Research on Spectroradiometer for Irradiance Calibration in Ageing Tests

被引:0
|
作者
Lin, Fangsheng [1 ]
Yin, Dejin [1 ]
Li, Tiecheng [1 ]
Ji, Muyao [1 ]
机构
[1] Shanghai Inst Measurement & Testing Technol, Shanghai 201203, Peoples R China
关键词
spectroradiometer; irradiance calibration; ageing test; xenon arc lamp;
D O I
10.1117/12.2541496
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Instead of natural exposing tests in atmosphere environment, we use artificial accelerated weathering apparatus to simulate solar radiation in ageing tests. Xenon arc lamps are most widely used to simulate spectral distribution of the sun in chambers. According to JJF1525-2015 Calibration Specification for Irradiance of Artificial Accelerated Weathering Apparatus of Xenon Arc Lamp, there are three methods to detect the irradiance: radiometer method, spectroradiometer method and standard lamp method. In this paper, we compare three methods and focus on the spectroradiometer method. Because different wavelength ranges need different radiometers, the best method to calibrate the irradiance in ageing tests is spectroradiometer method. By experiments, we verify the irradiance correspondence between 300nm similar to 400nm and 340nm or 420nm to test the accuracy of spectroradiometer. We also study the factors affecting the irradiance calibration by spectroradiometer. At last we conclude that spectroradiometer is more suitable for irradiance calibration in different ageing tests and provide guidance for daily calibration.
引用
收藏
页数:6
相关论文
共 50 条
  • [21] Infrared irradiance calibration
    Stephan D. Price
    Space Science Reviews, 2004, 113 : 409 - 456
  • [22] Infrared irradiance calibration
    Price, SD
    SPACE SCIENCE REVIEWS, 2004, 113 (3-4) : 409 - 456
  • [23] The Precision Solar Spectroradiometer (PSR) for direct solar irradiance measurements
    Grobner, Julian
    Kouremeti, Natalia
    SOLAR ENERGY, 2019, 185 : 199 - 210
  • [24] Global spectral irradiance array spectroradiometer validation according to WMO
    Zuber, Ralf
    Ribnitzky, Mario
    Tobar, Mario
    Lange, Kezia
    Kutscher, Dimitrij
    Schrempf, Michael
    Niedzwiedz, Angelika
    Seckmeyer, Gunther
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2018, 29 (10)
  • [25] The Calibration of Integrating Sphere with Spectroradiometer with Traceability to Goniophotometer
    Motycka, Martin
    Skoda, Jan
    Stepanek, Jaroslav
    PROCEEDINGS OF THE 2018 19TH INTERNATIONAL SCIENTIFIC CONFERENCE ON ELECTRIC POWER ENGINEERING (EPE), 2018,
  • [26] A calibration method of infrared LVF based spectroradiometer
    Liu, Jiaqing
    Han, Shunli
    Liu, Lei
    Hu, Dexin
    AOPC 2017: OPTICAL SPECTROSCOPY AND IMAGING, 2017, 10461
  • [27] Wavelength calibration of shortwave infrared flat spectroradiometer
    Li, Xin
    Zhang, Guowei
    Xun, Lina
    Xie, Ping
    Hong, Jin
    Zheng, Xiaobing
    Guangxue Xuebao/Acta Optica Sinica, 2008, 28 (05): : 902 - 906
  • [28] Research on high-accuracy in-flight spectral calibration of the solar backscattered ultraviolet spectroradiometer
    Li, Zhanfeng
    Wang, Shurong
    Huang, Yu
    Yu, Xiangyang
    Guangxue Xuebao/Acta Optica Sinica, 2013, 33 (02):
  • [29] Calibration of the total irradiance monitor
    Lawrence, GM
    Kopp, G
    Rottman, G
    Harder, J
    Woods, T
    Loui, H
    METROLOGIA, 2003, 40 (01) : S78 - S80
  • [30] Spectroradiometer spectral calibration, ISRF shapes, and related uncertainties
    Trim, Simon A.
    Mason, Kimberley
    Hueni, Andreas
    APPLIED OPTICS, 2021, 60 (18) : 5405 - 5417