Use of disposable graduated biopsy forceps improves accuracy of polyp size measurements during endoscopy

被引:22
|
作者
Jin, Hei-Ying [1 ]
Leng, Qiang [1 ]
机构
[1] Nanjing Univ Tradit Chinese Med, Affiliated Hosp 3, Natl Ctr Colorectal Surg, Jiangsu Integrated Colorectal Oncol Ctr, Nanjing 210001, Jiangsu, Peoples R China
关键词
Disposable graduated biopsy forceps; Polyp size measurement; Colonoscopy; Accuracy; LATERALLY SPREADING TUMORS; COLORECTAL-CANCER; SURVEILLANCE; DIAGNOSIS;
D O I
10.3748/wjg.v21.i2.623
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
AIM: To determine the accuracy of endoscopic polyp size measurements using disposable graduated biopsy forceps (DGBF). METHODS: Gradations accurate to 1 mm were assessed with the wire of disposable graduated biopsy forceps. When a polyp was noted, endoscopists determined the width of the polyp; then, the graduated biopsy forceps was inserted and the largest diameter of the tumor was measured. After excision, during surgery or endoscopy, the polyp was measured using the vernier caliper. RESULTS: One hundred and thirty-three colorectal polyps from 119 patients were studied. The mean diameter, by post-polypectomy measurement, was 0.92 +/- 0.69 cm; 83 were < 1 cm, 36 were between 1 and 2 cm, and 14 were > 2 cm. The mean diameter, by visual estimation, was 1.15 +/- 0.88 cm; compared to the actual size measured using vernier calipers, the difference was statistically significant. The mean diameter measured using the DGBF was 0.93 +/- 0.68 cm; compared to the actual size measured using vernier calipers, this difference was not statistically significant. The ratio between the mean size estimated by visual estimation and the actual size was significantly different from that between the mean size estimated using the DGBF and the actual size (1.26 +/- 0.30 vs 1.02 +/- 0.11). CONCLUSION: The accuracy of polyp size estimation was low by visual assessment; however, it improved when the DGBF was used.
引用
收藏
页码:623 / 628
页数:6
相关论文
共 48 条
  • [41] SAMPLING ERRORS IN PETCO(2) MEASUREMENTS DURING THE USE OF DRYDEN DISPOSABLE CO(2) ABSORBER CIRCUIT
    BENYAMIN, RM
    SALEM, MR
    JOSEPH, NJ
    ANESTHESIOLOGY, 1993, 79 (3A) : A561 - A561
  • [42] Measurement of small-bowel polyp size in patients with Peutz-Jeghers syndrome by using reference granules during video capsule endoscopy
    Racz, I.
    Janoki, M.
    Kovacs, V.
    ENDOSCOPY, 2007, 39 : E41 - E41
  • [43] The use of AndroVision eFlow improves the accuracy of sperm concentration determination in AI boars compared to conventional photometric measurements
    Grossfeld, R.
    Simmet, C.
    Hasenfuss, T.
    Jakop, U.
    Schulze, M.
    REPRODUCTION IN DOMESTIC ANIMALS, 2021, 56 : 18 - 18
  • [44] LOW ESOPHAGEAL BIOPSY RATES AND CONCOMITANT PPI USE DURING ENDOSCOPY MAY CONTRIBUTE TO UNDERDIAGNOSIS OF EOSINOPHILIC ESOPHAGITIS IN MEXICO
    Arenas-Martinez, Josealberto S.
    Dellon, Evan S.
    Jasso-Baltazar, Erick A.
    Chavez-Garcia, Luis C.
    Falcon-Antonio, Orlando E.
    Azamar-Llamas, Daniel
    Fernandez-Ramirez, Alfonso J.
    Ruiz-Manriquez, Jesus Alejandro
    Cedillo, Maria F. Garcia
    Zamora-Nava, Luis Eduardo
    Uscanga, Luis
    Coss-Adame, Enrique
    GASTROENTEROLOGY, 2023, 164 (06) : S358 - S359
  • [45] Use of Intraoperative CT Improves Accuracy of Spinal Navigation During Screw Fixation in Cervico-thoracic Region
    Habib, Nermine
    Filardo, Giuseppe
    Distefano, Daniela
    Candrian, Christian
    Reinert, Michael
    Scarone, Pietro
    SPINE, 2021, 46 (08) : 530 - 537
  • [46] Use of intraoperative fluorescein sodium fluorescence to improve the accuracy of tissue diagnosis during stereotactic needle biopsy of high-grade gliomas
    Rey-Dios, Roberto
    Hattab, Eyas M.
    Cohen-Gadol, Aaron A.
    ACTA NEUROCHIRURGICA, 2014, 156 (06) : 1071 - 1075
  • [47] Use of intraoperative fluorescein sodium fluorescence to improve the accuracy of tissue diagnosis during stereotactic needle biopsy of high-grade gliomas
    Roberto Rey-Dios
    Eyas M. Hattab
    Aaron A. Cohen-Gadol
    Acta Neurochirurgica, 2014, 156 : 1071 - 1075
  • [48] The use of 24-h ambulatory blood pressure monitoring (ABPM) during the first cycle of sunitinib improves the diagnostic accuracy and management of hypertension in patients with advanced renal cancer
    Bamias, A.
    Manios, E.
    Karadimou, A.
    Michas, F.
    Lainakis, G.
    Constantinidis, C.
    Deliveliotis, C.
    Zakopoulos, N.
    Dimopoulos, M. A.
    EUROPEAN JOURNAL OF CANCER, 2011, 47 (11) : 1660 - 1668