Benchmarking a GATE/Geant4 Monte Carlo model for proton beams in magnetic fields

被引:17
|
作者
Padilla-Cabal, Fatima [1 ,2 ]
Alejandro Fragoso, Jose [3 ]
Resch, Andreas Franz [1 ,2 ]
Georg, Dietmar [1 ,2 ]
Fuchs, Hermann [1 ,2 ]
机构
[1] Med Univ Vienna AKH, Dept Radiotherapy, Vienna, Austria
[2] Med Univ Vienna, Christian Doppler Lab Med Radiat Res Radiat Oncol, Vienna, Austria
[3] Higher Inst Technol & Appl Sci, Dept Nucl Phys, Havana, Cuba
基金
奥地利科学基金会;
关键词
beam modeling; magnetic fields; MC simulations; proton therapy; RADIOCHROMIC FILM DOSIMETRY; FLAT-BED SCANNERS; DOSE CALCULATION; EBT3; FILM; THERAPY; SIMULATION; CARBON; FEASIBILITY; VALIDATION; DEFLECTION;
D O I
10.1002/mp.13883
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeMagnetic resonance guidance in proton therapy (MRPT) is expected to improve its current performance. The combination of magnetic fields with clinical proton beam lines poses several challenges for dosimetry, treatment planning and dose delivery. Proton beams are deflected by magnetic fields causing considerable changes in beam trajectories and also a retraction of the Bragg peak positions. A proper prediction and compensation of these effects is essential to ensure accurate dose calculations. This work aims to develop and benchmark a Monte Carlo (MC) beam model for dose calculation of MRPT for static magnetic fields up to 1 T. MethodsProton beam interactions with magnetic fields were simulated using the GATE/Geant4 toolkit. The transport of charged particle in custom 3D magnetic field maps was implemented for the first time in GATE. Validation experiments were done using a horizontal proton pencil beam scanning system with energies between 62.4 and 252.7 MeV and a large gap dipole magnet (B = 0-1 T), positioned at the isocenter and creating magnetic fields transverse to the beam direction. Dose was measured with Gafchromic EBT3 films within a homogeneous PMMA phantom without and with bone and tissue equivalent material slab inserts. Linear energy transfer (LET) quenching of EBT3 films was corrected using a linear model on dose-averaged LET method to ensure a realistic dosimetric comparison between simulations and experiments. Planar dose distributions were measured with the films in two different configurations: parallel and transverse to the beam direction using single energy fields and spread-out Bragg peaks. The MC model was benchmarked against lateral deflections and spot sizes in air of single beams measured with a Lynx PT detector, as well as dose distributions using EBT3 films. Experimental and calculated dose distributions were compared to test the accuracy of the model. ResultsMeasured proton beam deflections in air at distances of 465, 665, and 1155 mm behind the isocenter after passing the magnetic field region agreed with MC-predicted values within 4 mm. Differences between calculated and measured beam full width at half maximum (FWHM) were lower than 2 mm. For the homogeneous phantom, measured and simulated in-depth dose profiles showed range and average dose differences below 0.2 mm and 1.2%, respectively. Simulated central beam positions and widths differed <1 mm to the measurements with films. For both heterogenous phantoms, differences within 1 mm between measured and simulated central beam positions and widths were obtained, confirming a good agreement of the MC model. ConclusionsA GATE/Geant4 beam model for protons interacting with magnetic fields up to 1 T was developed and benchmarked to experimental data. For the first time, the GATE/Geant4 model was successfully validated not only for single energy beams, but for SOBP, in homogeneous and heterogeneous phantoms. EBT3 film dosimetry demonstrated to be a powerful dosimetric tool, once the film response function is LET corrected, for measurements in-line and transverse to the beam direction in magnetic fields. The proposed MC beam model is foreseen to support treatment planning and quality assurance (QA) activities toward MRPT.
引用
收藏
页码:223 / 233
页数:11
相关论文
共 50 条
  • [1] Monte Carlo simulations in proton dosimetry with Geant4
    Moravek, Z
    Bogner, L
    [J]. RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S62 - S62
  • [2] Analytical and Monte-Carlo (GEANT4) calculations of collimator scatter of proton beams
    Ulmer, W.
    Schaffner, B.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2419 - 2419
  • [3] A Monte Carlo pencil beam scanning model for proton treatment plan simulation using GATE/GEANT4
    Grevillot, L.
    Bertrand, D.
    Dessy, F.
    Freud, N.
    Sarrut, D.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (16): : 5203 - 5219
  • [4] Proton Source Modeling for Geant4 Monte Carlo Simulations
    Barnes, S.
    McAuley, G.
    Wroe, A.
    Slater, J.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3756 - 3757
  • [5] The Geant4 Virtual Monte Carlo
    Hrivnacova, I.
    [J]. INTERNATIONAL CONFERENCE ON COMPUTING IN HIGH ENERGY AND NUCLEAR PHYSICS 2012 (CHEP2012), PTS 1-6, 2012, 396
  • [6] Benchmarking Gate/Geant4 for oxygen ion beams against experimental data
    Resch, A.
    Fuchs, H.
    Georg, D.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S427 - S427
  • [7] Analysis of the physical interactions of therapeutic proton beams in water with the use of Geant4 Monte Carlo calculations
    Moravek, Zdenek
    Bogner, Ludwig
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2009, 19 (03): : 174 - 181
  • [8] Monte Carlo calculation of beam quality correction factors in proton beams using TOPAS/GEANT4
    Baumann, Kilian-Simon
    Kaupa, Sina
    Bach, Constantin
    Engenhart-Cabillic, Rita
    Zink, Klemens
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (05):
  • [9] Monte Carlo simulation of the lateral spread for proton beams in voxelized water phantom using GEANT4 platform
    Kaanouch, Othmane
    Krim, Mustapha
    Ghazi, Ismail
    Saidi, Kamal
    Essaidi, El Mehdi
    Tantaoui, Meriem
    Saad, El Madani
    [J]. JAPANESE JOURNAL OF APPLIED PHYSICS, 2023, 62 (01)
  • [10] Proton transport in water and DNA components: A Geant4 Monte Carlo simulation
    Champion, C.
    Incerti, S.
    Tran, H. N.
    Karamitros, M.
    Shin, J. I.
    Lee, S. B.
    Lekadir, H.
    Bernal, M.
    Francis, Z.
    Ivanchenko, V.
    Fojon, O. A.
    Hanssen, J.
    Rivarola, R. D.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2013, 306 : 165 - 168