Multifrequency Magnetic Resonance Elastography for the Assessment of Renal Allograft Function

被引:41
|
作者
Garcia, Stephan Rodrigo Marticorena [1 ]
Fischer, Thomas [1 ]
Duerr, Michael [2 ]
Gueltekin, Emin [1 ]
Braun, Juergen [3 ]
Sack, Ingolf [1 ]
Guo, Jing [1 ]
机构
[1] Charite Univ Med Berlin, Dept Radiol, Berlin, Germany
[2] Charite Univ Med Berlin, Dept Nephrol, Berlin, Germany
[3] Charite Univ Med Berlin, Dept Med Informat, Berlin, Germany
关键词
multifrequency magnetic resonance elastography; kidney; renal allograft; shear waves; renal stiffness; renal dysfunction; MR ELASTOGRAPHY; TRANSIENT ELASTOGRAPHY; TISSUE ELASTICITY; KIDNEY; STIFFNESS; FIBROSIS; FEASIBILITY; DIAGNOSIS; HISTOLOGY; BIOPSY;
D O I
10.1097/RLI.0000000000000271
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The aim of this study was to apply multifrequency magnetic resonance elastography (MMRE) for assessment of kidney function based on renal stiffness of allografts in transplant recipients and native kidneys in controls. Methods: In this prospective study, MMRE was used to measure stiffness in transplant kidneys in 22 recipients (age range, 23-73 years; 7 females) and in native kidneys in 11 controls (age range, 26-55 years; 4 females) after internal review board approval. TheMMREwas performed on a 1.5 Tmagnetic resonance imaging scanner using 4 vibration frequencies from 40 to 70 Hz. Stiffness maps were computed by multifrequency reconstruction of the magnitude shear modulus (vertical bar G*vertical bar). Clinical markers such as glomerular filtration rate (GFR) and resistive index (RI) were acquired. Differences in renal stiffness among groups were compared by Mann-Whitney U test. Correlations were tested using Pearson correlation. Results: Functioning transplants had higher stiffness (vertical bar G*vertical bar = 9.00 +/- 1.71 kPa) than nonfunctioning transplants (vertical bar G*vertical bar = 5.88 +/- 1.71 kPa, P < 0.001) and native kidneys (vertical bar G*vertical bar = 6.63 +/- 1.63 kPa, P < 0.01). A cutoff value of 7.04 kPa provided sensitivity (83.33%) and specificity (86.67%) for detecting renal allograft dysfunction with an area under the receiver operating characteristic curve value of 0.9278 (95% confidence interval, 0.83-1.00). vertical bar G*vertical bar correlated positively with GFR (r = 0.52, P = 0.015) and negatively with RI (r = -0.52, P = 0.016). Conclusions: Multifrequency magnetic resonance elastography has good diagnostic accuracy in detecting renal allograft dysfunction. Renal stiffness is significantly lower in recipients with nonfunctioning transplant kidneys and correlates with clinical markers such as GFR and RI.
引用
收藏
页码:591 / 595
页数:5
相关论文
共 50 条
  • [31] Magnetic resonance elastography (MRE) detects medullary renal fibrosis
    Korsmo, Michael James
    Ebrahimi, Behzad
    Woollard, John R.
    Crane, John A.
    Eirin, Alfonso
    Krier, James D.
    Ehman, Richard L.
    Lerman, Lilach O.
    FASEB JOURNAL, 2012, 26
  • [32] NON-INVASIVE ASSESSMENT OF RENAL ALLOGRAFT CHRONICITY LESIONS WITH ULTRASOUND ELASTOGRAPHY
    Ma, M.
    Law, H.
    Tsi, K.
    Chan, K.
    Chan, G.
    Yap, D.
    Mok, M.
    Kwan, T.
    Choy, B.
    Chan, T.
    NEPHROLOGY, 2016, 21 : 160 - 161
  • [33] Magnetic Resonance Elastography
    Litwiller, Daniel V.
    Mariappan, Yogesh K.
    Ehman, Richard L.
    CURRENT MEDICAL IMAGING REVIEWS, 2012, 8 (01) : 46 - 55
  • [34] Kidney Transplantation Multiparametric Functional Magnetic Resonance Imaging for Assessment of Renal Allograft Pathophysiology in Mice
    Hueper, Katja
    Hensen, Bennet
    Gutberlet, Marcel
    Chen, Rongjun
    Hartung, Dagmar
    Barrmeyer, Amelie
    Meier, Martin
    Li, Wang
    Jang, Mi-Sun
    Mengel, Michael
    Wacker, Frank
    Rong, Song
    Gueler, Faikah
    INVESTIGATIVE RADIOLOGY, 2016, 51 (01) : 58 - 65
  • [35] Magnetic resonance elastography
    Muthupillai, R
    Ehman, RL
    NATURE MEDICINE, 1996, 2 (05) : 601 - 603
  • [36] Magnetic resonance elastography
    Sack, I.
    DEUTSCHE MEDIZINISCHE WOCHENSCHRIFT, 2008, 133 (06) : 247 - 251
  • [37] Magnetic resonance elastography for assessment of testicular stiffness in patients with varicocele Testis MR elastography
    Pocan, Suheyl
    Karakas, Levent
    ANNALS OF CLINICAL AND ANALYTICAL MEDICINE, 2023, 14 : 325 - 329
  • [38] COMPARISON OF TRANSIENT ELASTOGRAPHY AND MAGNETIC RESONANCE ELASTOGRAPHY FOR ASSESSMENT OF MASLD/MASH: A RETROSPECTIVE REVIEW
    Mathur, Anandita
    Paredes-Marin, Alejandra
    Ozkaya, Efe
    Rosberger, Sonam
    Sigel, Keith
    Taouli, Bachir
    Bansal, Meena
    HEPATOLOGY, 2024, 80 : S661 - S661
  • [39] Assessment of liver viscoelasticity using multifrequency MR elastography
    Asbach, Patrick
    Klatt, Dieter
    Hamhaber, Uwe
    Braun, Juergen
    Somasundaram, Rajan
    Hamm, Bernd
    Sack, Ingolf
    MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (02) : 373 - 379
  • [40] Preoperative assessment of meningioma stiffness using magnetic resonance elastography
    Murphy, Matthew C.
    Huston, John, III
    Glaser, Kevin J.
    Manduca, Armando
    Meyer, Fredric B.
    Lanzino, Giuseppe
    Morris, Jonathan M.
    Felmlee, Joel P.
    Ehman, Richard L.
    JOURNAL OF NEUROSURGERY, 2013, 118 (03) : 643 - 648