FASCAL 2: a new NIST facility for the calibration of the spectral irradiance of sources

被引:17
|
作者
Yoon, HW [1 ]
Proctor, JE [1 ]
Gibson, CE [1 ]
机构
[1] Natl Inst Stand & Technol, Opt Technol Div, Gaithersburg, MD 20899 USA
关键词
D O I
10.1088/0026-1394/40/1/308
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The facility for automated spectroradiometric calibrations (FASCAL) is the primary facility for calibration of spectral irradiance and spectral radiance at NIST and has been in continuous use since the early 1970s. Due to the increasing demands for spectroradiometric calibration, especially for supporting the monitoring of global environmental changes, a new facility, FASCAL 2, dedicated to calibrating spectral irradiance has been built. This facility will enable faster responses to calibration requests and, ultimately, result in lower uncertainties in the disseminated spectral irradiances. The FASCAL 2 facility is designed with the objective of achieving a signal-to-noise ratio exceeding 1000: 1 from 250 nm to 2500 nm in a bandwidth of 4 nm to 8 nm when measuring a 1000 W FEL lamp at a distance of 50 cm with a receiving aperture of 1 cm(2). The facility will also be capable of calibrating deuterium lamps from 200 nm to 400 rim. The facility has six independent source stations, with four of the stations dedicated to measurements with spectral irradiance lamps and two stations reserved for the realization of spectral irradiance scales and checking the accuracy of automated wavelength. After verifying that the calibrations of spectral irradiance performed in FASCAL and FASCAL 2 agree within their combined uncertainties, FASCAL 2 will become the primary NIST facility for calibration of spectral irradiance.
引用
收藏
页码:S30 / S34
页数:5
相关论文
共 50 条
  • [1] NIST facility for spectral irradiance and radiance responsivity calibrations with uniform sources
    Brown, SW
    Eppeldauer, GP
    Lykke, KR
    METROLOGIA, 2000, 37 (05) : 579 - 582
  • [2] Spectral irradiance and radiance responsivity calibrations using uniform sources (SIRCUS) facility at NIST
    Brown, SW
    Eppeldauer, GP
    Rice, JP
    Zhang, J
    Lykke, KR
    EARTH OBSERVING SYSTEMS IX, 2004, 5542 : 363 - 374
  • [3] Facility for calibration of ultraviolet and vacuum ultraviolet spectral irradiance
    Chen, Y. (guangweisun@163.com), 1600, Chinese Society of Astronautics (42):
  • [4] NIST low-background infrared spectral calibration facility
    Lorentz, SR
    Ebner, SC
    Walker, JH
    Datla, RU
    METROLOGIA, 1996, 32 (06) : 621 - 624
  • [5] NIST display colorimeter calibration facility
    Brown, S
    Ohno, Y
    FOURTH OXFORD CONFERENCE ON SPECTROSCOPY, 2003, 4826 : 137 - 145
  • [6] NIST calibration facility for display colorimeters
    Brown, SW
    Ohno, Y
    FLAT PANEL DISPLAY TECHNOLOGY AND DISPLAY METROLOGY, 1999, 3636 : 162 - 169
  • [7] Facility for spectral irradiance and radiance responsivity calibrations using uniform sources
    Brown, Steven W.
    Eppeldauer, George P.
    Lykke, Keith R.
    APPLIED OPTICS, 2006, 45 (32) : 8218 - 8237
  • [8] NIST facility for spectral calibration of detectors: Calibration of arsenic doped silicon blocked impurity band detectors
    Carter, AC
    Lorentz, SR
    Jung, TM
    Klemme, BJ
    Datla, RU
    INFRARED DETECTORS AND FOCAL PLANE ARRAYS VI, 2000, 4028 : 420 - 425
  • [9] Comparison of spectral irradiance scales between the NIST and the HUT
    Kübarsepp, T
    Yoon, HW
    Nevas, S
    Kärhä, P
    Ikonen, E
    METROLOGIA, 2002, 39 (04) : 399 - 402
  • [10] Infrared spectral emissivity characterization facility at NIST
    Hanssen, L
    Mekhontsev, S
    Khromchenko, V
    THERMOSENSE XXVI, 2004, 5405 : 1 - 12