Phantom evaluation of a navigation system for out-of-plane CT-guided puncture

被引:20
|
作者
Moncharmont, L. [1 ]
Moreau-Gaudry, A. [2 ]
Medici, M. [2 ]
Bricault, I. [1 ,3 ]
机构
[1] CHU Grenoble, Hop Nord, Serv Radiol & Imagerie Med, F-38043 Grenoble 9, France
[2] CHU Grenoble, Ctr Invest Clin, Unite Innovat Technol, F-38043 Grenoble 9, France
[3] Univ Grenoble Alpes, TIMC IMAG, F-38000 Grenoble, France
关键词
Computed tomography; Phantom study; Electromagnetic navigation; Interventional radiology; ELECTROMAGNETIC TRACKING; RADIOFREQUENCY ABLATION; UPPER ABDOMEN; BIOPSY; GUIDANCE; LUNG; INTERVENTIONS; FEEDBACK; FUSION;
D O I
10.1016/j.diii.2015.03.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The purpose of this phantom study was to assess a new real time electromagnetically-guided navigation system and compare it to standard computed tomography (CT) guidance. Material and methods: A prospective, randomized, comparative study was carried out over a two-day period. Operators without prior experience on the new navigation system sequentially attempted to puncture two 6 mm-diameter targets (one attempt for each target) with out-of-plane trajectories using both the standard CT-guided method and the new navigation station (NAV method). Results: Intention-to-treat analysis was performed for 54 operators. Twenty-two operators out of 54 (40.7%) reached the target on first attempt with the NAV method versus none (0%) using CT-guidance (P < 0.001). The median distance of the puncture from the center of the target was 3.7 mm [Q1-Q3 = 2-6.7] using NAV versus 15 mm [10-20] using CT-guidance (P < 0.001). Overall planning and puncture time were shorter using NAV: 76 s [50-118] versus 214 s [181-264] using CT-guidance (P < 0.001). Conclusion: Novice operators consistently performed faster and more accurate phantom punctures with out-of-plane trajectories using the electromagnetically-guided navigation system than with the standard CT-guided method. (C) 2015 Editions francaises radiologie. Published by Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:531 / 536
页数:6
相关论文
共 50 条
  • [1] Use of an Electromagnetic Navigation System on a Phantom as a Training Simulator for CT-Guided Procedures
    Sanchez, Yadiel
    Trifanov, Dmitly S.
    Kattapuram, Taj M.
    Tao, Haiyang
    Prabhakar, Anand M.
    Arellano, Ronald S.
    Uppot, Raul N.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2017, 14 (06) : 795 - 799
  • [2] CT-guided puncture
    Noldge, G
    Richter, GM
    Grenacher, L
    Brado, M
    Kauffmann, GW
    [J]. RADIOLOGE, 1996, 36 (09): : 683 - 691
  • [3] Robotic CT-guided out-of-plane needle insertion: comparison of angle accuracy with manual insertion in phantom and measurement of distance accuracy in animals
    Komaki, Toshiyuki
    Hiraki, Takao
    Kamegawa, Tetsushi
    Matsuno, Takayuki
    Sakurai, Jun
    Matsuura, Ryutaro
    Yamaguchi, Takuya
    Sasaki, Takanori
    Mitsuhashi, Toshiharu
    Okamoto, Soichiro
    Uka, Mayu
    Matsui, Yusuke
    Iguchi, Toshihiro
    Gobara, Hideo
    Kanazawa, Susumu
    [J]. EUROPEAN RADIOLOGY, 2020, 30 (03) : 1342 - 1349
  • [4] Robotic CT-guided out-of-plane needle insertion: comparison of angle accuracy with manual insertion in phantom and measurement of distance accuracy in animals
    Toshiyuki Komaki
    Takao Hiraki
    Tetsushi Kamegawa
    Takayuki Matsuno
    Jun Sakurai
    Ryutaro Matsuura
    Takuya Yamaguchi
    Takanori Sasaki
    Toshiharu Mitsuhashi
    Soichiro Okamoto
    Mayu Uka
    Yusuke Matsui
    Toshihiro Iguchi
    Hideo Gobara
    Susumu Kanazawa
    [J]. European Radiology, 2020, 30 : 1342 - 1349
  • [5] Augmented Reality-Assisted CT-Guided Puncture: A Phantom Study
    Vincent Van den Bosch
    Hizirwan Shukri Salim
    Njin-Zu Chen
    Otto Stroosma
    Philipp Bruners
    Christiane K. Kuhl
    Federico Pedersoli
    Peter Isfort
    [J]. CardioVascular and Interventional Radiology, 2022, 45 : 1173 - 1177
  • [6] Augmented Reality-Assisted CT-Guided Puncture: A Phantom Study
    Van den Bosch, Vincent
    Salim, Hizirwan Shukri
    Chen, Njin-Zu
    Stroosma, Otto
    Bruners, Philipp
    Kuhl, Christiane K.
    Pedersoli, Federico
    Isfort, Peter
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2022, 45 (08) : 1173 - 1177
  • [7] Intraoperative radiotherapy:: a CT-guided navigation system
    Tonus, C
    Strassmann, G
    Debertshäuser, D
    Kolotas, C
    Walter, S
    Zamboglou, N
    Nier, H
    [J]. CHIRURG, 2001, 72 (06): : 731 - 735
  • [8] CT-guided biopsies and interventions: Evaluation of a magnetic field navigation system
    Schoepf, UJ
    Helmberger, TK
    Holzknecht, N
    Ezekiel, G
    Segalescu, V
    Reiser, MF
    [J]. RADIOLOGY, 2001, 218 (02) : 615 - 616
  • [9] Evaluation of robotic-assisted navigation system for CT-guided thoracic and abdominal lesion puncture: A prospective clinical study
    Jing, Yifan
    Zhang, Jian
    Jin, Yong
    Bai, Xuming
    [J]. JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2024, 20 (04) : 1350 - 1356
  • [10] Intervention Planning Using a Laser Navigation System for CT-Guided Interventions: A Phantom and Patient Study
    Gruber-Rouh, Tatjana
    Lee, Clara
    Bolck, Jan
    Naguib, Nagy N. N.
    Schulz, Boris
    Eichler, Katrin
    Aschenbach, Rene
    Wichmann, Julian L.
    Vogl, Thomas. J.
    Zangos, Stephan
    [J]. KOREAN JOURNAL OF RADIOLOGY, 2015, 16 (04) : 729 - 735