Robustness of sweeping-window arc therapy treatment sequences against intrafractional tumor motion

被引:5
|
作者
Fleckenstein, Jens [1 ]
Hesser, Juergen [1 ]
Wenz, Frederik [1 ]
Lohr, Frank [1 ]
机构
[1] Heidelberg Univ, Univ Med Ctr Mannheim, Dept Radiat Oncol, D-68167 Mannheim, Germany
关键词
VMAT; robust planning; interplay effects; organ motion; time resolved dosimetry; INCORPORATING ORGAN MOTION; VOLUMETRIC MODULATED ARC; RESPIRATORY MOTION; STATISTICAL-ANALYSIS; RADIATION-THERAPY; DOSIMETRIC IMPACT; DOSE DISTRIBUTION; LUNG-CANCER; MONTE-CARLO; IMRT;
D O I
10.1118/1.4914166
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Due to the potentially periodic collimator dynamic in volumetric modulated arc therapy (VMAT) dose deliveries with the sweeping-window arc therapy (SWAT) technique, additional manifestations of dosimetric deviations in the presence of intrafractional motion may occur. With a fast multileaf collimator (MLC), and a flattening filter free dose delivery, treatment times close to 60 s per fraction are clinical reality. For these treatment sequences, the human breathing period can be close to the collimator sweeping period. Compared to a random arrangement of the segments, this will cause a further degradation of the dose homogeneity. Methods: Fifty VMAT sequences of potentially moving target volumes were delivered on a two dimensional ionization chamber array. In order to detect interplay effects along all three coordinate axes, time resolved measurements were performed twice-with the detector aligned in vertical (V) or horizontal (H) orientation. All dose matrices were then moved within a simulation software by a time-dependent motion vector. The minimum relative equivalent uniform dose EUDr,m for all breathing starting phases was determined for each amplitude and period. Furthermore, an estimation of periods with minimum EUD was performed. Additionally, LINAC logfiles were recorded during plan delivery. The MLC, jaw, gantry angle, and monitor unit settings were continuously saved and used to calculate the correlation coefficient between the target motion and the dose weighed collimator motion component for each direction (CC, LR, AP) separately. Results: The resulting EUDr,m were EUDr,m(CCV) = (98.3 +/- 0.6)%, EUDr,m(CCH) = (98.6 +/- 0.5)%, EUDr,m(AP(V)) = (97.7 +/- 0.9)%, and EUDr,m(LRH) = (97.8 +/- 0.9)%. The overall minimum relative EUD observed for 360. arc midventilation treatments was 94.6%. The treatment plan with the shortest period and a minimum relative EUD of less than 97% was found at T = 6.1 s. For a partial 120 degrees arc, an EUDr,m = 92.0% was found. In all cases, a correlation coefficient above 0.5 corresponded to a minimum in EUD. Conclusions: With the advent of fast VMAT delivery techniques, nonrobust treatment sequences for human breathing patterns can be generated. These sequences are characterized by a large correlation coefficient between a target motion component and the corresponding collimator dynamic. By iteratively decreasing the maximum allowed dose rate, a low correlation coefficient and consequentially a robust treatment sequence are ensured. (C) 2015 Author(s). All article content, except where otherwise noted, is licensed under a Creative Commons Attribution 3.0 Unported License.
引用
收藏
页码:1538 / 1545
页数:8
相关论文
共 50 条
  • [41] Synchronized moving aperture radiation therapy (SMART): Superimposing tumor motion on IMRT MLC leaf sequences under realistic delivery limitations
    Jiang, SB
    Neicu, T
    Zygmanski, P
    MEDICAL PHYSICS, 2003, 30 (06) : 1399 - 1399
  • [42] Treatment planning comparison of electron arc therapy and photon intensity modulated radiotherapy for Askin's tumor of chest wall
    Jamema, Swamidas V.
    Sharma, Pramod K.
    Laskar, Siddhartha
    Deshpande, Deepak D.
