Theoretical and experimental comparison of image signal and noise for dual-energy subtraction angiography and conventional x-ray angiography

被引:3
|
作者
Burton, Christiane S. [1 ,2 ]
Mayo, John R. [3 ,4 ]
Cunningham, Ian A. [1 ,2 ]
机构
[1] Univ Western Ontario, Imaging Res Labs, Robarts Res Inst, London, ON N6A 5B7, Canada
[2] Univ Western Ontario, Dept Med Biophys, London, ON N6A 5B7, Canada
[3] Vancouver Gen Hosp, Radiol, Vancouver, BC, Canada
[4] Univ British Columbia, Vancouver, BC V5Z 1M9, Canada
关键词
OPTIMIZATION; SYSTEMS;
D O I
10.1117/12.2082724
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Purpose: Cardiovascular disease is currently the leading cause of mortality worldwide. Digital subtraction angiography (DSA) is widely used to enhance the visibility of small vessels and vasculature obscurred by overlying bone and lung fields by subtracting a mask and contrast image. However, motion between these mask and contrast images can introduce artifacts that can render a study non-diagnostic. This makes DSA particularly unsuccessful for cardiac imaging. Energy subtraction angiography (ESA) or dual-energy was proposed in the past as an alternative for vascular imaging, however it was abandonned due to image quality that is inferior to DSA. Image quality for angiography comes down to iodine signal and noise. In this paper we investigate the fundamental iodine signal and noise analysis of ESA and compare it to DSA. Method: We developed a polyenergetic and monoenergetic theoretical model for iodine signal and noise for both ESA and DSA. We validated our polyenergetic model by experiment where ESA and DSA images of a vascular phantom were acquired using an x-ray system with a flat panel CsI Xmaru1215CF-MPTM (Rayence Co., Ltd., Republic of Korea) detector. For ESA low and high applied tube voltages of 50 kV and 120 kV (2.5 mmCu), respectively, and for DSA the applied tube voltage was 80 kV. Iodine signal-to-noise ratio (SNR) per entrance exposure was calculated for each iodine concentration for both ESA and DSA. Results: Our measured iodine SNR agreed well with theoretical calculations. Iodine SNR for ESA was relatively higher than DSA for low iodine mass loadings, and as iodine mass loading increases iodine SNR decreases. Conclusions: We have developed a model for iodine SNR for both DSA and ESA. Our model was validated with experiment and showed excellent agreement. We have shown that there is potential for obtaining iodine-specific images using ESA that are similar to DSA.
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Novel x-ray source for dual-energy subtraction angiography
    Tuffanelli, A
    Taibi, A
    Baldazzi, G
    Bollini, D
    Gombia, M
    Ramello, L
    Gambaccini, M
    [J]. MEDICAL IMAGING 2002: PHYSICS OF MEDICAL IMAGING, 2002, 4682 : 311 - 319
  • [2] Signal-difference-to-noise comparison of temporal subtraction, kV-switching dual-energy and photon-counting dual-energy x-ray angiography
    Aubert, Sarah
    Tanguay, Jesse
    [J]. MEDICAL PHYSICS, 2023, 50 (12) : 7400 - 7414
  • [3] Dual-energy direct bone removal CT angiography for evaluation of intracranial aneurysm or stenosis: comparison with conventional digital subtraction angiography
    Watanabe, Yoshiyuki
    Uotani, Kensuke
    Nakazawa, Tetsuro
    Higashi, Masahiro
    Yamada, Naoaki
    Hori, Yoshiro
    Kanzaki, Suzu
    Fukuda, Tetsuya
    Itoh, Toshihide
    Naito, Hiroaki
    [J]. EUROPEAN RADIOLOGY, 2009, 19 (04) : 1019 - 1024
  • [4] Experimental and theoretical studies of dual energy subtraction angiography (DESA) performed using laser-based x-ray source
    Krol, A
    Kieffer, JC
    Ichalaleb, Z
    Jiang, ZM
    Chamberlain, CC
    Scalzetti, E
    [J]. APPLICATIONS OF X RAYS GENERATED FROM LASERS AND OTHER BRIGHT SOURCES II, 2001, 4504 : 227 - 233
  • [5] Dual-energy direct bone removal CT angiography for evaluation of intracranial aneurysm or stenosis: comparison with conventional digital subtraction angiography
    Yoshiyuki Watanabe
    Kensuke Uotani
    Tetsuro Nakazawa
    Masahiro Higashi
    Naoaki Yamada
    Yoshiro Hori
    Suzu Kanzaki
    Tetsuya Fukuda
    Toshihide Itoh
    Hiroaki Naito
    [J]. European Radiology, 2009, 19 : 1019 - 1024
  • [6] Next-Generation Angiography: A Theoretical Comparison of Signal and Noise in Photon-Counting and Dual-Energy Approaches
    Tanguay, J.
    Cunningham, I.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [7] A CORRELATED NOISE-REDUCTION ALGORITHM FOR DUAL-ENERGY DIGITAL SUBTRACTION ANGIOGRAPHY
    MCCOLLOUGH, CH
    VANLYSEL, MS
    PEPPLER, WW
    MISTRETTA, CA
    [J]. MEDICAL PHYSICS, 1989, 16 (06) : 873 - 880
  • [8] Clinical evaluation of dual-energy bone removal in CT angiography of the head and neck: comparison with conventional bone-subtraction CT angiography
    Deng, K.
    Liu, C.
    Ma, R.
    Sun, C.
    Wang, X. -m.
    Ma, Z. -t.
    Sun, X. -l.
    [J]. CLINICAL RADIOLOGY, 2009, 64 (05) : 534 - 541
  • [9] The case for single-exposure angiography using energy-resolving photon-counting detectors: A theoretical comparison of signal and noise with conventional subtraction angiography
    Tanguay, Jesse
    Kim, Ho Kyung
    Cunningham, Ian. A.
    [J]. MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [10] CCD camera for dual-energy digital subtraction angiography
    Molloi, S
    Ersahin, A
    Qian, YJ
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1995, 14 (04) : 747 - 752