Collimator angle influence on dose distribution optimization for vertebral metastases using volumetric modulated arc therapy

被引:19
|
作者
Mancosu, Pietro [1 ]
Cozzi, Luca [2 ]
Fogliata, Antonella [2 ]
Lattuada, Paola [1 ]
Reggiori, Giacomo [1 ]
Cantone, Marie Claire [3 ]
Navarria, Pierina [1 ]
Scorsetti, Marta [1 ]
机构
[1] IRCCS Ist Clin Humanitas, Dept Radiat Oncol, I-20089 Milan, Rozzano, Italy
[2] Oncol Inst So Switzerland, Med Phys Unit, CH-6504 Bellinzona, Switzerland
[3] Univ Milan, Dept Phys, I-20133 Milan, Italy
关键词
RapidArc; spinal cord compression; volumetric modulated arc therapy (VMAT); planning optimization; RADIOTHERAPY; RAPIDARC; IMRT;
D O I
10.1118/1.3462560
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The cylindrical symmetry of vertebrae favors the use of volumetric modulated arc therapy in generating a dose "hole" on the center of the vertebrae limiting the dose to the spinal cord. The authors have evaluated if collimator angle is a significant parameter for dose distribution optimization in vertebral metastases. Methods: Three patients with one-three vertebrae involved were considered. Twenty-one differently optimized plans (nine single-arc and 12 double-arc plans) were performed, testing various collimator angle positions. Clinical target volume was defined as the whole vertebrae, excluding the spinal cord canal. The planning target volume (PTV) was defined as CTV+5 mm. Dose prescription was 5 x 4 Gy(2) with normalization to PTV mean dose. The dose at 1 cm(3) of spinal cord was limited to 11.5Gy. Results: The best plans in terms of target coverage and spinal cord sparing were achieved by two arcs and Arc1-80 degrees and Arc2-280 degrees collimator angles for all the cases considered (i.e., leaf travel parallel to the spinal cord primary orientation). If one arc is used, only 80 degrees reached the objectives. Conclusions: This study demonstrated the role of collimation rotation for the vertebrae metastasis irradiation, with the leaf travel parallel to the spinal cord primary orientation to be better than other solutions. Thus, optimal choice of collimator angle increases the optimization freedom to shape a desired dose distribution. (C) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3462560]
引用
收藏
页码:4133 / 4137
页数:5
相关论文
共 50 条
  • [31] Monte Carlo Dose Verification for Volumetric Modulated Arc Therapy in Multiple Brain Metastases
    Tominaga, H.
    Araki, F.
    Kanetake, N.
    Sakata, J.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3546 - 3547
  • [32] A Collimator Setting Optimization Algorithm for Dual-Arc Volumetric Modulated Arc Therapy in Pancreas Stereotactic Body Radiation Therapy
    Li, Xinyi
    Wu, Jackie
    Palta, Manisha
    Zhang, You
    Sheng, Yang
    Zhang, Jiahan
    Wang, Chunhao
    [J]. TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2019, 18
  • [33] Feasibility of dose painting using volumetric modulated arc optimization and delivery
    Korreman, Stine S.
    Ulrich, Silke
    Bowen, Steve
    Deveau, Michael
    Bentzen, Soren M.
    Jeraj, Robert
    [J]. ACTA ONCOLOGICA, 2010, 49 (07) : 964 - 971
  • [34] FEASIBILITY OF DOSE PAINTING USING VOLUMETRIC MODULATED ARC (RAPIDARC) OPTIMIZATION
    Korreman, S.
    Ulrich, S.
    Bowen, S.
    Deveau, M.
    Jeraj, R.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S554 - S554
  • [35] EFFECTS OF INTRAFRACTIONAL MOTION ON THE DOSE DISTRIBUTION IN VOLUMETRIC MODULATED ARC THERAPY (VMAT)
    Fleckenstein, J.
    Boggula, R.
    Stieler, F.
    Lennart, J.
    Wenz, F.
    Lohr, F.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S96 - S96
  • [36] Volumetric modulated arc therapy with dynamic collimator rotation for improved multileaf collimator tracking of the prostate
    Murtaza, Ghulam
    Toftegaard, Jakob
    Khan, Ehsan Ullah
    Poulsen, Per Rugaard
    [J]. RADIOTHERAPY AND ONCOLOGY, 2017, 122 (01) : 109 - 115
  • [37] VOLUMETRIC MODULATED ARC RADIOTHERAPY FOR RAPID DELIVERY OF STEREOTACTIC RADIOTHERAPY TO VERTEBRAL METASTASES
    Kuijper, I.
    Verbakel, W.
    Cuijpers, J.
    Haasbeek, N.
    Senan, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S90 - S90
  • [38] Effect of Gantry and Collimator Positional Errors On Volumetric Modulated Arc Therapy(VMAT)
    Holla, R.
    Ravikumar, V. Anand
    Pillai, B.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [39] Dynamic positioning accuracy of a novel multileaf collimator for volumetric modulated arc therapy
    Nakaguchi Y.
    Ono T.
    Onizuka R.
    Maruyama M.
    Shimohigashi Y.
    Kai Y.
    [J]. Radiological Physics and Technology, 2016, 9 (1) : 121 - 126
  • [40] IMPACT OF COLLIMATOR LEAF WIDTH FOR INTENSITY MODULATED RADIOTHERAPY AND VOLUMETRIC INTENSITY MODULATED ARC THERAPY
    Krayenbuehl, J.
    Linsenmeier, C.
    Kloeck, S.
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2011, 187 (08) : 518 - 518