Monte Carlo simulation and dosimetric verification of radiotherapy beam modifiers

被引:26
|
作者
Spezi, E [1 ]
Lewis, DG [1 ]
Smith, CW [1 ]
机构
[1] Velindre Hosp, Dept Med Phys, Cardiff CF14 2TL, S Glam, Wales
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2001年 / 46卷 / 11期
关键词
D O I
10.1088/0031-9155/46/11/316
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Monte Carlo simulation of beam modifiers such as physical wedges and compensating filters has been performed with a rectilinear voxel geometry module. A modified version of the EGS4/DOSXYZ code has been developed for this purpose. The new implementations have been validated against the BEAM Monte Carlo code using its standard component modules (CMs) in several geometrical conditions. No significant disagreements were found within the statistical errors of 0.5% for photons and 2% for electrons. The clinical applicability and flexibility of the new version of the code has been assessed through an extensive verification versus dosimetric data. Both Varian multi-leaf collimator (MLC) wedges and standard wedges have been simulated and compared against experiments for 6 MV photon beams and different field sizes. Good agreement was found between calculated and measured depth doses and lateral dose profiles along both wedged and unwedged directions for different depths and focus-to-surface distances. Furthermore, Monte Carlo-generated output factors for both open and wedged fields agreed with linac commissioning beam data within statistical uncertainties of the calculations (<3% at largest depths). Compensating filters of both low-density and high-density materials have also been successfully simulated. As a demonstration, a wax compensating filter with a complex three-dimensional concave and convex geometry has been modelled through a CT scan import. Calculated depth doses and lateral dose profiles for different field sizes agreed well with experiments. The code was used to investigate the performance of a commercial treatment planning system in designing compensators. Dose distributions in a heterogeneous water phantom emulating the head and neck region were calculated with the convolution-superposition method (pencil beam and collapsed cone implementations) and compared against those from the MC code developed herein. The new technique presented in this work is versatile, DICOM-RT compliant and accurate in the simulation of beam modulators. This paper addresses the need to reduce the sources of error in the modelling of beam modifiers since they remain a viable alternative to the MLC technique in the delivery of IMRT beams.
引用
收藏
页码:3007 / 3029
页数:23
相关论文
共 50 条
  • [41] Comparison of Monte Carlo Simulation and Measurement for Determination of Dosimetric Leaf Gap
    Popple, R.
    Sawkey, D.
    Yuan, Y.
    Sullivan, R.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E262 - E262
  • [42] Dosimetric comparison between different radiotherapy protocols for prostate cancer using Geant4 Monte Carlo simulation
    Syahir, M. K.
    Fahmi, M. R.
    Hashikin, N. A. A.
    [J]. 11TH INTERNATIONAL SEMINAR ON MEDICAL PHYSICS (ISMP) 2019, 2020, 1497
  • [43] MONTE CARLO VERIFICATION OF INTENSITY MODULATED RADIOTHERAPY TREATMENTS USING EPIDS
    Perera, L.
    Boyd, J.
    Doody, J.
    Moody, M.
    Sandford, P.
    Fielding, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S439 - S439
  • [44] Simple beam models for Monte Carlo photon beam dose calculations in radiotherapy
    Fix, MK
    Keller, H
    Rüegsegger, P
    Born, EJ
    [J]. MEDICAL PHYSICS, 2000, 27 (12) : 2739 - 2747
  • [45] Dosimetric data for broad beam, pencil beam and Monte Carlo models of clinical proton beams
    Mazal, A
    Kooy, H
    Nauraye, C
    Schaffner, B
    Rosenthal, S
    Kollipara, S
    Paganetti, H
    Ferrand, R
    Rosenwald, JC
    Chen, G
    [J]. RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S321 - S321
  • [46] Monte Carlo dosimetric verification for IMRT QA using MLC log files and EPID
    Luo, W
    Li, J
    Fourkal, E
    Fan, J
    Yang, J
    Price, R
    Ma, C
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 1985 - 1985
  • [47] Dosimetric verification of IMRT treatment planning using Monte Carlo simulations for prostate cancer
    Yang, J
    Li, J
    Chen, L
    Price, R
    McNeeley, S
    Qin, L
    Wang, L
    Xiong, W
    Ma, CM
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (05): : 869 - 878
  • [48] Verification of the Simplified Monte Carlo Method for Simulation in An Inhomogeneous Phantom
    Mizutani, S.
    Takada, Y.
    Kohno, R.
    Hotta, K.
    Akimoto, T.
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3523 - 3523
  • [49] Beam hardening correction based on Monte Carlo simulation
    Zeng, G
    Yu, ZQ
    Yan, YL
    [J]. HIGH ENERGY PHYSICS AND NUCLEAR PHYSICS-CHINESE EDITION, 2006, 30 (02): : 178 - 182
  • [50] Monte Carlo simulation of single beam dose profile
    LIU Xiaowei
    WEN Guanghao
    LI Mianfeng(Zhongshan University
    [J]. Nuclear Science and Techniques, 1999, (02) : 87 - 90