Quantitative measurement and real-time tracking of high intensity focused ultrasound using phase-sensitive optical coherence tomography: Feasibility study

被引:2
|
作者
Le, Nhan [1 ]
Song, ShaoZhen [2 ]
Nabi, Ghulam [3 ]
Wang, Ruikang [4 ]
Huang, Zhihong [1 ]
机构
[1] Univ Dundee, Dept Mech Engn, Dundee, Scotland
[2] Univ Washington, Biophoton & Imaging Lab, Seattle, WA 98195 USA
[3] Univ Dundee, Div Canc Res, Dundee, Scotland
[4] Univ Washington, Dept Bioengn, Seattle, WA 98195 USA
关键词
Acoustic radiation force impulse; high intensity focused ultrasound; shear-wave imaging; ultrasound;
D O I
10.1080/02656736.2016.1190036
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Phase-sensitive optical coherence tomography (PhS-OCT) is proposed, as a new high intensity focused ultrasound (HIFU) imaging guidance to detect and track HIFU focus inside 1% agar samples in this work. The experiments studied the effect of varying HIFU power on the induction of shear wave, which can be implemented as a new technique to monitor focused ultrasound surgery (FUS). Method: A miniature HIFU transducer (1.02 MHz, 20 mm aperture diameter, 15 mm radius of curvature) was produced in-house, pressure-field mapped, and calibrated. The transducer was then embedded inside a 1% agar phantom, which was placed under PhS-OCT for observation, under various HIFU power settings (acoustic power, and number of cycles per pulse). Shear wave was induced on the sample surface by HIFU and was captured in full under PhS-OCT. The lowest HIFU acoustic power output for the detection of shear wave was found to be 0.36 W (1.02MHz, 100 cycles/pulse), or with the number of cycles/pulse as low as 20 (1.02 MHz, 0.98 W acoustic power output). Results: A linear relationship between acoustic power output and the maximum shear wave displacement was found in the first study. The second study explores a non-linear correlation between the (HIFU) numbers of cycles per pulse, and the maximum shear wave displacement. Conclusion: PhS-OCT demonstrates excellent tracking and detection of HIFU-induced shear wave. The results could benefit other imaging techniques in tracking and guiding HIFU focus. Further studies will explore the relationship between the physical transducer characteristics and the HIFU-induced shear wave.
引用
下载
收藏
页码:713 / 722
页数:10
相关论文
共 50 条
  • [31] Real-Time Position and Speed Monitoring of Trains Using Phase-Sensitive OTDR
    Peng, Fei
    Duan, Ning
    Rao, Yun-Jiang
    Li, Jin
    IEEE PHOTONICS TECHNOLOGY LETTERS, 2014, 26 (20) : 2055 - 2057
  • [32] Real-time distributed optical fiber vibration sensing based on phase-sensitive OTDR
    Liu, Xin
    Wang, Yu
    Zhang, Jianguo
    Wang, Dong
    Wang, Pengfei
    Bai, Qing
    Jin, Baoquan
    FIBER OPTIC SENSING AND OPTICAL COMMUNICATION, 2018, 10849
  • [33] Measurement and visualization of stimulus-evoked tissue dynamics in mouse barrel cortex using phase-sensitive optical coherence tomography
    Tang, Peijun
    Li, Yuandong
    Rakymzhan, Adiya
    Xie, Zhiying
    Wang, Ruikang K.
    BIOMEDICAL OPTICS EXPRESS, 2020, 11 (02): : 699 - 710
  • [34] Measurement of in vivo basal-turn vibrations of the organ of Corti using phase-sensitive Fourier domain optical coherence tomography
    Ramamoorthy, Sripriya
    Zhang, Yuan
    Petrie, Tracy
    Chen, Fangyi
    Subhash, Hrebesh Molly
    Choudhury, Niloy
    Wang, Ruikang
    Jacques, Steven L.
    Nuttall, Alfred L.
    PHOTONIC THERAPEUTICS AND DIAGNOSTICS IX, 2013, 8565
  • [35] Pupil Tracking for Real-Time Motion Corrected Anterior Segment Optical Coherence Tomography
    Carrasco-Zevallos, Oscar M.
    Nankivil, Derek
    Viehland, Christian
    Keller, Brenton
    Izatt, Joseph A.
    PLOS ONE, 2016, 11 (08):
  • [36] Real-Time Monitoring and Quantitative Evaluation of Cavitation Bubbles Induced by High Intensity Focused Ultrasound Using B-Mode Imaging
    Yu Jie
    Chen Chu-Yi
    Chen Gong
    Guo Xia-Sheng
    Ma Yong
    Tu Juan
    Zhang Dong
    CHINESE PHYSICS LETTERS, 2014, 31 (03)
  • [37] Real-time observation of coronary artery perforation by intravascular ultrasound and optical coherence tomography
    Watabe, Hiroaki
    Sato, Akira
    Usami, Kyohei
    Hiraya, Daigo
    Hoshi, