Error bounds on ultrasonic scatterer size estimates

被引:43
|
作者
Chaturvedi, P
Insana, MF
机构
[1] Department of Radiology, University of Kansas, Medical Center, Kansas City
来源
关键词
D O I
10.1121/1.415958
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Precision errors that occur in estimating the average scatterer size from pulse-echo ultrasound waveforms are examined in detail. The method-independent lower bound on estimation error is found from the Cramer-Rao inequality for comparison with the predicted error for the measurement technique currently used to estimate scatterer sizes in soft biological tissues. The probability density function for the estimate is also derived. From these statistical analyses, strategies for designing experiments that minimize the error are discussed. It is shown that compared with biological variability, measurement errors in scatterer size estimates are relatively large. Consequently, there is reason to continue searching for more efficient estimators. Although the analysis and results are derived for Gaussian correlation models that have been used to study the function and structure of kidneys, generalization to include correlation models for other tissues is straightforward. (C) 1996 Acoustical Society of America.
引用
收藏
页码:392 / 399
页数:8
相关论文
共 50 条