Experimental and clinical evaluation of iterative reconstruction (OSEM) in dynamic PET: Quantitative characteristics and effects on kinetic modeling

被引:0
|
作者
Boellaard, R [1 ]
van Lingen, A [1 ]
Lammertsma, AA [1 ]
机构
[1] Univ Hosp Vrije Univ, Clin PET Ctr, NL-1007 MB Amsterdam, Netherlands
关键词
iterative reconstruction; dynamic PET; quantitative accuracy; kinetic modeling;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to investigate the quantitative properties and effects of ordered-subset expectation maximization (OSEM) on kinetic modeling compared with filtered backprojection (FBP) in dynamic PET studies. Both phantom and patient studies were performed. Methods: For phantom studies dynamic two-dimensional emission scans with 10-min frames and 20-min scan intervals were acquired over a 14-h period using an HR+ PET scanner. Various phantoms were scanned: 2-, 5-, 10-, and 20-cm-diameter phantoms filled with an F-18 solution (300 kBq/ml) and a NEMA phantom filled with an F-18 background (40 kBq/ml) and a cold or C-11 insert (450 kBq/ml). Transmission (Tx) scans of 5-60 min were acquired. Data were reconstructed using FBP Hanning 0.5 and OSEM with 2-12 iterations and 12 or 24 subsets. Quantitative accuracy and noise characteristics were assessed. For patient studies, five cardiac, three oncologic, and three brain dynamic F-18-FDG scans were used. Five reconstructions were performed: FBP Hanning 0.5, and OSEM 2 x 12 and OSEM 4 x 16 with and without 5-mm full width at half maximum smoothing. Time-activity curves were calculated using volumes of interest. The input function was derived from arterial sampling. Metabolic rate of glucose (MRglu) was calculated with a standard two-tissue compartment model and Patlak analysis. Results: Contribution of Tx noise to the reconstructed image was smaller for OSEM than for FBP. Differences in signal-to-noise ratio between FBP and OSEM depended on number of iterations and phantom size, Bias with OSEM was observed for regions enclosed within a 5- to 10-fold hotter background. For cardiac studies OSEM 2 x 12 and OSEM 4 x 16 resulted in 13% and 21% higher pixel values and 9% and 15% higher MRglu values compared with FBP. Smoothing decreased all these values to 2%, Similar results were found for most tumor studies. For brain studies MRglu of FBP and OSEM 4 x 16 agreed within 2%. Use of OSEM image-derived input functions for cardiac PET studies resulted in a decrease in calculated MRglu of about 15%. Conclusion: For most PET studies OSEM has equal quantitative accuracy as FBP. The higher pixel and MRglu values are explained by the better resolution of OSEM. However, OSEM does not provide accurate image-derived input functions for FDG cardiac PET studies because of bias in regions located within a hotter background.
引用
收藏
页码:808 / 817
页数:10
相关论文
共 50 条
  • [21] A quantitative clinical evaluation of simultaneous reconstruction of attenuation and activity in time-of-flight PET
    Haiqiong Zhang
    Jingnan Wang
    Nan Li
    Yue Zhang
    Jie Cui
    Li Huo
    Hui Zhang
    BMC Medical Imaging, 23
  • [22] A quantitative clinical evaluation of simultaneous reconstruction of attenuation and activity in time-of-flight PET
    Zhang, Haiqiong
    Wang, Jingnan
    Li, Nan
    Zhang, Yue
    Cui, Jie
    Huo, Li
    Zhang, Hui
    BMC MEDICAL IMAGING, 2023, 23 (01)
  • [23] Machine learning based quantitative analysis: an alternative method for traditional kinetic modeling based quantitative analysis of dynamic PET data
    Pan, Leyun
    Zhang, Xun
    Koralli, Panagiota
    Cheng, Caixia
    Wang, Eric Ke
    Ye, Yunming
    Dimitrakopoulou-Strauss, Antonia
    JOURNAL OF NUCLEAR MEDICINE, 2019, 60
  • [24] Performance characteristics and quantitative clinical evaluation of the digital Siemens Biograph Vision PET/CT
    Van Sluis, Joyce
    Boellaard, Ronald
    de Jong, Johan
    Willemsen, Antoon
    Somasundaram, Ananthi
    van Snick, Paul
    Borra, Ronald
    Dierckx, Rudi
    Stormezand, Gilles
    Glaudemans, Andor
    Noordzij, Walter
    JOURNAL OF NUCLEAR MEDICINE, 2019, 60
  • [25] Machine learning-based kinetic modeling: a robust and reproducible solution for quantitative analysis of dynamic PET data
    Pan, Leyun
    Cheng, Caixia
    Haberkorn, Uwe
    Dimitrakopoulou-Strauss, Antonia
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (09): : 3566 - 3581
  • [26] Evaluation of Quantitative Ga-68 PSMA PET/CT Repeatability of Recurrent Prostate Cancer Lesions Using Both OSEM and Bayesian Penalized Likelihood Reconstruction Algorithms
    Roef, Mark J.
    Rijnsdorp, Sjoerd
    Brouwer, Christel
    Wyndaele, Dirk N.
    Arends, Albert J.
    DIAGNOSTICS, 2021, 11 (06)
  • [27] Evaluation of quantitative modeling methods in whole-body, dynamic [11C]-erlotinib PET
    Petrulli, Joseph Ryan
    Zheng, Mingqiang
    Huang, Yiyun
    Nabulsi, Nabeel B.
    Goldberg, Sarah B.
    Contessa, Joseph N.
    Morris, Evan D.
    AMERICAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2021, 11 (02): : 143 - +
  • [28] Bias evaluation and reduction in 3D OP-OSEM reconstruction in dynamic equilibrium PET studies with 11C-labeled for binding potential analysis
    Brambilla, Claudia Regio
    Scheins, Jurgen
    Issa, Ahlam
    Tellmann, Lutz
    Herzog, Hans
    Kops, Elena Rota
    Shah, N. Jon
    Neuner, Irene
    Lerche, Christoph W.
    PLOS ONE, 2021, 16 (01):
  • [29] Experimental evaluation of iterative reconstruction for whole-body F-18 PET in a 3-and 4-ring PET/CT system
    Ressner, M.
    Gustafsson, D.
    Gustafsson, A.
    Jonsson, C.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2011, 38 : S263 - S264
  • [30] Quantitative comparison of analytic and iterative reconstruction methods in 2-and 3-dimensional dynamic cardiac 18F-FDG PET
    Lubberink, M
    Boellaard, R
    van der Weerdt, AP
    Visser, FC
    Lammertsma, AA
    JOURNAL OF NUCLEAR MEDICINE, 2004, 45 (12) : 2008 - 2015