Acoustic power measurement of high intensity focused ultrasound in medicine based on radiation force

被引:35
|
作者
Shou, Wende [1 ]
Huang, Xiaowei
Duan, Shimei
Xia, Rongmin
Shi, Zhonglong
Geng, Xiaoming
Li, Faqi
机构
[1] Shanghai Jiao Tong Univ, Dept Biomed Engn, Shanghai 200030, Peoples R China
[2] Chongqing Univ Med Sci, Inst Ultrasound Engn Med, Chongqing, Peoples R China
基金
中国国家自然科学基金;
关键词
HIFU; acoustic power; measurement; radiation force;
D O I
10.1016/j.ultras.2006.06.034
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
How to measure the acoustic power of HIFU is one of the most important tasks in its medical application. In the paper a whole series of formula for calculating the radiation force related to the acoustic power radiated by a single element focusing transducer and by the focusing transducer array were given. Various system of radiation force balance ( RFB) to measure the acoustic power of HIFU in medicine were designed and applied in China. In high power experiments, the dependence of radiation force acting the absorbing target on the target position at the beam axis of focusing transducer was fined. There is a peak value of `` radiation force'' acting the absorbing target in the focal region when the acoustic power through the focal plane exceeds some threshold. In order to avoid this big measurement error caused by the ` peak effect' in focal region, the distance between the absorbing target of RFB and the focusing transducer or transducer array was defined to be equal to or less than 0.7 times of the focal length in the National Standard of China for the measurements of acoustic power and field characteristics of HIFU. More than six different therapeutic equipments of HIFU have been examined by RFB for measuring the acoustic power since 1998. These results show that RFB with the absorbing target is valid in the acoustic power range up to 500 W with good linearity for the drive voltage squared of focusing transducer or array. The uncertainty of measurement is within +/- 15%. (c) 2006 Elsevier B. V. All rights reserved.
引用
收藏
页码:E17 / E20
页数:4
相关论文
共 50 条
  • [31] Effect of difference in shear modulus of biological tissue on heat source distribution of high-intensity focused ultrasound estimated by acoustic radiation force imaging
    Yabata, Hiroki
    Umemura, Shin-ichiro
    Yoshizawa, Shin
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2021, 60 (SD)
  • [32] Radiation force calculation and acoustic power measurement for a cylindrical concave transducer based on the ray acoustic model
    Lili Yu
    Wende Shou
    Chun Hui
    Bing Hu
    Journal of the Korean Physical Society, 2012, 61 : 544 - 550
  • [33] Radiation force calculation and acoustic power measurement for a cylindrical concave transducer based on the ray acoustic model
    Yu, Lili
    Shou, Wende
    Hui, Chun
    Hu, Bing
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2012, 61 (04) : 544 - 550
  • [34] Image-Guided Measurement of Radiation Force Induced by Focused Ultrasound Beams
    Sahoo, Abhishek
    He, Huijing
    Darrow, David
    Chen, Clark C. C.
    Ebbini, Emad S. S.
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2023, 70 (02) : 138 - 146
  • [35] Optical fiber-based acoustic intensity microphone for high-intensity airborne ultrasound measurement
    Hoshi, Takayuki
    O-oka, Yoshiki
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2024, 63 (04)
  • [36] Acoustic-electroelastic modeling of piezoelectric disks in high-intensity focused ultrasound power transfer systems
    Bhargava, Aarushi
    Shahab, Shima
    JOURNAL OF INTELLIGENT MATERIAL SYSTEMS AND STRUCTURES, 2021, 32 (11) : 1215 - 1233
  • [37] Temperature Elevation of Biological Tissue Model Exposed by Focused Ultrasound with Acoustic Radiation Force
    Nitta, Naotaka
    Kudo, Nobuki
    Akiyama, Iwaki
    NONLINEAR ACOUSTICS: STATE-OF-THE-ART AND PERSPECTIVES (ISNA 19), 2012, 1474 : 264 - 267
  • [38] Measurement and numerical simulation of high intensity focused ultrasound field in water
    Kang Il Lee
    Journal of the Korean Physical Society, 2017, 71 : 665 - 669
  • [39] Focused ultrasound setup for the study of acoustic radiation force induced biological effects in cells
    Chandrasekaran, Sri Nivas
    Holm, Sverre
    Prieur, Fabrice
    2017 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2017,
  • [40] Measurement of the Field Characteristics from High Intensity Focused Ultrasound Transducer
    Umchid, Sumet
    2014 7TH BIOMEDICAL ENGINEERING INTERNATIONAL CONFERENCE (BMEICON), 2014,