Evaluation of Turbidimetric Inhibition Immunoassay (TINIA) and HPLC Methods for Glycated Haemoglobin Determination

被引:25
|
作者
Genc, Sema [1 ]
Omer, Beyhan [1 ]
Aycan-Ustyol, Esra [1 ]
Ince, Nurhan [2 ]
Bal, Fatih [1 ]
Gurdol, Figen [1 ]
机构
[1] Istanbul Univ, Istanbul Fac Med, Dept Biochem, TR-34093 Istanbul, Turkey
[2] Istanbul Univ, Istanbul Fac Med, Dept Publ Hlth, TR-34093 Istanbul, Turkey
关键词
HbA1c; turbidimetric immunoassay (TINIA); high-performance liquid chromatography (HPLC); diabetes mellitus; GLYCOHEMOGLOBIN STANDARDIZATION; UNITED-STATES; A(1C); CHROMATOGRAPHY; INTERFERENCE; VARIANTS; HBA(1C); SYSTEM; ASSAYS; JAPAN;
D O I
10.1002/jcla.21550
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background Various factors may affect the accuracy of hemoglobin (Hb) A1c measurements that are widely used to monitor glycemic control in diabetic patients. This study was aimed to compare the values of HbA1c obtained by two different methods, Roche Tina-quant second and thirdgeneration HbA1c assays based on the turbidimetric inhibition immunoassay (TINIA), and high-performance liquid chromatography (HPLC) cation-exchange method used by Arkray Adams HA-8160 analyzer. Methods Measurements of HbA1c were carried out in blood samples from 2,917 patients using above-mentioned methods. Linear regression was used for the correlation analysis and linear equations. BlandAltman plots were performed from method comparison data using MedCalc statistical software. Results For the low control, the second generation Tina-quant assay had within-run and between-run CVs 0.8% and 0.9%; for the high control within-run and between-run CVs were 1% and 0.96%, respectively. HPLC method for the low control had within-run CV 1% and between-run CV 1.3%; for the high control within-run CV was 0.6% and between-run CV was 0.9%. Conclusion There was a good concordance between the results of TINIA and HPLC methods (y = 1.091x 0.363; r(2) = 0.96). J. Clin. Lab. Anal. 26:481-485, 2012. (C) 2012 Wiley Periodicals, Inc.
引用
收藏
页码:481 / 485
页数:5
相关论文
共 45 条
  • [1] Comparison of Turbidimetric Inhibition Immunoassay, High-Performance Liquid Chromatography, and Capillary Electrophoresis Methods for Glycated Hemoglobin Determination
    Gilani, Mehwish
    Aamir, Mohammad
    Akram, Ammad
    Haroon, Zujaja Hina
    Ijaz, Aamir
    Khadim, Muhammad Tahir
    LABORATORY MEDICINE, 2020, 51 (06) : 579 - 584
  • [2] A comparison between turbidimetric inhibition immunoassay and capillary electrophoresis in glycated hemoglobin (HbA1c) measurement
    Aksungar, Fehime Benli
    Serteser, Mustafa
    Coskun, Abdurrahman
    Unsal, Ibrahim
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2013, 51 (08) : E191 - E193
  • [3] Comparison of IFCC and NGSP methods for determination of glycated haemoglobin using advanced regression techniques
    Mrazova, Viera
    Mocak, Jan
    Bednarova, Adriana
    Balla, Jan
    CENTRAL EUROPEAN JOURNAL OF CHEMISTRY, 2010, 8 (06): : 1216 - 1222
  • [4] Evaluation and interference study of hemoglobin A1c measured by turbidimetric inhibition immunoassay
    Chang, J
    Hoke, C
    Ettinger, B
    Penerian, G
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1998, 109 (03) : 274 - 278
  • [5] THE DIFFERENCE IN THE INTERFERENCE OF FETAL HEMOGLOBIN (HBF) BETWEEN CATION-EXCHANGE HPLC AND TURBIDIMETRIC INHIBITION IMMUNOASSAY HBA1C METHODS
    Zhu, Y.
    Williams, L. M.
    Andrews, D. M.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S408 - S408
  • [6] Simultaneous determination of glycated haemoglobin, a long term biomarker of diabetes mellitus, and total haemoglobin by isotope dilution and HPLC-ICP-MS
    del Castillo Busto, M. Estela
    Montes-Bayon, Maria
    Anon, Elena
    Sanz-Medel, Alfredo
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2008, 23 (05) : 758 - 764
  • [7] Determination of fumonisins in corn: Evaluation of competitive immunoassay and HPLC techniques
    Sydenham, EW
    Shephard, GS
    Thiel, PG
    Bird, C
    Miller, BM
    JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 1996, 44 (01) : 159 - 164
  • [8] Evaluation of a Commercially Available Human Serum Amyloid A (SAA) Turbidimetric Immunoassay for Determination of Feline SAA Concentration
    A. E. Hansen
    M. K. Schaap
    M. Kjelgaard-Hansen
    Veterinary Research Communications, 2006, 30 : 863 - 872
  • [9] Evaluation of a commercially available human serum amyloid A (SAA) turbidimetric immunoassay for determination of feline SAA concentration
    Hansen, A. E.
    Schaap, M. K.
    Kjelgaard-Hansen, M.
    VETERINARY RESEARCH COMMUNICATIONS, 2006, 30 (08) : 863 - 872
  • [10] DETERMINATION OF WEATHERED ATRAZINE RESIDUES IN SOIL BY ENZYME-IMMUNOASSAY AND HPLC - AN EVALUATION STUDY
    SCHEWES, R
    WUST, S
    LEPSCHYVONGLEISSENTHALL, J
    MAIDL, FX
    SUSS, A
    HOCK, B
    FISCHBECK, G
    ANALYTICAL LETTERS, 1994, 27 (03) : 487 - 494