Determination of arsenic and cadmium in crude oil by direct sampling graphite furnace atomic absorption spectrometry

被引:26
|
作者
de Jesus, Alexandre [1 ]
Zmozinski, Ariane Vanessa [1 ]
Ferreira Damin, Isabel Cristina [2 ]
Silva, Marcia Messias [1 ,3 ]
Rodrigues Vale, Maria Goreti [1 ,3 ]
机构
[1] Univ Fed Rio Grande do Sul, Inst Quim, BR-91501970 Porto Alegre, RS, Brazil
[2] Fac Dom Bosco Porto Alegre, BR-90520280 Porto Alegre, RS, Brazil
[3] Univ Fed Bahia, Inst Natl Ciencia & Tecnol CNPq, INCT Energia & Ambiente, BR-40170115 Salvador, BA, Brazil
关键词
Arsenic; Cadmium; Direct sampling; Graphite furnace atomic absorption spectrometry; PETROLEUM; VANADIUM; METALS; COPPER; NICKEL; LEAD; STABILIZATION; NAPHTHA; FILTER;
D O I
10.1016/j.sab.2012.03.010
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
In this work, a direct sampling graphite furnace atomic absorption spectrometry method has been developed for the determination of arsenic and cadmium in crude oil samples. The samples were weighed directly on the solid sampling platforms and introduced into the graphite tube for analysis. The chemical modifier used for both analytes was a mixture of 0.1% Pd + 0.06% Mg + 0.06% Triton X-100. Pyrolysis and atomization curves were obtained for both analytes using standards and samples. Calibration curves with aqueous standards could be used for both analytes. The limits of detection obtained were 5.1 mu g kg(-1) for arsenic and 0.2 mu g kg(-1) for cadmium, calculated for the maximum amount of sample that can be analyzed (8 mg and 10 mg) for arsenic and cadmium, respectively. Relative standard deviations lower than 20% were obtained. For validation purposes, a calibration curve was constructed with the SRM 1634c and aqueous standards for arsenic and the results obtained for several crude oil samples were in agreement according to paired t-test. The result obtained for the determination of arsenic in the SRM against aqueous standards was also in agreement with the certificate value. As there is no crude oil or similar reference material available with a certified value for cadmium, a digestion in an open vessel under reflux using a "cold finger" was adopted for validation purposes. The use of paired t-test showed that the results obtained by direct sampling and digestion were in agreement at a 95% confidence level. Recovery tests were carried out with inorganic and organic standards and the results were between 88% and 109%. The proposed method is simple, fast and reliable, being appropriated for routine analysis. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:86 / 91
页数:6
相关论文
共 50 条
  • [31] Determination of As in Vegetable Oil and Biodiesel by Graphite Furnace Atomic Absorption Spectrometry
    Vieira, Mariana Antunes
    Castro de Oliveira, Ligia Claudia
    Goncalves, Rodrigo Araujo
    de Souza, Vanderlea
    de Campos, Reinaldo Calixto
    ENERGY & FUELS, 2009, 23 (12) : 5942 - 5946
  • [32] Determination of lead and arsenic in copper aspirinate by graphite furnace atomic absorption spectrometry
    Qiu, HL
    Liu, JS
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2001, 21 (04) : 555 - 557
  • [33] Solid sampling-graphite furnace atomic absorption spectrometry for the direct determination of boron in plant tissues
    Resano, M.
    Briceno, J.
    Aramendia, M.
    Belarra, M. A.
    ANALYTICA CHIMICA ACTA, 2007, 582 (02) : 214 - 222
  • [34] Direct Sampling Graphite Furnace Atomic Absorption Spectrometry - A Suitable Tool for the Determination of Metallic Contaminants in Pitch
    Frohlich, Angelica C.
    Pereira, Leticia S. F.
    Junges, Andrizia F.
    Flores, Erico M. M.
    Paniz, Jose N. G.
    Duarte, Fabio A.
    BULLETIN OF THE CHEMICAL SOCIETY OF JAPAN, 2021, 94 (07) : 1963 - 1969
  • [35] Determination of trace impurities in aluminum nitride by direct solid sampling graphite furnace atomic absorption spectrometry
    Paz de Mattos, Julio Cezar
    Rodrigues, Luiz Frederico
    de Moraes Flores, Erico Marlon
    Krivan, Viliam
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2011, 66 (08) : 637 - 643
  • [36] Determination of cadmium in rice by microwave digestion and graphite furnace atomic absorption spectrometry
    Amakawa, E
    Ogiwara, T
    Ohhashi, N
    Kamata, K
    Suzuki, S
    JOURNAL OF THE FOOD HYGIENIC SOCIETY OF JAPAN, 2000, 41 (01): : 66 - 69
  • [37] Direct determination of lead and cadmium in soil by slurry-sampling graphite furnace atomic absorption spectrometry using matrix modification technique
    Sun, HW
    Wen, XH
    Liang, SX
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2006, 26 (05) : 950 - 954
  • [38] Determination of manganese and nickel in slurry sampling by graphite furnace atomic absorption spectrometry
    Pereira, Luciano Almeida
    Borges, Simone Soares de O.
    Castro, Mauricio Costa
    Borges Neto, Waldomiro
    Windmoller, Claudia Carvalhinho
    Borba da Silva, Jose Bento
    CANADIAN JOURNAL OF CHEMISTRY, 2008, 86 (04) : 312 - 316
  • [39] Determination of cadmium in lichens by solid sampling graphite furnace atomic absorption spectrometry (SS-GF-AAS)
    Pavel Coufalík
    Antonín Uher
    Ondřej Zvěřina
    Josef Komárek
    Environmental Monitoring and Assessment, 2020, 192
  • [40] Determination of cadmium in lichens by solid sampling graphite furnace atomic absorption spectrometry (SS-GF-AAS)
    Coufalik, Pavel
    Uher, Antonin
    Zverina, Ondrej
    Komarek, Josef
    ENVIRONMENTAL MONITORING AND ASSESSMENT, 2020, 192 (04)