Evaluating the four-dimensional cone beam computed tomography with varying gantry rotation speed

被引:13
|
作者
Yoganathan, S. A. [1 ]
Maria, K. J. [1 ]
Ali, Shajahan Mohamed [2 ]
Agarwal, Arpita [1 ]
Mishra, Surendra P. [2 ]
Kumar, Shaleen [1 ]
机构
[1] Sanjay Gandhi Postgrad Inst Med Sci, Deparment Radiotherapy, Lucknow, Uttar Pradesh, India
[2] Dr Ram Manohar Lohia Inst Med Sci, Dept Radiat Oncol, Lucknow, Uttar Pradesh, India
来源
BRITISH JOURNAL OF RADIOLOGY | 2016年 / 89卷 / 1060期
关键词
RADIATION-THERAPY; ADAPTIVE RADIOTHERAPY; CT RECONSTRUCTION; TARGET MARGINS; INTRA-FRACTION; MOTION MODEL; TUMOR MOTION; LUNG; VOLUME; VERIFICATION;
D O I
10.1259/bjr.20150870
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The purpose of this work was to evaluate the four-dimensional cone beam CT (4DCBCT) imaging with different gantry rotation speed. Methods: All the 4DCBCT image acquisitions were carried out in Elekta XVI Symmetry (TM) system (Elekta AB, Stockholm, Sweden). A dynamic thorax phantom with tumour mimicking inserts of diameter 1, 2 and 3 cm was programmed to simulate the respiratory motion (4 s) of the target. 4DCBCT images were acquired with different gantry rotation speeds (36 degrees, 50 degrees, 75 degrees, 100 degrees, 150 degrees and 200 degrees min(-1)). Owing to the technical limitation of 4DCBCT system, average cone beam CT (CBCT) images derived from the 10 phases of 4DCBCT were used for the internal target volume (ITV) contouring. ITVs obtained from average CBCT were compared with the four-dimensional CT (4DCT). In addition, the image quality of 4DCBCT was also evaluated for various gantry rotation speeds using Catphan (R) 600 (The Phantom Laboratory Inc., Salem, NY). Results: Compared to 4DCT, the average CBCT underestimated the ITV. The ITV deviation increased with increasing gantry speed (210.8% vs 217.8% for 36 degrees and 200 degrees min(-1) in 3-cm target) and decreasing target size (217.8% vs 226.8% for target diameter 3 and 1 cm in 200 degrees min(-1)). Similarly, the image quality indicators such as spatial resolution, contrast-to-noise ratio and uniformity also degraded with increasing gantry rotation speed. Conclusion: The impact of gantry rotation speed has to be considered when using 4DCBCT for ITV definition. The phantom study demonstrated that 4DCBCT with slow gantry rotation showed better image quality and less ITV deviation. Advances in knowledge: Usually, the gantry rotation period of Elekta 4DCBCT system is kept constant at 4min (50 degrees min(-1)) for acquisition, and any attempt of decreasing/increasing the acquisition duration requires careful investigation. In this study, the 4DCBCT images with different gantry rotation speed were evaluated.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] Quantitation of the four-dimensional computed tomography process
    Lu, W
    Parikh, P
    ElNaqa, I
    Nystrom, M
    Hubenschmidt, J
    Wahab, S
    Mutic, S
    Singh, A
    Christensen, G
    Bradley, JD
    Low, DA
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01): : S286 - S286
  • [42] Slow Gantry Rotation Acquisition Protocol for Four-Dimensional Digital Tomosynthesis
    Maurer, J.
    Pan, T.
    Yin, F.
    MEDICAL PHYSICS, 2009, 36 (06) : 2725 - +
  • [43] Four-dimensional computed tomography: musculoskeletal applications
    Wong, Murray T.
    Wiens, Charmaine
    Kuczynski, Michael
    Manske, Sarah
    Schneider, Prism S.
    CANADIAN JOURNAL OF SURGERY, 2022, 65 (03) : E388 - E393
  • [44] Impact of Norm Selections on the Performance of Four-dimensional Cone-beam Computed Tomography (4DCBCT) using PICCS
    Li, Yinsheng
    Tang, Jie
    Chen, Guang-Hong
    MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [45] Comparison of internal target volumes defined by three-dimensional, four-dimensional, and cone-beam computed tomography images of a motion phantom
    Wang, Yu
    Liu, Tao
    Chen, Huiqin
    Bai, Penggang
    Zhan, Qiong
    Liang, Xiaohua
    ANNALS OF TRANSLATIONAL MEDICINE, 2020, 8 (22)
  • [46] Evaluating the mandible with cone-beam computed tomography
    Stratemann, Scott A.
    Huang, John C.
    Maki, Koutaro
    Hatcher, David C.
    Miller, Arthur J.
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2010, 137 (04) : S58 - S70
  • [47] Variability of four-dimensional computed tomography patient models
    Sonke, Jan-Jakob
    Lebesque, Joos
    Van Herk, Marcel
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 70 (02): : 590 - 598
  • [48] Four-Dimensional Computed Tomography for Localisation of Parathyroid Adenomas
    Leung, H. S.
    Liu, S. Y. W.
    Wong, K. T.
    Yu, S. M.
    King, A. D.
    HONG KONG JOURNAL OF RADIOLOGY, 2023, 26 (03): : 185 - 193
  • [49] Radiation dose reduction in four-dimensional computed tomography
    Li, T
    Schreibmann, E
    Thorndyke, B
    Tillman, G
    Boyer, A
    Koong, A
    Goodman, K
    Xing, L
    MEDICAL PHYSICS, 2005, 32 (12) : 3650 - 3660
  • [50] Mechanical Valve Evaluation with Four-Dimensional Computed Tomography
    Numata, Satoshi
    Tsutsumi, Yasushi
    Monta, Osamu
    Yamazaki, Sachiko
    Seo, Hiroyuki
    Yoshida, Shohei
    Samura, Takaaki
    Ohashi, Hirokazu
    JOURNAL OF HEART VALVE DISEASE, 2013, 22 (06): : 837 - 842