Collimator angle influence on dose distribution optimization for vertebral metastases using volumetric modulated arc therapy

被引:19
|
作者
Mancosu, Pietro [1 ]
Cozzi, Luca [2 ]
Fogliata, Antonella [2 ]
Lattuada, Paola [1 ]
Reggiori, Giacomo [1 ]
Cantone, Marie Claire [3 ]
Navarria, Pierina [1 ]
Scorsetti, Marta [1 ]
机构
[1] IRCCS Ist Clin Humanitas, Dept Radiat Oncol, I-20089 Milan, Rozzano, Italy
[2] Oncol Inst So Switzerland, Med Phys Unit, CH-6504 Bellinzona, Switzerland
[3] Univ Milan, Dept Phys, I-20133 Milan, Italy
关键词
RapidArc; spinal cord compression; volumetric modulated arc therapy (VMAT); planning optimization; RADIOTHERAPY; RAPIDARC; IMRT;
D O I
10.1118/1.3462560
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The cylindrical symmetry of vertebrae favors the use of volumetric modulated arc therapy in generating a dose "hole" on the center of the vertebrae limiting the dose to the spinal cord. The authors have evaluated if collimator angle is a significant parameter for dose distribution optimization in vertebral metastases. Methods: Three patients with one-three vertebrae involved were considered. Twenty-one differently optimized plans (nine single-arc and 12 double-arc plans) were performed, testing various collimator angle positions. Clinical target volume was defined as the whole vertebrae, excluding the spinal cord canal. The planning target volume (PTV) was defined as CTV+5 mm. Dose prescription was 5 x 4 Gy(2) with normalization to PTV mean dose. The dose at 1 cm(3) of spinal cord was limited to 11.5Gy. Results: The best plans in terms of target coverage and spinal cord sparing were achieved by two arcs and Arc1-80 degrees and Arc2-280 degrees collimator angles for all the cases considered (i.e., leaf travel parallel to the spinal cord primary orientation). If one arc is used, only 80 degrees reached the objectives. Conclusions: This study demonstrated the role of collimation rotation for the vertebrae metastasis irradiation, with the leaf travel parallel to the spinal cord primary orientation to be better than other solutions. Thus, optimal choice of collimator angle increases the optimization freedom to shape a desired dose distribution. (C) 2010 American Association of Physicists in Medicine. [DOI: 10.1118/1.3462560]
引用
收藏
页码:4133 / 4137
页数:5
相关论文
共 50 条
  • [1] Collimator Angle Optimization in Volumetric Modulated Arc Therapy
    Andersen, A.
    Johnson, C.
    Bartlett, G.
    Das, I.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3627 - 3627
  • [2] IS COLLIMATOR ANGLE SIGNIFICANT FOR DOSE DISTRIBUTION OPTIMIZATION IN VERTEBRAL METASTASES USING RAPIDARC?
    Lattuada, P.
    Mancosu, P.
    Lobefalo, F.
    Navarria, P.
    Reggiori, G.
    Fogliata, A.
    Cozzi, L.
    Cantone, M. C.
    Urso, G.
    Scorsetti, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S503 - S504
  • [3] The importance of collimator angle error in volumetric-modulated arc therapy
    Puchades-Puchades, Vicente
    Serna-Berna, Alfredo
    Mata-Colodro, Fernando
    Ramos-Amores, David
    Alcaraz-Banos, Miguel
    [J]. JOURNAL OF RADIOTHERAPY IN PRACTICE, 2015, 14 (03) : 323 - 330
  • [4] Dosimetric Dependence On the Collimator Angle in Prostate Volumetric Modulated Arc Therapy
    Khan, M.
    Rehman, J.
    Khan, M.
    Chow, J.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 366 - 367
  • [5] Impact of Collimator Angle on Deliverability of Volumetric Modulated Arc Therapy (VMAT)
    Hong, D.
    Bundalevski, I.
    Wong, K.
    Mokhtar, B. E.
    Ashamalla, H.
    Parameritis, J.
    Kyriacou, A.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E632 - E632
  • [6] Effect of the Collimator Angle on Dosimetric Verification of Volumetric Modulated Arc Therapy
    Kim, Yong Ho
    Park, Ha Ryung
    Kim, Won Taek
    Kim, Dong Won
    Ki, Yongkan
    Lee, Juhye
    Bae, Jinsuk
    Park, Dahl
    Jeon, Hosang
    Nam, Ji Ho
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2015, 67 (01) : 243 - 247
  • [7] Effect of the collimator angle on dosimetric verification of volumetric modulated arc therapy
    Yong Ho Kim
    Ha Ryung Park
    Won Taek Kim
    Dong Won Kim
    Yongkan Ki
    Juhye Lee
    Jinsuk Bae
    Dahl Park
    Hosang Jeon
    Ji Ho Nam
    [J]. Journal of the Korean Physical Society, 2015, 67 : 243 - 247
  • [8] Influence of Gantry Angle, Multi-Leaf Collimator (MLC), and Jaw Position Errors to Dose Distribution in Volumetric Modulated Arc Therapy (VMAT)
    Haga, A.
    Nakagawa, K.
    Shiraishi, K.
    Itoh, S.
    Saegusa, S.
    Imae, T.
    Sasaki, K.
    Kida, S.
    Terahara, A.
    Yoda, K.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [9] An investigation of the dose distribution effect related with collimator angle in volumetric arc therapy of prostate cancer
    Tas, Bora
    Bilge, Hatice
    Ozturk, Sibel Tokdemir
    [J]. JOURNAL OF MEDICAL PHYSICS, 2016, 41 (02) : 100 - 105
  • [10] Volumetric modulated arc therapy optimization including dynamic collimator rotation
    Fix, M. K.
    Frei, D.
    Volken, W.
    Terribilini, D.
    Manser, P.