    Shrivastava, Shyam K.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 (03) : 257 - 265
  • [43] IS THERE A SAFE THERAPEUTIC WINDOW FOR DELIVERY OF CHEMOTHERAPY PRIOR TO INITIATION OF SURGERY AND/OR RADIATION-THERAPY FOR TREATMENT OF THE PRIMARY TUMOR IN CHILDREN WITH RHABDOMYOSARCOMA
    JAFFE, N
    ROTT, EJ
    WOO, SY
    MAOR, MH
    SMITH, M
    BLACK, T
    ANDRASSY, RJ
    INTERNATIONAL JOURNAL OF ONCOLOGY, 1995, 7 (04) : 855 - 862
  • [44] Comparing the Robustness of Intensity-modulated Proton Therapy and Proton-arc Therapy Against Interplay Effects of 4D Robust-optimised Plans for Lung Stereotactic Body Radiotherapy
    Wei, W. G.
    Yu, H.
    Xiao, Q.
    Li, Z. B.
    Li, J.
    Zhang, X. Y.
    Wu, Y. C.
    Qin, T. L.
    Zeng, X. H.
    Song, Y.
    Li, G. J.
    Bai, S.
    CLINICAL ONCOLOGY, 2025, 39
  • [45] Does Pancreatic Head Tumor Motion on the 4D-CT Accurately Predict the Tumor Motion during the Treatment for Patients Receiving Gated Stereotactic Body Radiation Therapy (SBRT)?
    Chen, S.
    Lin, C.
    Schubert, L.
    Li, S.
    Deng, H.
    Enke, C.
    Zhou, S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 81 (02): : S87 - S88
  • [46] A Dosimetric Comparison Between Conventionally Fractionated Tomotherapy, Volumetric Modulated Arc Therapy, Sliding Window and Proton Therapy, and Hypofractionated Robotic Radiosurgery Treatment by Low-Risk Prostate Carcinoma
    Scobioala, S.
    Kittel, C.
    Haverkamp, U.
    Channaoui, M.
    Habibeh, O.
    Elsayad, K.
    Eich, H. T. T.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E281 - E281
  • [47] Stereotactic Ablative Radiotherapy Using CALYPSO® Extracranial Tracking for Intrafractional Tumor Motion Management-A New Potential Local Treatment for Unresectable Locally Advanced Pancreatic Cancer? Results from a Retrospective Study
    Kaucic, Hrvoje
    Kosmina, Domagoj
    Schwarz, Dragan
    Mack, Andreas
    Sobat, Hrvoje
    Cehobasic, Adlan
    Leipold, Vanda
    Andrasek, Iva
    Avdicevic, Asmir
    Mlinaric, Mihaela
    CANCERS, 2022, 14 (11)
  • [48] Quantification of plan robustness against different uncertainty sources for classical and anatomical robust optimized treatment plans in head and neck cancer proton therapy
    Cubillos-Mesias, Macarena
    Troost, Esther G. C.
    Lohaus, Fabian
    Agolli, Linda
    Rehm, Maximilian
    Richter, Christian
    Stuetzer, Kristin
    BRITISH JOURNAL OF RADIOLOGY, 2020, 93 (1107):
  • [49] Dosimetric comparison of volumetric-arc therapy versus sliding window intensity-modulated radiotherapy in postoperative treatment for primary soft tissue sarcoma of the thigh
    Ortiz Gonzalez, Irene
    Morera Cano, Daniel
    Roncero Sanchez, Raquel
    Mateos Salvador, Pedro
    Valencia Blanco, Lourdes
    Vidal Borras, Meritxell
    Aymar Salis, Neus
    Gadea Quinteiro, Jon
    Jimenez Jimenez, Esther
    Font Gelabert, Joan
    Pardo Masferrer, Jose
    ASIA-PACIFIC JOURNAL OF CLINICAL ONCOLOGY, 2019, 15 (06) : 371 - 376
  • [50] Dosimetric Impact of Breathing Motion in Lung Stereotactic Body Radiotherapy Treatment Using Image-Modulated Radiotherapy and Volumetric Modulated Arc Therapy
    Rao, Min
    Wu, Jianzhou
    Cao, Daliang
    Wong, Tony
    Mehta, Vivek
    Shepard, David
    Ye, Jinsong
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 83 (02): : E251 - E256