Microdosimetric specification of the radiation quality of a d(48.5)+Be fast neutron therapy beam produced by a superconducting cyclotron

被引:9
|
作者
Kota, C [1 ]
Maughan, RL [1 ]
机构
[1] WAYNE STATE UNIV,DETROIT,MI 48201
关键词
radiotherapy; neutrons; microdosimetry; radiation quality;
D O I
10.1118/1.597896
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The proportional counter microdosimetric technique has been employed to quantify variations in the quality of a d(48.5)+Be fast neutron beam passing through a homogeneous water phantom. Single event spectra have been measured as a function of spatial location in the water phantom and field size. The measured spectra have been separated into component spectra corresponding to the gamma, recoil proton and alpha plus heavy recoil ion contribution to the total absorbed dose. The total absorbed dose normalized to the ''monitor units'' used in daily clinical use has been calculated from the measured spectra and compared to the data measured with calibrated ion chambers. The present measurements agree with the ion chamber data to within 5%. The RBE of the neutron beam is assumed to be proportional to the microdosimetric parameter gamma* for the dose ranges pertinent to fractionated neutron therapy. The relative variations in gamma*, assumed to be representative of variations in the RBE are mapped as a function of field size and spatial location in the phantom. A variation in the RBE of about 4% for points within and 8% for points outside a 10 cmX10 cm field is observed. The variations in the RBE within the beam are caused by an increase in the gamma component with depth. An increase in the RBE of about 4% is observed with increasing field size which is attributed to a change in the neutron spectrum. Compared to the uncertainties in the prescribed dose, associated with uncertainties in the clinically used RBE, variation in the RBE between various tissues, and other dosimetric uncertainities caused by factors such as patient inhomogeneities, patient setup errors, patient motion, etc., the measured spatial RBE variations are not considered significant enough to be incorporated into the treatment planning scheme. (C) 1996 American Association of Physicists in Medicine.
引用
收藏
页码:1591 / 1599
页数:9
相关论文
共 38 条
  • [31] EARLY CLINICAL-EVALUATION OF NEUTRON BEAM RADIATION-THERAPY WITH THE D-D REACTION
    GRIEM, ML
    KUCHNIR, FT
    REFT, CS
    SUTTON, HG
    AMERICAN JOURNAL OF CLINICAL ONCOLOGY-CANCER CLINICAL TRIALS, 1983, 6 (02): : 142 - 142
  • [32] MICRODOSIMETRIC INVESTIGATIONS ON COLLIMATED FAST-NEUTRON BEAMS FOR RADIATION-THERAPY .1. MEASUREMENTS OF MICRODOSIMETRIC SPECTRA AND PARTICLE DOSE FRACTIONS IN A WATER PHANTOM FOR FAST-NEUTRONS FROM 14-MEV DEUTERONS ON BERYLLIUM
    FIDORRA, J
    BOOZ, J
    PHYSICS IN MEDICINE AND BIOLOGY, 1981, 26 (01): : 27 - 41
  • [33] RADIATION QUALITY INVESTIGATIONS AROUND A NEUTRON THERAPY BEAM, USING A HIGH-PRESSURE IONIZATION-CHAMBER
    MAKRIGIORGOS, GM
    SMATHERS, JB
    MYERS, LT
    BENEDICT, SH
    MEDICAL PHYSICS, 1990, 17 (01) : 84 - 86
  • [34] Novel Technologies and Theoretical Models in Radiation Therapy of Cancer Patients Using 6.3 MeV Fast Neutrons Produced by U-120 Cyclotron
    Musabaeva, L. I.
    Startseva, Zh. A.
    Gribova, O. V.
    Velikaya, V. V.
    Lisin, V. A.
    PHYSICS OF CANCER: INTERDISCIPLINARY PROBLEMS AND CLINICAL APPLICATIONS (PC'16), 2016, 1760
  • [35] LOW-PRESSURE TISSUE EQUIVALENT PROPORTIONAL-COUNTERS APPLIED TO THE PROBLEM OF RADIATION QUALITY IN FAST-NEUTRON THERAPY
    MENZEL, HG
    SCHUHMACHER, H
    STRAHLENTHERAPIE, 1984, 160 (02) : 125 - 126
  • [36] Spatial variation of radiation quality during moving beam therapy with 14 MeV [d(0.25)+T] neutrons
    Herskind, C
    Loncol, T
    Höver, KH
    RADIATION PROTECTION DOSIMETRY, 2002, 99 (1-4) : 365 - 368
  • [37] PRELIMINARY-REPORT OF THE M-D-ANDERSON-HOSPITAL TEXAS-A AND M VARIABLE ENERGY CYCLOTRON FAST-NEUTRON THERAPY PILOT-STUDY
    PETERS, LJ
    HUSSEY, DH
    FLETCHER, GH
    BAUMANN, PA
    OLSON, MH
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1979, 132 (04) : 637 - 642
  • [38] Patient-reported Quality of Life in Prostate Cancer Patients Treated With 3D Conformal Intensity Modulated or Proton Beam Radiation Therapy
    Gray, P. J.
    Paly, J. J.
    Yeap, B.
    Sanda, M. G.
    Talcott, J. A.
    Sandler, H. M.
    Michalski, J. M.
    Hamstra, D. A.
    Bekelman, J. E.
    Efstathiou, J. A.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 84 (03): : S13 - S13