Modeling the Agility MLC in the Monaco treatment planning system

被引:11
|
作者
Snyder, Michael [1 ,2 ]
Halford, Robert [2 ]
Knill, Cory [2 ]
Adams, Jeffrey N. [2 ]
Bossenberger, Todd [2 ]
Nalichowski, Adrian [2 ]
Hammoud, Ahmad [2 ]
Burmeister, Jay [1 ,2 ]
机构
[1] Wayne State Univ, Sch Med, Dept Oncol, Detroit, MI 48201 USA
[2] Karmanos Canc Inst, Dept Radiat Oncol, Detroit, MI USA
来源
关键词
MLC modeling; IMRT commissioning; Monaco; MODULATED RADIATION-THERAPY; CARLO DOSE CALCULATION; QUALITY-ASSURANCE; IMRT; ALGORITHM; VMC;
D O I
10.1120/jacmp.v17i3.6044
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We investigate the relationship between the various parameters in the Monaco MLC model and dose calculation accuracy for an Elekta Agility MLC. The vendor-provided MLC modeling procedure - completed first with external vendor participation and then exclusively in-house - was used in combination with our own procedures to investigate several sets of MLC modeling parameters to determine their effect on dose distributions and point-dose measurements. Simple plans provided in the vendor procedure were used to elucidate specific mechanical characteristics of the MLC, while ten complex treatment plans - five IMRT and five VMAT - created using TG-119-based structure sets were used to test clinical dosimetric effects of particular parameter choices. EDR2 film was used for the vendor fields to give high spatial resolution, while a combination of MapCHECK and ion chambers were used for the in-house TG-119-based procedures. The vendor-determined parameter set provided a reasonable starting point for the MLC model and largely delivered acceptable gamma pass rates for clinical plans - including a passing external evaluation using the IROC H&N phantom. However, the vendor model did not provide point-dose accuracy consistent with that seen in other treatment systems at our center. Through further internal testing it was found that there existed many sets of MLC parameters, often at opposite ends of their allowable ranges, that provided similar dosimetric characteristics and good agreement with planar and point-dose measurements. In particular, the leaf offset and tip leakage parameters compensated for one another if adjusted in opposite directions, which provided a level curve of acceptable parameter sets across all plans. Interestingly, gamma pass rates of the plans were less dependent upon parameter choices than point-dose measurements, suggesting that MLC modeling using only gamma evaluation may be generally an insufficient approach. It was also found that exploring all parameters of the very robust MLC model to find the best match to the vendor-provided QA fields can reduce the pass rates of the TG-119-based clinical distributions as compared to simpler models. A wide variety of parameter sets produced MLC models capable of meeting RPC passing criteria for their H&N IMRT phantom. The most accurate models were achievable using a combination of vendor-provided and in-house procedures. The potential existed for an over-modeling of the Agility MLC in an effort to obtain the fine structure of certain quality assurance fields, which led to a reduction in agreement between calculation and measurement of more typical clinical dose distributions.
引用
收藏
页码:190 / 202
页数:13
相关论文
共 50 条
  • [1] Agility MLC transmission optimization in the Monaco treatment planning system
    Roche, Michael
    Crane, Robert
    Powers, Marcus
    Crabtree, Timothy
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (05): : 473 - 482
  • [2] Optimizing Agility MLC model in the Monaco treatment planning system
    Miguel, F. San
    Chamorro, P.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S1316 - S1317
  • [3] Transmission probability filter optimization for Agility MLC in Monaco treatment planning system
    Thewes, Lena
    Eckl, Miriam
    Schneider, Frank
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (09):
  • [4] Dosimetric comparison of Thoracic SBRT planning in Monaco and Eclipse for the Elekta Agility MLC
    Villamizar, R. J. Plata
    Ribeiro, M.
    Torzsok, C. K.
    Broque, H.
    Ruiz, A.
    Rodriguez-Mongua, J. L.
    Aponte, J.
    Marangoni, F.
    Pino, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2023, 182 : S1781 - S1782
  • [5] Treatment couch modeling in Elekta Monaco treatment planning system
    Huertas, C.
    Ferrer, C.
    Huerga, C.
    Mas, I.
    Serrada, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S389 - S390
  • [6] iBEAM evo Couch Modeling and Validation for Monaco Treatment Planning System
    Thiyagarajan, R.
    Mariagandhi, E.
    Kataria, T.
    Manigandan, D.
    TamilSelvan
    Kamaraj, D.
    Veni, M.
    Nambiraj, N. Arunai
    [J]. MEDICAL PHYSICS, 2018, 45 (06) : E303 - E303
  • [7] Validation of Elekta couch modeling for dose calculation in the Monaco treatment planning system
    Poli, E.
    Reis, P.
    Prudencio, L.
    Galhardas, J.
    Ribeiro, T.
    Malveiro, R.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S991 - S991
  • [8] A Test Suite for Optimization and Verification of MLC Treatment Planning System Modeling
    Popple, R.
    Barnes, M.
    Covington, E.
    Farrey, K.
    Moran, J.
    Perez, M.
    Raxter, D.
    Ritter, T.
    [J]. MEDICAL PHYSICS, 2017, 44 (06) : 2968 - 2968
  • [9] Case Studies with the Monaco Treatment Planning System
    Culberson, W.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [10] Modelling the MLC Agility transmission radiation through the closed leaf gap in the Monaco TPS
    Vicedo Gonzalez, M. A.
    Garcia Hernandez, M. T.
    Planes Meseguer, D.
    Garcia Molla, R.
    Borderia Navarro, B.
    Calatayud Cuesta, P.
    Alcaraz Lozano, M. L.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S1296 - S1297