Evaluation of an automated breast 3D-ultrasound system by comparing it with hand-held ultrasound (HHUS) and mammography

被引:35
|
作者
Golatta, Michael [1 ]
Baggs, Christina [1 ]
Schweitzer-Martin, Mirjam [1 ]
Domschke, Christoph [1 ]
Schott, Sarah [1 ]
Harcos, Aba [1 ]
Scharf, Alexander [1 ]
Junkermann, Hans [1 ]
Rauch, Geraldine [2 ]
Rom, Joachim [1 ]
Sohn, Christof [1 ]
Heil, Joerg [1 ]
机构
[1] Univ Breast Unit, D-69120 Heidelberg, Germany
[2] Heidelberg Univ, Inst Med Biometry & Informat, Heidelberg, Germany
关键词
Ultrasound; Breast; Automated three-dimensional breast ultrasound; 3D scanning; BI-RADS((R)); VOLUME SCANNER; LESION DETECTION; DENSE BREASTS; SONOGRAPHY; CANCER; AGREEMENT; DIAGNOSIS; US; CLASSIFICATION; CARCINOMA;
D O I
10.1007/s00404-014-3509-9
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Automated three-dimensional (3D) breast ultrasound (US) systems are meant to overcome the shortcomings of hand-held ultrasound (HHUS). The aim of this study is to analyze and compare clinical performance of an automated 3D-US system by comparing it with HHUS, mammography and the clinical gold standard (defined as the combination of HHUS, mammography and-if indicated-histology). Nine hundred and eighty three patients (=1,966 breasts) were enrolled in this monocentric, explorative and prospective cohort study. All examinations were analyzed blinded to the patientsA ' history and to the results of the routine imaging. The agreement of automated 3D-US with HHUS, mammography and the gold standard was assessed with kappa statistics. Sensitivity, specificity and positive and negative predictive value were calculated to assess the test performance. Blinded to the results of the gold standard the agreement between automated 3D-US and HHUS or mammography was fair, given by a Kappa coefficient of 0.31 (95 % CI [0.26;0.36], p < 0.0001) and 0.25 (95 % CI [0.2;0.3], p < 0.0001), respectively. Our results showed a high negative predictive value (NPV) of 98 %, a high specificity of 85 % and a sensitivity of 74 % based on the cases with US-guided biopsy. Including the cases where the lesion was seen in a second-look automated 3D-US the sensitivity improved to 84 % (NPV = 99 %, specificity = 85 %). The results of this study let us suggest, that automated 3D-US might be a helpful new tool in breast imaging, especially in screening.
引用
收藏
页码:889 / 895
页数:7
相关论文
共 50 条
  • [21] Validation of a hand-held ultrasound device in the evaluation of aortic stenosis
    Zubiaur, Jon
    Castro, Adrian de Margarida
    Perez-Barquin, Raquel
    Gonzalez, Manuel Lozano
    Martin Gorria, Gonzalo
    Ruiz Guerrero, Luis Javier
    Teira Calderon, Andrea
    Santiago Setien, Ignacio
    Serrano Lozano, David
    Gonzalez Vilchez, Francisco
    Vazquez de Prada Tiffe, Jose Antonio
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2025, 41 (02): : 377 - 385
  • [22] Automated Breast Volume Scanner Is More Valuable Than Hand-Held Ultrasound in Diagnosis of Small Breast cancer
    Yan, Lixia
    Jing, Luxia
    Lu, Qing
    Wang, Xi
    Mao, Wujian
    Wang, Peilei
    Zhan, Mengna
    Huang, Beijian
    ULTRASOUND QUARTERLY, 2024, 40 (01) : 66 - 73
  • [23] Diagnostic value of an automated breast volume scanner compared with a hand-held ultrasound: a meta-analysis
    Zhang, Xiaohui
    Chen, Juan
    Zhou, Yidong
    Mao, Feng
    Lin, Yan
    Shen, Songjie
    Sun, Qiang
    Ouyang, Zhaolian
    GLAND SURGERY, 2019, 8 (06) : 698 - +
  • [24] Performance evaluation of a medical robotic 3D-ultrasound imaging system
    Janvier, Marie-Ange
    Durand, Louis-Gilles
    Cardinal, Marie-Helene Roy
    Renaud, Isabelle
    Chayer, Boris
    Bigras, Pascal
    de Guise, Jacques
    Soulez, Gilles
    Cloutier, Guy
    MEDICAL IMAGE ANALYSIS, 2008, 12 (03) : 275 - 290
  • [25] CLINICAL-EVALUATION OF A HAND-HELD ULTRASOUND DEVICE IN PEDIATRIC CARDIOLOGY
    GUSSENHOVEN, WJ
    KAPPEN, L
    LIGTVOET, CM
    EGGINK, JH
    ULTRASOUND IN MEDICINE AND BIOLOGY, 1980, 6 (03): : 271 - &
  • [26] Hand-held ultrasound in the evaluation of lower extremity deep vein thrombosis
    Pellerito, JS
    Hong, RJ
    Farren, L
    Torre, J
    Burke, BJ
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2004, 182 (04) : 2 - 2
  • [27] DIAGNOSTIC PERFORMANCE OF AUTOMATED BREAST ULTRASOUND AS A REPLACEMENT FOR A HAND-HELD SECOND-LOOK ULTRASOUND FOR BREAST LESIONS DETECTED INITIALLY ON MAGNETIC RESONANCE IMAGING
    Chae, Eun Young
    Shin, Hee Jung
    Kim, Hyun Ji
    Yoo, Hyunkyung
    Baek, Seunghee
    Cha, Joo Hee
    Kim, Hak Hee
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2013, 39 (12): : 2246 - 2254
  • [28] A Prospective Comparative Evaluation of Handheld Ultrasound Examination (HHUS) or Automated Ultrasound Examination (ABVS) in Women with Dense Breast
    Brunetti, Nicole
    De Giorgis, Sara
    Tosto, Simona
    Garlaschi, Alessandro
    Rescinito, Giuseppe
    Massa, Barbara
    Calabrese, Massimo
    Tagliafico, Alberto Stefano
    DIAGNOSTICS, 2022, 12 (09)
  • [29] A Hand-Held Ultrasound Imaging System for Point-of-Care Applications
    Han, Ho-san
    Cho, Jeong
    Song, Jae-hee
    Kim, Gi-duck
    Song, Tai-kyong
    Kim, Kang-sik
    2006 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-5, PROCEEDINGS, 2006, : 1099 - +
  • [30] Analysis of 107 breast lesions with automated 3D ultrasound and comparison with mammography and manual ultrasound
    Kotsianos-Hermle, D.
    Hiltawsky, K. M.
    Wirth, S.
    Fischer, T.
    Friese, K.
    Reiser, M.
    EUROPEAN JOURNAL OF RADIOLOGY, 2009, 71 (01) : 109 - 115