The Swiss IMRT dosimetry intercomparison using a thorax phantom

被引:26
|
作者
Schiefer, H. [1 ]
Fogliata, A. [2 ]
Nicolini, G. [2 ]
Cozzi, L. [2 ]
Seelentag, W. W. [1 ]
Born, E. [3 ,4 ]
Hasenbalg, F. [3 ,4 ]
Roth, J. [5 ]
Schnekenburger, B. [5 ]
Muench-Berndl, K. [6 ]
Vallet, V. [7 ]
Pachoud, M. [7 ]
Reiner, B. [8 ]
Dipasquale, G. [9 ]
Krusche, B. [10 ]
Fix, M. K. [3 ,4 ]
机构
[1] Kantonsspital St Gallen, Klin Radioonkol, CH-9007 St Gallen, Switzerland
[2] Oncol Inst So Switzerland, Med Phys Unit, CH-6500 Bellinzona, Switzerland
[3] Univ Hosp Bern, Inselspital, Div Med Radiat Phys, CH-3010 Bern, Switzerland
[4] Univ Bern, CH-3010 Bern, Switzerland
[5] Univ Spital Basel, Inst Radioonkol, CH-4031 Basel, Switzerland
[6] Lindenhofspital Bern, Klin Radioonkol, CH-3001 Bern, Switzerland
[7] Univ Inst Appl Radiophys, CH-1007 Lausanne, Switzerland
[8] Univ Zurich Hosp, CH-8091 Zurich, Switzerland
[9] Hop Univ Geneve, Serv Radiooncol, CH-1211 Geneva, Switzerland
[10] Univ Basel, Inst Phys, CH-4056 Basel, Switzerland
关键词
dose comparison; radiotherapy; intensity modulated; heterogeneity; thermoluminescence dosimetry; DOSE CALCULATION ALGORITHMS; MODULATED RADIATION-THERAPY; MONTE-CARLO; EQUIVALENT HETEROGENEITIES; QUALITY-ASSURANCE; CLINICAL-TRIALS; RADIOTHERAPY; CANCER; LUNG; IMPLEMENTATION;
D O I
10.1118/1.3460795
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: In 2008, a national intensity modulated radiation therapy (IMRT) dosimetry intercomparison was carried out for all 23 radiation oncology institutions in Switzerland. It was the aim to check the treatment chain focused on the planning, dose calculation, and irradiation process. Methods: A thorax phantom with inhomogeneities was used, in which thermoluminescence dosimeter (TLD) and ionization chamber measurements were performed. Additionally, absolute dosimetry of the applied beams has been checked. Altogether, 30 plan-measurement combinations have been used in the comparison study. The results have been grouped according to dose calculation algorithms, classified as "type a" or "type b," as proposed by Knoos et al. ["Comparison of dose calculation algorithms for treatment planning in external photon beam therapy for clinical situations," Phys. Med. Biol. 51, 5785-5807 (2006)]. Results: Absolute dosimetry check under standard conditions: The mean ratio between the dose derived from the single field measurement and the stated dose, calculated with the treatment planning system, was 1.007 +/- 0.010 for the ionization chamber and 1.002 +/- 0.014 (mean +/- standard deviation) for the TLD measurements. IMRT Plan Check: In the lung tissue of the planning target volume, a significantly better agreement between measurements (TLD, ionization chamber) and calculations is shown for type b algorithms than for type a (p < 0.001). In regions outside the lungs, the absolute differences between TLD measured and stated dose values, relative to the prescribed dose, vertical bar(D(m) - D(s)) / D(prescribed)vertical bar, are 1.9 +/- 0.4% and 1.4 +/- 0.3%, respectively. These data show the same degree of accuracy between the two algorithm types if low-density medium is not present. Conclusions: The results demonstrate that the performed intercomparison is feasible and confirm the calculation accuracies of type a and type b algorithms in a water equivalent and low-density environment. It is now planned to offer the intercomparison on a regular basis to all Swiss institutions using IMRT techniques. (C) 2010 American Association of Physicists in Medicine. [DOT: 10.1118/1.3460795]
引用
收藏
页码:4424 / 4431
页数:8
相关论文
共 50 条
  • [1] A DOSIMETRY INTERCOMPARISON PHANTOM FOR INTRAOPERATIVE RADIOTHERAPY
    Armoogum, K.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S396 - S396
  • [2] A dosimetry intercomparison phantom for intraoperative radiotherapy
    Armoogum, Kris
    Watson, Colin
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2008, 18 (02): : 120 - 127
  • [3] Hctergeneity dose calculation accuracy in IMRT using an anthropomorphic thorax phantom
    Davidson, S.
    Prado, K.
    Ibbott, G.
    Followill, D.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2106 - 2106
  • [4] A comprehensive study on the heterogeneity dose calculation accuracy in IMRT using an anthropomorphic thorax phantom
    Davidson, S.
    Popple, R.
    Ibbott, G.
    Followill, D.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2520 - 2520
  • [5] END-TO-END IMRT EVALUATION USING SEMI-ANTHROPOMORPHIC THORAX PHANTOM
    Udrea, Andreea A.
    Dumitrache, Mihai T.
    Dumitrache, Alina
    Vlasceanu, Maria
    Popescu, Aurel I.
    [J]. ROMANIAN REPORTS IN PHYSICS, 2022, 74 (01)
  • [6] Implementation of IMRT and VMAT using Delta4 phantom and Portal dosimetry as dosimetry verification tools
    Daci, Lulzime
    Malkaj, Partizan
    [J]. 9TH INTERNATIONAL PHYSICS CONFERENCE OF THE BALKAN PHYSICAL UNION (BPU-9), 2016, 1722
  • [7] How good is film dosimetry for phantom verification in IMRT?
    Dunia, R
    Sanford, R
    Chung, H
    Al-Ghazi, M
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 1268 - 1268
  • [8] Dosimetry in an IMRT phantom designed for a remote monitoring program
    Han, Youngyih
    Shin, Eun Hyuk
    Lim, Chunil
    Kang, Se-Kwon
    Park, Sung Ho
    Lah, Jeong-Eun
    Suh, Tae-suk
    Yoon, Myonggeun
    Lee, Se Byeong
    Cho, Sang Hyun
    Ibbott, Geoffrey S.
    Ju, Sang Gyu
    Ahn, Yong Chan
    [J]. MEDICAL PHYSICS, 2008, 35 (06) : 2519 - 2527
  • [9] Dosimetry in an IMRT phantom designed for a remote auditing program
    Han, Y.
    Shin, E. H.
    Yoon, M. K.
    Lee, S. B.
    Ju, S. G.
    Lim, C. I.
    Park, S. H.
    Kang, S. K.
    Lah, J. E.
    Cho, S. H.
    Ibbott, G.
    Cahn, Y.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2444 - 2444
  • [10] The Czech multicenter dosimetric and treatment planning intercomparison of IMRT with a pelvic phantom
    Koniarova, I.
    Horakova, I.
    Dufek, V.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S278 - S278