Comparison of two point-of-care testing (POCT) devices for fetal lactate during labor

被引:16
|
作者
Heinis, Ayesha M. F. [2 ]
Dinnissen, Jacqueline [1 ]
Spaanderman, Marc E. A. [2 ]
Lotgering, Fred K. [2 ]
Gunnewiek, Jacqueline M. T. Klein [1 ,3 ]
机构
[1] Radboud Univ Nijmegen, Med Ctr, Dept Clin Chem, NL-6525 ED Nijmegen, Netherlands
[2] Radboud Univ Nijmegen, Med Ctr, Dept Obstet & Gynecol, NL-6525 ED Nijmegen, Netherlands
[3] Hosp Gelderse Vallei, Dept Clin Chem & Hematol, Ede, Netherlands
关键词
fetal; lactate; point-of-care testing; scalp; SCALP BLOOD LACTATE; METABOLIC-ACIDOSIS; PH;
D O I
10.1515/CCLM.2011.734
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Point-of-care testing (POCT) of fetal scalp blood lactate is used as an alternative for pH analysis. Lactate measurements have not been standardized and values vary with each device used. The aim of this study was to evaluate the performance of two POCT lactate meters for intrapartum use. Methods: Analytical performance of StatStrip Lactate (Nova Biomedical) and Lactate Pro (Arkray) was evaluated using CLSI EP10. Both POCT meters were compared with our lactate reference method (RapidLab 860; Siemens Healthcare Diagnostics) using fetal scalp and neonatal cord blood. Deming regression analysis was performed. Results: StatStrip Lactate coefficients of variation (CVs) were 5.1%, 5.0% and 2.6% at 0.9, 7.5 and 14.1 mmol/L lactate, respectively. CVs for Lactate Pro were 10.7%, 5.2% and 5.7% at 1.7, 4.1 and 6.4 mmol/L lactate, respectively. Consecutive lactate measurements in 37 fetal scalp and 122 cord blood samples revealed different test characteristics for the two POCT devices. In fetal scalp blood: StatStrip Lactate=1.13*RapidLab-0.39 (R-2=0.907) and Lactate Pro=0.95*RapidLab-0.03 (R-2=0.823). In cord artery blood: StatStrip Lactate=1.08*RapiclLab-0.09 (R-2=0.810) and Lactate Pro=0.72*RapidLab+0.59 (R-2=0.807). Conclusions: Overall performance of both Lactate Pro and StatStrip Lactate was good, with StatStrip Lactate having smallest CVs and closest correlation to our reference method. Both StatStrip Lactate and Lactate Pro can be used as a lactate POCT device for obstetric use.
引用
收藏
页码:89 / 93
页数:5
相关论文
共 50 条
  • [21] Special Topic: Point-of-Care Testing (POCT) and In Vitro Diagnostics (IVDs)
    Li, Chen-zhong
    JOURNAL OF ANALYSIS AND TESTING, 2019, 3 (01) : 1 - 2
  • [22] Special Topic: Point-of-Care Testing (POCT) and In Vitro Diagnostics (IVDs)
    Chen-zhong Li
    Journal of Analysis and Testing, 2019, 3 : 1 - 2
  • [23] Evaluation of two devices for point-of-care testing of haemoglobin in neonatal pigs
    Kutter, Annette P. N.
    Mauch, Jacqueline Y.
    Riond, Barbara
    Martin-Jurado, Olga
    Spielmann, Nelly
    Weiss, Markus
    Bettschart-Wolfensberger, Regula
    LABORATORY ANIMALS, 2012, 46 (01) : 65 - 70
  • [24] Flexible Analytical Devices for Point-of-Care Testing
    Tsutsui, Hideaki
    Lillehoj, Peter B.
    SLAS TECHNOLOGY, 2020, 25 (01): : 6 - 8
  • [25] From Point-of-Care Testing (PoCT) and vaccination to changing the care paradigm: the role of primary care
    Petelos, E.
    Lionis, C.
    EUROPEAN JOURNAL OF PUBLIC HEALTH, 2020, 30
  • [26] Preparation and characterization of a homogeneous immunoassay for point-of-care testing (POCT) of procalcitonin (PCT)
    Li, Zhaoying
    Zhai, Weixiang
    Wang, Lu
    Liu, Jiyang
    Li, Chunjie
    Xu, Liang
    ANALYTICAL METHODS, 2023, 15 (38) : 5002 - 5009
  • [27] Editorial: Integrated Point-of-Care Testing (POCT) Systems: Recent Progress and Applications
    Wang, Lirong
    Xu, Tailin
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2022, 10
  • [28] Glow-type chemiluminescent hydrogels for point-of-care testing (POCT) of cholesterol
    Ning, Yu
    Lu, Fan
    Liu, Yucheng
    Yang, Sisi
    Wang, Fubing
    Ji, Xinghu
    He, Zhike
    ANALYST, 2021, 146 (15) : 4775 - 4780
  • [29] A comparison of two point of care lactate meters for use in fetal blood sampling
    Ridenour, R. V.
    Gada, R. P.
    Bryant, S. C.
    Brost, B. C.
    Karon, B. S.
    CLINICAL CHEMISTRY, 2008, 54 (06) : A231 - A231
  • [30] Practicability of fetal scalp blood sampling during labor using microtubes and a point-of-care (POC) lactate testing device: difficulty assessment, sampling time and failure rates
    Reif, Philipp
    Ulrich, Daniela
    Lakovschek, Ioana
    Tappauf, Carmen
    Lang, Uwe
    Schoell, Wolfgang
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2015, 53 (08) : E195 - E197