High power calibration of commercial power meters using an NIST-traceable secondary standard

被引:0
|
作者
Lander, ML [1 ]
Bagford, JO [1 ]
Seibert, DB [1 ]
Hull, RJ [1 ]
机构
[1] WRIGHT LAB,MAT DIRECTORATE,WRIGHT PATTERSON AFB,OH 45433
关键词
laser power measurement; power meter calibration; NIST standard;
D O I
10.2351/1.4745410
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
For the past 19 years, the Laser Hardened Materials Evaluation Laboratory (LHMEL), located at Wright-Patterson AFB, OH, has maintained a number of calibrated high energy power measurement devices capable of measuring CO2 powers up to 150 kW. These devices, calibrated annually to the National Institute for Standards & Technology (NIST) high energy standard calorimeters, serve as secondary standards for in-house calibration of LHMEL's commercial power heads and calorimeters. Recent discussions with academic and industrial laser users have identified an industry need for traceable calibration sources above the 1 kW level. This paper describes the methods used to calibrate LHMEL's commercial power measurement devices using the LHMEL secondary standards and also describes the process by which other laser users can access these same calibration capabilities to confirm the accuracy of their own power heads and calorimeters. The LHMEL calibration capability offers academic and industrial laser users the opportunity to economically calibrate existing power measurement devices on demand to an NIST-traceable secondary standard. The method described has been applied to calibrate systems up to 100 kW while maintaining measurement accuracies to within +/-6.3%.
引用
收藏
页码:103 / 107
页数:5
相关论文
共 50 条
  • [21] Multicenter study of quantitative PET system harmonization using NIST-traceable 68Ge/68Ga cross-calibration kit
    Miwa, Kenta
    Wagatsuma, Kei
    Iimori, Takashi
    Sawada, Koichi
    Kamiya, Takashi
    Sakurai, Minoru
    Miyaji, Noriaki
    Murata, Taisuke
    Sato, Eisuke
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2018, 52 : 98 - 103
  • [22] Calibration of wafer temperature to NIST traceable standards using an isothermal cavity
    Vandenabeele, P
    Renken, W
    RAPID THERMAL AND INTEGRATED PROCESSING VII, 1998, 525 : 115 - 120
  • [23] Using Commercial Source Measure Units for Traceable RF Power Measurements
    Gray, Daniel C.
    Hagerstrom, Aaron M.
    Roberts, Zenn C.
    Long, Christian J.
    2024 103RD ARFTG MICROWAVE MEASUREMENT CONFERENCE, ARFTG 2024, 2024,
  • [24] Confirmation of ethanol compressed gas standard concentrations by an NIST-traceable, absolute chemical method and comparison with wet breath alcohol simulators
    Silverman, LD
    Wong, K
    Miller, S
    JOURNAL OF ANALYTICAL TOXICOLOGY, 1997, 21 (05) : 369 - 372
  • [25] Traceable technique to calibrate current coils for calibration of the power clamp meters in AC current range up to 1000 A
    Olencki, Andrzej
    Mroz, Piotr
    MEASUREMENT, 2017, 109 : 366 - 372
  • [26] Comparison of optical-power meters between the NIST and the PTB
    Vayshenker, I
    Haars, H
    Li, X
    Lehman, JH
    Livigni, DJ
    METROLOGIA, 2000, 37 (04) : 349 - 350
  • [27] High-precision linearity standard for high-power optical-fiber power meters
    Mukai, S.
    Amemiya, K.
    Endo, M.
    2008 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 2008, : 288 - 289
  • [28] A traceable mains-frequency power standard
    Corney, AC
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1999, 48 (02) : 418 - 421
  • [29] A traceable mains-frequency power standard
    Corney, AC
    1998 CONFERENCE ON PRECISION ELECTROMAGNETIC MEASUREMENTS DIGEST, 1998, : 281 - 282
  • [30] CALIBRATION OF FIBER OPTICAL POWER METERS AT PTB
    STOCK, KD
    NEW DEVELOPMENTS AND APPLICATIONS IN OPTICAL RADIOMETRY, 1989, 92 : 159 - 166