A 3D pencil-beam-based superposition algorithm for photon dose calculation in heterogeneous media

被引:104
|
作者
Tillikainen, L. [1 ]
Helminen, H. [1 ]
Torsti, T. [1 ]
Siljamaki, S. [1 ]
Alakuijala, J. [1 ]
Pyyry, J. [1 ]
Ulmer, W. [2 ]
机构
[1] Varian Med Syst Finland Oy, FIN-00270 Helsinki, Finland
[2] Varian Med Syst Imaging Lab GmbH, CH-5408 Baden, Switzerland
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2008年 / 53卷 / 14期
关键词
D O I
10.1088/0031-9155/53/14/008
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In thiswork, a novel three-dimensional superposition algorithm for photon dose calculation is presented. The dose calculation is performed as a superposition of pencil beams, which are modified based on tissue electron densities. The pencil beams have been derived from Monte Carlo simulations, and are separated into lateral and depth-directed components. The lateral component is modeled using exponential functions, which allows accurate modeling of lateral scatter in heterogeneous tissues. The depth-directed component represents the total energy deposited on each plane, which is spread out using the lateral scatter functions. Finally, convolution in the depth direction is applied to account for tissue interface effects. The method can be used with the previously introduced multiple-source model for clinical settings. The method was compared against Monte Carlo simulations in several phantoms including lung-and bone-type heterogeneities. Comparisons were made for several field sizes for 6 and 18 MV energies. The deviations were generally within (2%, 2 mm) of the field central axis dmax. Significantly larger deviations (up to 8%) were found only for the smallest field in the lung slab phantom for 18MV. The presented method was found to be accurate in a wide range of conditions making it suitable for clinical planning purposes.
引用
收藏
页码:3821 / 3839
页数:19
相关论文
共 50 条
  • [31] 3D photon dose calculation: Verification of the components by automated testing
    Pijpelink, J
    Linton, N
    Schulze, C
    PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, 1997, : 212 - 212
  • [32] Evaluating the accuracy of a three-term pencil beam algorithm in heterogeneous media
    Chapman, J. W.
    Knutson, N. C.
    Fontenot, J. D.
    Newhauser, W. D.
    Hogstrom, K. R.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (03): : 1172 - 1191
  • [33] A Fast Finite Size Pencil Beam Algorithm for Dose Calculation Using GPUs
    Arhjoul, L.
    Jinag, L.
    Solberg, T.
    Despres, P.
    Mao, W.
    MEDICAL PHYSICS, 2012, 39 (06) : 4020 - 4021
  • [34] ACCELERATION OF A PENCIL-BEAM DOSE CALCULATION ALGORITHM WITH GRAPHICS PROCESSING UNITS
    Arhjoul, L.
    Hissoiny, S.
    Ozell, B.
    Despres, P.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S70 - S70
  • [35] PENCIL BEAM VS COLLAPSED CONE TREATMENT PLANNING SYSTEM CALCULATION ALGORITHM IN 3D PLANNING OF LUNG CANCER.
    Paschalis, T.
    Silyvridou, A.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S582 - S582
  • [37] Accuracy of patient dose calculation for lung IMRT: A comparison of Monte Carlo, convolution/superposition, and pencil beam computations
    Vanderstraeten, Barbara
    Reynaert, Nick
    Paelinck, Leen
    Madani, Indira
    De Wagter, Carlos
    De Gersem, Werner
    De Neve, Wilfried
    Thierens, Hubert
    MEDICAL PHYSICS, 2006, 33 (09) : 3149 - 3158
  • [38] Photon Dose Calculation Method Based on Monte Carlo Finite-Size Pencil Beam Model in Accurate Radiotherapy
    Zheng, Huaqing
    Sun, Guangyao
    Li, Gui
    Cao, Ruifen
    Pei, Xi
    Hu, Liqin
    Song, Gang
    Wu, Yican
    COMMUNICATIONS IN COMPUTATIONAL PHYSICS, 2013, 14 (05) : 1415 - 1422
  • [39] 3D photon dose calculation: From a scientific tool to routine planning
    Schulze, C
    Pijpelink, J
    Linton, N
    Bortfeld, T
    Schlegel, W
    PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, 1997, : 43 - 45
  • [40] Pacemakers in radiotherapy - how accurate is the calculation of dose on cardiac pacemakers with a Pencil Beam algorithm?
    Scheermann, J.
    Walke, M.
    Gademann, G.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2009, 185 : 165 - 165