [2] The feasibility study and characterization of a two-dimensional diode array in "magic phantom" for high dose rate brachytherapy quality assurance
    Espinoza, A.
    Beeksma, B.
    Petasecca, M.
    Fuduli, I.
    Porumb, C.
    Cutajar, D.
    Corde, S.
    Jackson, M.
    Lerch, M. L. F.
    Rosenfeld, A. B.
    [J]. MEDICAL PHYSICS, 2013, 40 (11)
  • [3] Evaluation of a 2D diode array for IMRT quality assurance
    Létourneau, D
    Gulam, M
    Yan, D
    Oldham, M
    Wong, JW
    [J]. RADIOTHERAPY AND ONCOLOGY, 2004, 70 (02) : 199 - 206
  • [4] Pretreatment verification of high dose rate brachytherapy plans using the 'magic phantom' system
    Espinoza, A.
    Petasecca, M.
    Cutajar, D.
    Fuduli, I.
    Howie, A.
    Bucci, J.
    Corde, S.
    Jackson, M.
    Zaider, M.
    Lerch, M. L. F.
    Rosenfeld, A. B.
    [J]. BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2015, 1 (02):
  • [5] Initial Evaluation of a 2D Diode Array for Quality Assurance of Radiosurgery Treatments
    Carlos Ronquillo-Gomez, Roberto
    Hernandez-Bojorquez, Mariana
    Javier Uriarte-Rivera, Hector
    de la Mata, Dolores
    Moreno-Jimenez, Sergio
    Flores-Vazquez, Fabiola
    [J]. XV MEXICAN SYMPOSIUM ON MEDICAL PHYSICS, 2019, 2090
  • [6] On the potential of 2D ion chamber arrays for high-dose rate remote afterloading brachytherapy quality assurance
    Lekatou, Aristea
    Peppa, Vasiliki
    Karaiskos, Pantelis
    Pantelis, Evangelos
    Papagiannis, Panagiotis
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (08):
  • [7] 2D linear array device as a quality assurance tool in brachytherapy applications
    Gowri, B.
    Radha, C. Anu
    Ramasubramanian, V.
    [J]. INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2017, 15 (04): : 371 - 376
  • [8] Evaluation of the Performance of a Multiple-Array-Diode Detector for Quality Assurance Tests in High-Dose-Rate Brachytherapy with Ir-192 Source
    Harpool, K.
    Herman, T. De la Fuente
    Ahmad, S.
    Ali, I.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3468 - 3468
  • [9] Design and testing of a phantom and instrumented gynecological applicator based on GaN dosimeter for use in high dose rate brachytherapy quality assurance
    Guiral, P.
    Ribouton, J.
    Jalade, P.
    Wang, R.
    Galvan, J. -M.
    Lu, G. -N.
    Pittet, P.
    Rivoire, A.
    Gindraux, L.
    [J]. MEDICAL PHYSICS, 2016, 43 (09) : 5240 - 5251
  • [10] Quality assurance of asymmetric jaw alignment using 2D diode array
    Kim, Sun Mo
    Chmielewski, Renata
    Abbas, Ahmar
    Yeung, Ivan W. T.
    Moseley, Douglas J.
    [J]. MEDICAL PHYSICS, 2013, 40 (12)