Gas chromatographic determination of ethyl glucuronide in hair: Comparison between tandem mass spectrometry and single quadrupole mass spectrometry

被引:25
|
作者
Cappelle, Delphine [1 ]
Neels, Hugo [1 ,2 ]
Yegles, Michel [3 ]
Paulus, Jeff [3 ]
van Nuijs, Alexander L. N. [1 ]
Covaci, Adrian [1 ]
Crunelle, Cleo L. [1 ]
机构
[1] Toxicol Univ Antwerp, B-2610 Antwerp, Belgium
[2] ZNA Stuivenberg Hosp, Toxicol Lab, B-2060 Antwerp, Belgium
[3] Lab Natl Sante, Serv Toxicol, L-3555 Dudelange, Luxembourg
关键词
Ethyl glucuronide; Alcohol; Hair; GC-MS/MS; Negative ion chemical ionization; Validation; ETG;
D O I
10.1016/j.forsciint.2014.11.022
中图分类号
DF [法律]; D9 [法律]; R [医药、卫生];
学科分类号
0301 ; 10 ;
摘要
Ethyl glucuronide (EtG), a minor metabolite of ethanol, accumulates in hair and is currently used as a long-term marker for the detection of chronic and excessive alcohol consumption. Sensitive methods are required to differentiate teetotalers from moderate drinkers according to the established cut-off (i.e., 7 pg/mg hair). The aim of this study was to develop a sensitive method using gas chromatography coupled to tandem mass spectrometry (GC-MS/MS) operated in the negative ion chemical ionization (NICI) mode. The validated method was applied to hair samples from teetotalers, moderate and excessive alcohol consumers, and results were compared to a previously validated GC-NICI-MS method. The developed GC-NICI-MS/MS method showed linearity over a range from 2 to 400 pg/mg hair, with a limit of detection (LOD) of 0.05 pg/mg hair and a lower limit of quantification (LLOQ) of 0.2 pg/mg hair, compared to an LOD of 0.5 pg/mg hair and LLOQ of 1.5 pg/mg hair obtained with GC-NICI-MS. Furthermore, lower background noise was observed using GC-NICI-MS/MS. Comparison of results of hair samples (n = 58) obtained by GC-NICI-MS and GC-NICI-MS/MS showed no significant difference between both methods (paired-sample t-test, p > 0.05; mean CV = 1.0%). The differences between both methods were larger for EtG concentrations < 30 mg/pg hair (mean CV = 1.7%) than for EtG concentrations > 30 mg/pg hair (mean CV = 0.7%). This suggests a higher selectivity of GC-NICI-MS/MS at lower concentrations. In conclusion, by using GC-NICI-MS/MS, a higher analytical selectivity and an improved signal to noise ratio, can be achieved. Although GC-NICI-MS would not change the interpretation of the EtG concentrations, the present GC-NICI-MS/MS method should preferentially be used for the determination of EtG in hair, especially when differentiating between teetotalers and moderate drinkers according to the current cut-off (i.e., 7 pg/mg hair). (C) 2014 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:20 / 24
页数:5
相关论文
共 50 条
  • [41] TANDEM QUADRUPOLE-TIME-OF-FLIGHT INSTRUMENT FOR MASS-SPECTROMETRY MASS-SPECTROMETRY
    GLISH, GL
    GOERINGER, DE
    ANALYTICAL CHEMISTRY, 1984, 56 (13) : 2291 - 2295
  • [42] Determination of Acrylamide in Bread by Gas Chromatography - Tandem Mass Spectrometry
    Negoita, Mioara
    Catana, Monica
    Iorga, Enuta
    Catana, Luminita
    Adascalului, Alina
    Belc, Nastasia
    ROMANIAN BIOTECHNOLOGICAL LETTERS, 2014, 19 (04): : 9561 - 9568
  • [43] Haloxyfop determination by gas chromatography/tandem mass spectrometry in eggs
    Lehner, Andreas F.
    Zyskowski, Justin
    Fulton, Richard M.
    Buchweitz, John P.
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2020, 34 (19)
  • [44] Comparison of Liquid and Gas Chromatographic Mass Spectrometry for the Determination of Multiclass Pesticides at Low Concentrations
    Anagnostopoulos, C. J.
    Balagiannis, G.
    Miliadis, G. E.
    SPECTROSCOPY LETTERS, 2012, 45 (03) : 202 - 218
  • [45] Determination of polychlorinated biphenyls in ambient air by gas chromatography coupled to triple quadrupole tandem mass spectrometry
    Rosa Martínez Ocaña
    Angela Mena Granero
    Francisco Javier Egea Gonzalez
    Antonia Garrido Frenich
    José Luis Martínez Vidal
    Patricia Plaza Bolaños
    Analytical and Bioanalytical Chemistry, 2008, 390 : 1413 - 1423
  • [46] Evaluation of gas chromatography-tandem quadrupole mass spectrometry for the determination of organochlorine pesticides in fats and oils
    Patel, K
    Fussell, RJ
    Hetmanski, M
    Goodall, DM
    Keely, BJ
    JOURNAL OF CHROMATOGRAPHY A, 2005, 1068 (02) : 289 - 296
  • [47] Assessment of PCBs exposure in human hair using double focusing high resolution mass spectrometry and single quadrupole mass spectrometry
    Barbounis, Emmanouil G.
    Tzatzarakis, Manolis N.
    Alegakis, Athanasios K.
    Kokkinaki, Aikaterini
    Karamanos, Nikos
    Tsakalof, Andreas
    Tsatsakis, Aristidis M.
    TOXICOLOGY LETTERS, 2012, 210 (02) : 225 - 231
  • [48] Determination of polyhalogenated carbazoles in soil using gas chromatography-triple quadrupole tandem mass spectrometry
    Tao, Wuqun
    Zhou, Zhiguang
    Shen, Li
    Zhu, Chaofei
    Zhang, Wanglong
    Xu, Li
    Guo, Zhingling
    Xu, Tong
    Xie, Heidi Qunhui
    Zhao, Bin
    SCIENCE OF THE TOTAL ENVIRONMENT, 2020, 710 (710)
  • [49] Determination of polychlorinated biphenyls in ambient air by gas chromatography coupled to triple quadrupole tandem mass spectrometry
    Ocana, Rosa Martinez
    Granero, Angela Mena
    Gonzalez, Francisco Javier Egea
    Frenich, Antonia Garrido
    Vidal, Jose Luis Martinez
    Bolanos, Patricia Plaza
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2008, 390 (05) : 1413 - 1423
  • [50] Determination of ethyl glucuronide in human urine by solid phase extraction-gas chromatography-mass spectrometry
    Yu Tianxiao
    Li Qing
    Wan Tao
    Li Jianbo
    Ding Shijia
    CHINESE JOURNAL OF CHROMATOGRAPHY, 2011, 29 (02) : 172 - 175