Acoustic power measurement of high intensity focused ultrasound in medicine based on radiation force

被引:35
|
作者
Shou, Wende [1 ]
Huang, Xiaowei
Duan, Shimei
Xia, Rongmin
Shi, Zhonglong
Geng, Xiaoming
Li, Faqi
机构
[1] Shanghai Jiao Tong Univ, Dept Biomed Engn, Shanghai 200030, Peoples R China
[2] Chongqing Univ Med Sci, Inst Ultrasound Engn Med, Chongqing, Peoples R China
基金
中国国家自然科学基金;
关键词
HIFU; acoustic power; measurement; radiation force;
D O I
10.1016/j.ultras.2006.06.034
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
How to measure the acoustic power of HIFU is one of the most important tasks in its medical application. In the paper a whole series of formula for calculating the radiation force related to the acoustic power radiated by a single element focusing transducer and by the focusing transducer array were given. Various system of radiation force balance ( RFB) to measure the acoustic power of HIFU in medicine were designed and applied in China. In high power experiments, the dependence of radiation force acting the absorbing target on the target position at the beam axis of focusing transducer was fined. There is a peak value of `` radiation force'' acting the absorbing target in the focal region when the acoustic power through the focal plane exceeds some threshold. In order to avoid this big measurement error caused by the ` peak effect' in focal region, the distance between the absorbing target of RFB and the focusing transducer or transducer array was defined to be equal to or less than 0.7 times of the focal length in the National Standard of China for the measurements of acoustic power and field characteristics of HIFU. More than six different therapeutic equipments of HIFU have been examined by RFB for measuring the acoustic power since 1998. These results show that RFB with the absorbing target is valid in the acoustic power range up to 500 W with good linearity for the drive voltage squared of focusing transducer or array. The uncertainty of measurement is within +/- 15%. (c) 2006 Elsevier B. V. All rights reserved.
引用
收藏
页码:E17 / E20
页数:4
相关论文
共 50 条
  • [21] Magnetic Resonance Acoustic Radiation Force Imaging (MR-ARFI) for the monitoring of High Intensity Focused Ultrasound (HIFU) ablation in anisotropic tissue
    Choquet, Karine
    Vappou, Jonathan
    Cabras, Paolo
    Ishak, Ounay
    Gangi, Afshin
    Breton, Elodie
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2023, 36 (05) : 737 - 747
  • [22] Comparison between thermal strain and acoustic radiation force imaging methods for estimation of heat source distribution of high-intensity focused ultrasound
    Obara, Nozomi
    Umemura, Shin-ichiro
    Yoshizawa, Shin
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2021, 60 (SD)
  • [23] A study on correction method of total acoustic power in presence of shielding in focused ultrasound treatment using acoustic radiation force imaging
    Numahata, Erika
    Yoshizawa, Shin
    Umemura, Shin-ichiro
    Ishii, Takuro
    Saijo, Yoshifumi
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2023, 62 (SJ)
  • [24] Accurate acoustic power measurement for low-intensity focused ultrasound using focal axial vibration velocity
    Tao, Chenyang
    Guo, Gepu
    Ma, Qingyu
    Tu, Juan
    Zhang, Dong
    Hu, Jimin
    JOURNAL OF APPLIED PHYSICS, 2017, 122 (01)
  • [25] Accurate acoustic power measurement for low-intensity focused ultrasound using focal axial vibration velocity
    Guo, Gepu (guogepu@njnu.edu.cn), 1600, American Institute of Physics Inc. (122):
  • [26] Reduction of acoustic reverberation noise from therapeutic transducer for estimation of heat source distribution by acoustic radiation force imaging for high-intensity focused ultrasound treatment
    Kodama, Konosuke
    Mori, Shohei
    Yoshizawa, Shin
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2025, 64 (03)
  • [27] Acoustic radiation force and streaming induced by focused nonlinear ultrasound in a dissipative medium
    Rudenko, OV
    Sarvazyan, AP
    Emelianov, SY
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1996, 99 (05): : 2791 - 2798
  • [28] Characterization of high intensity focused ultrasound transducers using acoustic streaming
    Hariharan, Prasanna
    Myers, Matthew R.
    Robinson, Ronald A.
    Maruvada, Subha H.
    Sliwa, Jack
    Banerjee, Rupak K.
    Journal of the Acoustical Society of America, 2008, 123 (03): : 1706 - 1719
  • [29] Characterization of high intensity focused ultrasound transducers using acoustic streaming
    Hariharan, Prasanna
    Myers, Matthew R.
    Robinson, Ronald A.
    Maruvada, Subha H.
    Sliwa, Jack
    Banerjee, Rupak K.
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2008, 123 (03): : 1706 - 1719
  • [30] A Localization Method of Lesions Induced by High Intensity Focused Ultrasound Based on Acoustic Backscatter Change
    Zheng, Xinliang
    Vaezy, Shahram
    2008 30TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-8, 2008, : 3673 - 3676