Phantom evaluation of a navigation system for out-of-plane CT-guided puncture

被引:20
|
作者
Moncharmont, L. [1 ]
Moreau-Gaudry, A. [2 ]
Medici, M. [2 ]
Bricault, I. [1 ,3 ]
机构
[1] CHU Grenoble, Hop Nord, Serv Radiol & Imagerie Med, F-38043 Grenoble 9, France
[2] CHU Grenoble, Ctr Invest Clin, Unite Innovat Technol, F-38043 Grenoble 9, France
[3] Univ Grenoble Alpes, TIMC IMAG, F-38000 Grenoble, France
关键词
Computed tomography; Phantom study; Electromagnetic navigation; Interventional radiology; ELECTROMAGNETIC TRACKING; RADIOFREQUENCY ABLATION; UPPER ABDOMEN; BIOPSY; GUIDANCE; LUNG; INTERVENTIONS; FEEDBACK; FUSION;
D O I
10.1016/j.diii.2015.03.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The purpose of this phantom study was to assess a new real time electromagnetically-guided navigation system and compare it to standard computed tomography (CT) guidance. Material and methods: A prospective, randomized, comparative study was carried out over a two-day period. Operators without prior experience on the new navigation system sequentially attempted to puncture two 6 mm-diameter targets (one attempt for each target) with out-of-plane trajectories using both the standard CT-guided method and the new navigation station (NAV method). Results: Intention-to-treat analysis was performed for 54 operators. Twenty-two operators out of 54 (40.7%) reached the target on first attempt with the NAV method versus none (0%) using CT-guidance (P < 0.001). The median distance of the puncture from the center of the target was 3.7 mm [Q1-Q3 = 2-6.7] using NAV versus 15 mm [10-20] using CT-guidance (P < 0.001). Overall planning and puncture time were shorter using NAV: 76 s [50-118] versus 214 s [181-264] using CT-guidance (P < 0.001). Conclusion: Novice operators consistently performed faster and more accurate phantom punctures with out-of-plane trajectories using the electromagnetically-guided navigation system than with the standard CT-guided method. (C) 2015 Editions francaises radiologie. Published by Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:531 / 536
页数:6
相关论文
共 50 条
  • [21] Electromagnetic Navigation System for CT-Guided Biopsy of Small Lesions
    Appelbaum, Liat
    Sosna, Jacob
    Nissenbaum, Yizhak
    Benshtein, Alexander
    Goldberg, S. Nahum
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 196 (05) : 1194 - 1200
  • [22] CT-guided Navigation in Shoulder Arthroplasty
    Wanner, John P.
    Maslow, Jed
    Byram, Ian
    [J]. TECHNIQUES IN SHOULDER AND ELBOW SURGERY, 2019, 20 (02): : 68 - 73
  • [23] Augmented reality visualization for CT-guided interventions: Feasibility and phantom evaluation
    Das, M
    Sauer, F
    Schoepf, UJ
    Khamene, A
    Vogt, SK
    Silverman, SG
    [J]. RADIOLOGY, 2002, 225 : 483 - 483
  • [24] Puncture Cube Patient-Mounted Navigation System versus Freehand Method for CT-Guided Needle Placement: Study on a Neoprene Covered Elliptical Cylinder Gelatin Phantom
    Wessendorf, Joel
    Scheschenja, Michael
    Bastian, Moritz B.
    Gjini, Tefta
    Viniol, Simon
    Owczarek, Andreas
    Koenig, Alexander M.
    Jedelska, Jarmila
    Mahnken, Andreas H.
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2024,
  • [25] CT-guided Abdominal Biopsy Training Phantom
    Taylor, Austin
    Pham, Tony
    Zhao, Zhuo
    Montayre, Rommel
    Xu, Sheng
    Wood, Bradford
    Tse, Zion
    [J]. 2018 INTERNATIONAL SYMPOSIUM ON MEDICAL ROBOTICS (ISMR), 2018,
  • [26] Out-of-plane computed-tomography-guided biopsy using a magnetic-field-based navigation system
    Wallace, MJ
    Gupta, S
    Hicks, ME
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 29 (01) : 108 - 113
  • [27] Out-of-Plane Computed-Tomography-Guided Biopsy Using a Magnetic-Field-Based Navigation System
    Michael J. Wallace
    Sanjay Gupta
    Marshall E. Hicks
    [J]. CardioVascular and Interventional Radiology, 2006, 29 : 108 - 113
  • [28] Augmented reality visualization for CT-guided interventions: System description, feasibility, and initial evaluation in an abdominal phantom
    Das, Marco
    Sauer, Frank
    Schoepf, U. Joseph
    Khamene, Ali
    Vogt, Sebastian K.
    Schaller, Stefan
    Kikinis, Ron
    van Sonnenberg, Eric
    Silverman, Stuart G.
    [J]. RADIOLOGY, 2006, 240 (01) : 230 - 235
  • [29] Augmented reality system for CT-guided interventions: System description and initial phantom trials
    Sauer, F
    Schoepf, UJ
    Khamene, A
    Vogt, S
    Das, M
    Silverman, SG
    [J]. MEDICAL IMAGING 2003: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND DISPLAY, 2003, 5029 : 384 - 394
  • [30] AngleNav: MEMS Tracker to Facilitate CT-Guided Puncture
    Li, Rui
    Xu, Sheng
    Pritchard, William F.
    Karanian, John W.
    Krishnasamy, Venkatesh P.
    Wood, Bradford J.
    Tse, Zion Tsz Ho
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2018, 46 (03) : 452 - 463