A fully 3D iterative image reconstruction algorithm incorporatin data corrections

被引:0
|
作者
Iatrou, M [1 ]
Ross, SG [1 ]
Manjeshwar, RM [1 ]
Stearns, CW [1 ]
机构
[1] Gen Elect Global Res Ctr, Niskayuna, NY 12309 USA
关键词
D O I
暂无
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
In this study, we implemented a fully 3D Maximum Likelihood Ordered Subsets Expectation Maximization (ML-OSEM) reconstruction algorithm with two methods for corrections of randoms, and scatter coincidences: (a) Measured data were pre-corrected for randoms and scatter, and (b) Corrections were incorporated into the iterative algorithm. In 3D PET acquisitions, the random and scatter coincidences constitute a significant fraction of the measured coincidences. ML-OSEM reconstruction algorithms make assumptions of Poisson distributed data. Pre-corrections for random and scatter coincidences result in deviations from that assumption, potentially leading to increased noise and inconsistent convergence. Incorporating the corrections inside the loop of the iterative reconstruction preserves the Poisson nature of the data. We performed Monte Carlo simulations with different randoms fractions and reconstructed the data with the two methods. We also reconstructed clinical patient images. The two methods were compared quantitatively through contrast and noise measurements. The results indicate that for high levels of randoms, incorporating the corrections inside the iterative loop results in superior image quality.
引用
收藏
页码:2493 / 2497
页数:5
相关论文
共 50 条
  • [1] Parallelization of a fully 3D CT iterative reconstruction algorithm
    Keesing, Daniel B.
    O'Sullivan, Joseph A.
    Politte, David G.
    Whiting, Bruce R.
    [J]. 2006 3RD IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: MACRO TO NANO, VOLS 1-3, 2006, : 1240 - +
  • [2] Fast iterative image reconstruction of 3D PET data
    Kinahan, PE
    Michel, C
    Defrise, M
    Townsend, DW
    Sibomana, M
    Lonneux, M
    Newport, DF
    Luketich, JD
    [J]. 1996 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1997, : 1918 - 1922
  • [3] Assessment of image quality with a fast fully 3D reconstruction algorithm
    Daube-Witherspoon, ME
    Matej, S
    Karp, JS
    [J]. 2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 2238 - 2242
  • [4] ACCELERATED FULLY 3D ITERATIVE RECONSTRUCTION IN SPECT
    Backfrieder, Werner
    Zwettler, Gerald A.
    [J]. 23RD EUROPEAN MODELING & SIMULATION SYMPOSIUM, EMSS 2011, 2011, : 100 - 104
  • [5] Fast iterative image reconstruction methods for fully 3D multispectral bioluminescence tomography
    Ahn, Sangtae
    Chaudhari, Abhijit J.
    Darvas, Felix
    Bouman, Charles A.
    Leahy, Richard M.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (14): : 3921 - 3942
  • [6] Missing data estimation for fully 3D spiral CT image reconstruction
    Keesing, Daniel B.
    O'Sullivan, Joseph A.
    Politte, David G.
    Whiting, Bruce R.
    Snyder, Donald L.
    [J]. MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2007, 6510
  • [7] Optimization of PET image quality by means of 3D data acquisition and iterative image reconstruction
    Doll, J
    Zaers, J
    Trojan, H
    Bellemann, ME
    Adam, LE
    Haberkorn, U
    Brix, G
    [J]. NUKLEARMEDIZIN-NUCLEAR MEDICINE, 1998, 37 (02): : 62 - 67
  • [8] Fast Segment Iterative Algorithm for 3D Reconstruction
    Mesko, Matej
    Krsak, Emil
    [J]. 2014 10TH INTERNATIONAL CONFERENCE ON DIGITAL TECHNOLOGIES (DT), 2014, : 238 - 242
  • [9] Fully 3D implementation of the end-to-end deep image prior-based PET image reconstruction using block iterative algorithm
    Hashimoto, Fumio
    Onishi, Yuya
    Ote, Kibo
    Tashima, Hideaki
    Yamaya, Taiga
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (15):
  • [10] Fully 3D iterative CT reconstruction using polar coordinates
    Thibaudeau, Christian
    Leroux, Jean-Daniel
    Fontaine, Rejean
    Lecomte, Roger
    [J]. MEDICAL PHYSICS, 2013, 40 (11)