Calibration and Optimization of an Insertion Electromagnetic Flowmeter With the Aid of Numerical Analysis

被引:1
|
作者
Ma, Haibo [1 ]
Ma, Shilei [2 ]
Moss, Robert [2 ]
Fisher, Mark [2 ]
Silaen, Armin K. [1 ]
Zhou, Chenn Q. [1 ]
机构
[1] Purdue Univ Northwest, Ctr Innovat Visualizat & Simulat CIVS, Hammond, IN 46323 USA
[2] Dwyer Instruments, Michigan City, IN 46360 USA
关键词
computational foundations for engineering optimization; computer aided design; computer aided engineering; multiphysics modeling and simulation; physics-based simulations; MAGNETIC-FIELD DATA; VELOCITY PROFILE; 2-PHASE FLOW; LAMINAR-FLOW; SLUG FLOW; COMPUTATION; SIMULATIONS; PREDICTION; METER;
D O I
10.1115/1.4052579
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Accurate measurement of the volumetric flowrate of working liquids is essential for process control, as well as energy consumption evaluation. Electromagnetic flowmeters have gained popularity in applications where low-pressure drop and low maintenance are required. Dwyer Instruments, Inc. recently developed an adjustable insertion electromagnetic flowmeter (IEF) featuring accurate and reliable measurement. However, unexpected and non-repeatable behavior in the K-factor was observed during the calibration process. The K-factor is the coefficient used to convert the measured electric potential to the flow velocity in pipes, and the non-repeatable behavior imposes challenges for precise measurement. A one-way coupled magnetohydrodynamics model was developed to reduce the effort and time of onsite troubleshooting and optimization. By modeling the measurement process, the transition of flow regimes and the regeneration of the boundary layer on the electrode surface were identified as the causes of the non-repeatable issue. Then, a series of parametric studies were performed to provide reliable solutions. A new design with further embedded electrodes to allow the smooth transition between boundary layers was recommended. The field test showed excellent repeatability by using the new design, and the non-repeatable issue was entirely resolved. The improvement in the IEF design was implemented in production in less than one week, and it reduced the calibration time by 50%.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] CFD ANALYSIS AND IMPROVEMENT ON AN INSERTION ELECTROMAGNETIC FLOWMETER
    Ma, Haibo
    Silaen, Armin K.
    Zhou, Chenn Q.
    Ma, Shilei
    Moss, Robert
    Fisher, Mark
    [J]. PROCEEDINGS OF THE ASME 2020 INTERNATIONAL MECHANICAL ENGINEERING CONGRESS AND EXPOSITION, IMECE2020, VOL 6, 2020,
  • [2] NUMERICAL-ANALYSIS OF ELECTROMAGNETIC FLOWMETER
    BAKER, RC
    [J]. PROCEEDINGS OF THE INSTITUTION OF ELECTRICAL ENGINEERS-LONDON, 1973, 120 (09): : 1039 - 1043
  • [3] CALIBRATION TECHNIQUES FOR ELECTROMAGNETIC FLOWMETER
    MAKIN, GS
    BERGEL, DH
    WELBOURN, RB
    [J]. BRITISH JOURNAL OF SURGERY, 1968, 55 (01) : 74 - &
  • [4] Electromagnetic flowmeter used as calibration standard
    [J]. Pollution Equipment News, 2000, 33 (04):
  • [5] A TRANSIENT ELECTROMAGNETIC FLOWMETER AND CALIBRATION FACILITY
    LEFEBVRE, PJ
    DURGIN, WW
    [J]. JOURNAL OF FLUIDS ENGINEERING-TRANSACTIONS OF THE ASME, 1990, 112 (01): : 12 - 15
  • [6] Numerical calibration of a Lorentz force flowmeter
    Wang, Xiaodong
    Kolesnikov, Yurii
    Thess, Andre
    [J]. MEASUREMENT SCIENCE AND TECHNOLOGY, 2012, 23 (04)
  • [7] Optimization of electromagnetic flowmeter excitation coil
    [J]. Yue, Shihong (shyue1999@tju.edu.cn), 1953, Central South University of Technology (47):
  • [8] CALIBRATION AND USE OF SQUARE-WAVE ELECTROMAGNETIC FLOWMETER
    TETIRICK, JE
    MENGOLI, L
    [J]. SURGERY, 1963, 54 (04) : 621 - 626
  • [9] CALIBRATION CHARACTERISTICS OF PULSED-FIELD ELECTROMAGNETIC FLOWMETER
    BECK, R
    MORRIS, JA
    ASSALI, NS
    [J]. AMERICAN JOURNAL OF MEDICAL ELECTRONICS, 1965, 4 (02): : 87 - &
  • [10] Dry calibration procedure of electromagnetic flowmeter for open channels
    Warsaw Univ of Technology, Warszawa, Poland
    [J]. Conf Rec IEEE Instrum Meas Technol Conf, (1011-1016):