Gamma-Ray Point-Source Detection in Unknown Background Using 3D-Position-Sensitive Semiconductor Detectors

被引:11
|
作者
Wahl, Christopher G. [1 ]
He, Zhong [1 ]
机构
[1] Univ Michigan, Dept Nucl Engn & Radiol Sci, Ann Arbor, MI 48109 USA
关键词
Gamma rays; maximum likelihood detection; radiation detection algorithms; radiation imaging; PORTAL MONITORS; LARGE-AREA; LIKELIHOOD; SOFTWARE;
D O I
10.1109/TNS.2011.2113355
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Three-dimensional-position-sensitive semiconductor detectors provide both energy and locations of gamma-ray interactions in the detector. From these data, Compton-imaging methods can reconstruct the spatial source distribution around the detector as a function of energy. Here, the generalized likelihood ratio test (GLRT) and the source-intensity test (SIT) are applied to automate detection and identification of point sources of radiation in distributed background by using the combined energy and direction information from these detectors. Throughout, the model makes no assumptions of the background intensity or energy distribution. However, directional information enables detection and identification without prior knowledge of the background spectrum and the detection of isotopes also appearing in the background. Spatially, background is modeled as uniform, but only slight degradation in performance was observed when the background was twice as intense in one hemisphere. Experimental detection and identification performance as a function of false-alarm probability was measured for three weak isotopes using experimental data collected on a 20 mm x 20 mm x 15 mm CdZnTe detector. Both the GLRT and SIT performed better than the commercial software Genie 2000 (Canberra), which uses only spectral information, when the possible source direction was unknown. Prior knowledge of the possible source direction further improves both detection and identification.
引用
收藏
页码:605 / 613
页数:9
相关论文
共 50 条
  • [31] SOURCE POSITION EFFECT IN GAMMA-RAY ENERGY MEASUREMENTS WITH GE(LI) DETECTORS
    SHIZUMA, K
    INOUE, H
    YOSHIZAWA, Y
    SAKAI, E
    KATAGIRI, M
    NUCLEAR INSTRUMENTS & METHODS, 1978, 157 (01): : 117 - 126
  • [32] Transient Behavior in TlBr Gamma-Ray Detectors and Its Analysis Using 3-D Position Sensing
    Thrall, Crystal L.
    Kaye, William R.
    He, Zhong
    Kim, Hadong
    Cirignano, Leonard
    Shah, Kanai
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (02) : 1162 - 1167
  • [33] GAMMA-RAY DOSE BUILDUP FACTORS AT GROUND-LEVEL FOR A POINT-SOURCE IN AIR
    TAKADA, K
    SAITO, K
    MORIUCHI, S
    RADIATION PROTECTION DOSIMETRY, 1985, 11 (04) : 237 - 244
  • [34] POINT-SOURCE SEARCH TECHNIQUES IN ULTRA-HIGH ENERGY GAMMA-RAY ASTRONOMY
    ALEXANDREAS, DE
    BERLEY, D
    BILLER, S
    DION, GM
    GOODMAN, JA
    HAINES, TJ
    HOFFMAN, CM
    HORCH, E
    LU, XQ
    SINNIS, C
    YODH, GB
    ZHANG, W
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 328 (03): : 570 - 577
  • [35] Detection of gamma ray polarization using a 3D position sensitive CdZnTe detector
    Xu, D
    He, Z
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 376 - 380
  • [36] Coded Aperture Gamma-Ray Imaging Using Pixelated Semiconductor Detectors
    Joshi, S.
    Kaye, W.
    Jaworski, J.
    He, Z.
    MEDICAL PHYSICS, 2015, 42 (06) : 3202 - 3203
  • [37] CALCULATION OF THE POINT-SOURCE EFFECTIVE COLLIMATOR APERTURE FOR COLLIMATED X-RAY AND GAMMA-RAY TELESCOPES
    OWENS, A
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1985, 239 (03): : 623 - 629
  • [38] Enhancing the crystal discrimination for position-sensitive gamma-ray detectors by incorporating interaction position information
    Wei, Q.
    Yan, Z.
    Jiang, N.
    Xu, T.
    Lyu, Z.
    Ma, T.
    RADIATION DETECTION TECHNOLOGY AND METHODS, 2025,
  • [39] Three-dimensional point-source localization method based on position-sensitive neutron detectors
    Liang Q.
    Li J.
    Liu G.
    Wang B.
    Zhou H.
    Tian Y.
    He Jishu/Nuclear Techniques, 2024, 47 (03):
  • [40] Analysis of detector response using 3-D position-sensitive CZT gamma-ray spectrometers
    Zhang, F
    He, Z
    Xu, D
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (06) : 3098 - 3104