Evaluation of residual patient position variation for spinal radiosurgery using the Novalis image guided system

被引:30
|
作者
Jin, Jian-Yue [1 ]
Ryu, Samuel [1 ,2 ]
Rock, Jack [2 ]
Faber, Kathleen [1 ]
Chen, Qing [1 ]
Ajlouni, Munther [1 ]
Movsas, Benjamin [1 ]
机构
[1] Henry Ford Hosp, Dept Radiat Oncol, Detroit, MI 48202 USA
[2] Henry Ford Hosp, Dept Neurosurg, Detroit, MI 48202 USA
关键词
spinal radiotherapy; image guided radiotherapy; position uncertainty; image fusion; intrafraction motion;
D O I
10.1118/1.2839097
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The Novalis system has been demonstrated to achieve accurate target localization on anthropomorphic phantoms. However, other factors, such as rotational deviation, patient intrafraction motion, and image fusion uncertainty due to patient body deformation, could contribute additional position uncertainty for actual patients. This study evaluates such position uncertainty for spinal radiosurgery patients. Materials and Methods: Fifty-two consecutive spinal radiosurgery patients were included in the study. Rotational deviation was evaluated from 6-deg of freedom (6D) fusion results for all patients. The combined uncertainty of patient motion and image fusion was determined from fusion results of additional kV x-ray images acquired before, during, and after treatment for 25 of the 52 patients. The uncertainty of image fusion was also evaluated by performing 6D fusion ten different times with various regions of interest in the images selected for fusion. This was performed for two patients with L3 and T2 lesions, respectively, for comparison. Results: The mean rotational deviation was 0.7 +/- 1.8, 0.7 +/- 1.5, and 0.7 +/- 1.6 deg along the yaw, roll, and pitch directions, respectively. The combined uncertainty from patient motion and image fusion was 0.1 +/- 0.9, 0.2 +/- 1.2, and 0.2 +/- 1.0 mm in the anteroposterior (AP), longitudinal, and lateral directions, respectively. The uncertainty (standard deviation) due to image fusion was less than 0.28 mm in any direction for the L3 lesion and 0.8 mm in the AP direction for the T2 lesion. Conclusion: Overall position uncertainty for spinal radiosurgery patients has been evaluated. Rotational deviation and patient motion were the main factors contributed to position uncertainty for actual patient treatment. (C) 2008 American Association of Physicists in Medicine.
引用
收藏
页码:1087 / 1093
页数:7
相关论文
共 50 条
  • [31] EVALUATION OF IMAGE-GUIDED POSITIONING FOR FRAMELESS INTRACRANIAL RADIOSURGERY
    Lamba, Michael
    Breneman, John C.
    Warnick, Ronald E.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 74 (03): : 913 - 919
  • [32] Image-guided hypo-fractionated stereotactic radiosurgery to spinal lesions
    Ryu, SI
    Chang, SD
    Kim, DH
    Murphy, MJ
    Le, QT
    Martin, DP
    Adler, JR
    NEUROSURGERY, 2001, 49 (04) : 838 - 846
  • [33] Analysis of Intrafraction Motion in Image-Guided Stereotactic Radiosurgery of Spinal metastases
    Svestad, J. G.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S466 - S466
  • [34] Image-guided and intensity-modulated radiosurgery for patients with spinal metastasis
    Ryu, S
    Yin, FF
    Rock, J
    Zhu, JE
    Chu, A
    Kagan, E
    Rogers, L
    Ajlouni, M
    Rosenblum, M
    Kim, JH
    CANCER, 2003, 97 (08) : 2013 - 2018
  • [35] Image-Guided Stereotactic Spinal Radiosurgery with a Conventional Linear-Accelerator
    Wang, J.
    Rice, R.
    Mundt, A.
    Murphy, K.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [36] CyberKnife image-guided stereotactic radiosurgery in the treatment of benign spinal tumors
    Dodd, RL
    Gibbs, I
    Adler, J
    Chang, SD
    JOURNAL OF NEUROSURGERY, 2004, 100 (04) : 787 - 788
  • [37] Patterns of patient movement during frameless image-guided radiosurgery
    Murphy, MJ
    Chang, SD
    Gibbs, IC
    Le, QT
    Hai, J
    Kim, D
    Martin, DP
    Adler, JR
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 55 (05): : 1400 - 1408
  • [38] Measurement of total clinical accuracy for an image guided radiosurgery system
    Wu, X
    Main, W
    Schwade, J
    MEDICAL PHYSICS, 2004, 31 (06) : 1901 - 1901
  • [39] Preliminary Results of Stereotactic Radiotherapy for Spinal Lesions using the Novalis System
    Mori, Yoshimasa
    Hashizume, Chisa
    Shibamoto, Yuta
    Kobayashi, Tatsuya
    Hayashi, Naoki
    Tomita, Natsuo
    Yoshida, Jun
    RADIOSURGERY, VOL 7, 2010, 7 : 378 - 383
  • [40] DETERMINATION OF TOLERANCE TABLE FOR TREATMENT COUCH POSITION USING AN IMAGE GUIDED SYSTEM
    Perez Moreno, J. M.
    Zucca, D.
    Fernandez Leton, P.
    Garcia Ruiz-Zorrilla, J.
    Minambres, A.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S461 - S462