Comparison of IFCC and NGSP methods for determination of glycated haemoglobin using advanced regression techniques

被引:0
|
作者
Mrazova, Viera [1 ]
Mocak, Jan [1 ]
Bednarova, Adriana [1 ]
Balla, Jan [2 ]
机构
[1] Univ SS Cyril & Methodius, Dept Chem, SK-91701 Trnava, Slovakia
[2] Analyt Diagnost Lab, SK-08001 Presov, Slovakia
来源
CENTRAL EUROPEAN JOURNAL OF CHEMISTRY | 2010年 / 8卷 / 06期
关键词
Glycated haemoglobin; Diabetes mellitus; Method comparison; Advanced regression; CLINICAL-CHEMISTRY; LINEAR-REGRESSION; STATISTICAL-METHODS; UNITED-STATES; HUMAN BLOOD; HBA(1C); STANDARDIZATION; AGREEMENT; VALUES; SWEDEN;
D O I
10.2478/s11532-010-0096-x
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
The objective of this work was a correct statistical comparison of two assay methods for determination of glycated haemoglobin HbA1c. The immunoturbidimetric determination of HbA1c was performed in two ways: using an automatic analyser Hitachi 912, calibrated according to the IFCC reference system (International Federation of Clinical Chemistry and Laboratory Medicine) and using an analyser Advia 1200 and the NGSP reference system (National Glycohaemoglobin Standardization Program). For statistical comparison of these two analytical methods several advanced regression methods were used, which respect random errors of both compared methods. Specifically, Deming regression with and without weights, orthogonal regression, and Passing-Bablok regression were employed. The results demonstrate that the investigated analytical assay methods do not correspond to each other. The summarized results indicate usefulness of better harmonisation of two existing reference systems.
引用
收藏
页码:1216 / 1222
页数:7
相关论文
共 50 条
  • [1] Evaluation of Turbidimetric Inhibition Immunoassay (TINIA) and HPLC Methods for Glycated Haemoglobin Determination
    Genc, Sema
    Omer, Beyhan
    Aycan-Ustyol, Esra
    Ince, Nurhan
    Bal, Fatih
    Gurdol, Figen
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2012, 26 (06) : 481 - 485
  • [2] Evaluating new HbA1c methods for adoption by the IFCC and NGSP reference networks using international quality targets
    Lenters-Westra, Erna
    English, Emma
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2017, 55 (09) : 1426 - 1434
  • [3] Comparison of haemoglobin estimates using direct & indirect cyanmethaemoglobin methods
    Bansal, Priyanka Gupta
    Toteja, Gurudayal Singh
    Bhatia, Neena
    Gupta, Sanjeev
    Kaur, Manpreet
    Adhikari, Tulsi
    Garg, Ashok Kumar
    INDIAN JOURNAL OF MEDICAL RESEARCH, 2016, 144 : 566 - 571
  • [4] SPATIAL REGRESSION METHODS IN DENDROCLIMATOLOGY - A REVIEW AND COMPARISON OF 2 TECHNIQUES
    COOK, ER
    BRIFFA, KR
    JONES, PD
    INTERNATIONAL JOURNAL OF CLIMATOLOGY, 1994, 14 (04) : 379 - 402
  • [5] USING LINEAR-REGRESSION FOR METHODS COMPARISON
    PAULSON, R
    WACHTEL, M
    LABORATORY MEDICINE, 1995, 26 (07) : 464 - 469
  • [6] Glycated hemoglobin (HbA1c) determination: A comparison of assay methods.
    Rebolledo, OR
    Dato, SMA
    Brizzio, AM
    Gagliardino, JJ
    ACTA BIOQUIMICA CLINICA LATINOAMERICANA, 2001, 35 (03): : 355 - 361
  • [7] Determination of HbA1c in the presence of haemoglobin variants:: Comparison of three HPLC techniques
    Carta, M
    Dall'Olio, G
    Soffiati, G
    EUROPEAN JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1997, 35 (12): : 923 - 925
  • [8] System reliability methods using advanced sampling techniques
    Schueremans, L
    Van Gemert, D
    SAFETY AND RELIABILITY, VOLS 1 AND 2, 2003, : 1425 - 1432
  • [9] A comparison of advanced regression techniques for predicting ship CO2 emissions
    Lepore, Antonio
    dos Reis, Marco Seabra
    Palumbo, Biagio
    Rendall, Ricardo
    Capezza, Christian
    QUALITY AND RELIABILITY ENGINEERING INTERNATIONAL, 2017, 33 (06) : 1281 - 1292
  • [10] Advanced regression methods for combustion modelling using principal components
    Isaac, Benjamin J.
    Thornock, Jeremy N.
    Sutherland, James
    Smith, Philip J.
    Parente, Alessandro
    COMBUSTION AND FLAME, 2015, 162 (06) : 2592 - 2601