Electrochemical Treatment of Contaminated Titanium Surfaces in Vitro: An Approach for Implant Surface Decontamination

被引:15
|
作者
Al-Hashedi, Ashwaq Ali [1 ,2 ]
Laurenti, Marco [3 ]
Abdallah, Mohamed-Nur [1 ]
Albuquerque, Rubens F., Jr. [4 ]
Tamimi, Faleh [1 ]
机构
[1] McGill Univ, Fac Dent, 3640 Univ St, Montreal, PQ H3A 0C7, Canada
[2] Sanaa Univ, Dept Prosthodont, Fac Dent, Wadi Dhaher Rd, Sanaa, Yemen
[3] Univ Complutense Madrid, Dept Phys Chem, Ave Seneca,2, E-28040 Madrid, Spain
[4] Univ Sao Paulo, Fac Dent Ribeirao Preto, 253 Ave Prof Dr Zeferino Vaz, Ribeirao Preto, SP, Brazil
来源
关键词
Ti implants; microbial biofilm; peri-implantitis; electrochemical treatment; decontamination; INDUCED PERI-IMPLANTITIS; STAPHYLOCOCCUS-EPIDERMIDIS BIOFILMS; CURRENT-INDUCED DETACHMENT; TOTAL HIP-ARTHROPLASTY; DENTAL IMPLANT; RE-OSSEOINTEGRATION; ELECTRIC-CURRENT; FOLLOW-UP; STAINLESS-STEEL; PURE TITANIUM;
D O I
10.1021/acsbiomaterials.6b00265
中图分类号
TB3 [工程材料学]; R318.08 [生物材料学];
学科分类号
0805 ; 080501 ; 080502 ;
摘要
Bacterial contamination on titanium implants can cause inflammation and eventually implant failure. Currently used methods for decontamination of implants have demonstrated limited success. Metal surfaces can be disinfected electrochemically. However, the effect of electrochemical treatments on biofilm-contaminated titanium is largely unknown. We hypothesized that electrochemical treatments are able to safely remove organic contamination and bacteria from titanium implants without altering their surfaces. This study was designed to assess the electrochemical properties of bacteria-contaminated surfaces in order to develop new treatments to clean titanium. Surface morphology, composition, bacterial load, and electrochemical properties of polished titanium discs were analyzed before and after biofilm contamination and subsequent decontamination with various electrochemical methods. The effect of the combination of the electrochemical with titanium brush cleaning was also evaluated. Results were then analyzed and compared to baseline readings (prior to contamination) using repeated measures ANOVA. Biofilm contamination increased the levels of carbon, nitrogen, and live bacteria on titanium surfaces while reducing their open circuit potential and corrosion resistance. Optimized electrochemical treatments with alternating current (-2.3 mA, + 22.5 mu A) and voltages (1.8 V), were bactericidal and able to completely decontaminate saliva-contaminated titanium surfaces within 5 min while preserving surface integrity and histological quality of mammalian tissues. Furthermore, with the aid of mechanical brushing, the optimized electrochemical treatment was able to achieve complete decontamination of biofilm-contaminated Ti surfaces. The electrochemical treatment seems to be promising and well worth investigating for the clinical management of peri-implant infections.
引用
下载
收藏
页码:1504 / 1518
页数:15
相关论文
共 50 条
  • [21] Comparative In Vitro Study of Surface Treatment of Grade II Titanium Biomedical Implant
    Oscar, Decco
    Victor, Beltran
    Jesica, Zuchuat
    Romina, Gudino
    VI LATIN AMERICAN CONGRESS ON BIOMEDICAL ENGINEERING (CLAIB 2014), 2014, 49 : 183 - 186
  • [22] Efficacy of combined chemical and electrochemical decontamination treatments on contaminated healing abutments and their effect on surface topography: An in vitro study
    Kyaw, Thiha Tin
    Abdou, Ahmed
    Nakata, Hidemi
    Pimkhaokham, Atiphan
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2022, 24 (05) : 696 - 708
  • [23] Electrochemical removal of biofilms from titanium dental implant surfaces
    Schneider, Sebastian
    Rudolph, Michael
    Bause, Vanessa
    Terfort, Andreas
    BIOELECTROCHEMISTRY, 2018, 121 : 84 - 94
  • [24] Surface property alterations and osteoblast attachment to contaminated titanium surfaces after different surface treatments: An in vitro study
    Lee, Bor-Shiunn
    Shih, Kuang-Shao
    Lai, Chern-Hsiung
    Takeuchi, Yasuo
    Chen, Yi-Wen
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2018, 20 (04) : 583 - 591
  • [25] Effects of the Er:YAG laser irradiation on titanium implant materials and contaminated implant abutment surfaces
    Matsuyama, T
    Aoki, A
    Oda, S
    Yoneyama, T
    Ishikawa, I
    JOURNAL OF CLINICAL LASER MEDICINE & SURGERY, 2003, 21 (01): : 7 - 17
  • [26] Organic–Inorganic Surface Modifications for Titanium Implant Surfaces
    Lise T. de Jonge
    Sander C. G. Leeuwenburgh
    Joop G. C. Wolke
    John A. Jansen
    Pharmaceutical Research, 2008, 25 : 2357 - 2369
  • [27] Decontamination of Titanium Surface Using Different Methods: An In Vitro Study
    Yan, Sifan
    Li, Min
    Komasa, Satoshi
    Agariguchi, Akinori
    Yang, Yuanyuan
    Zeng, Yuhao
    Takao, Seiji
    Zhang, Honghao
    Tashiro, Yuichiro
    Kusumoto, Tetsuji
    Kobayashi, Yasuyuki
    Chen, Liji
    Kashiwagi, Kosuke
    Matsumoto, Naoyuki
    Okazaki, Joji
    Kawazoe, Takayoshi
    MATERIALS, 2020, 13 (10)
  • [28] Titanium Surface Decontamination: A Systematic Review of In Vitro Comparative Studies
    Francis, Samy
    Perrotti, Vittoria
    Piattelli, Adriano
    Quaranta, Alessandro
    Laculli, Flavia
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2022, 37 (01) : 76 - +
  • [29] Efficacy of an Er:YAG laser in the decontamination of dental implant surfaces: An in vitro study
    AlMoharib, Hani S.
    Steffensen, Bjorn
    Zoukhri, Driss
    Finkelman, Matthew
    Gyurko, Robert
    JOURNAL OF PERIODONTOLOGY, 2021, 92 (11) : 1613 - 1621
  • [30] Microgrooves and Microrugosities in Titanium Implant Surfaces: An In Vitro and In Vivo Evaluation
    Gehrke, Sergio Alexandre
    Cavalcanti de Lima, Jose Henrique
    Rodriguez, Fernando
    Luis Calvo-Guirado, Jose
    Aramburu Junior, Jaime
    Perez-Diaz, Leticia
    Mazon, Patricia
    Manuel Aragoneses, Juan
    De Aza, Piedad N.
    MATERIALS, 2019, 12 (08)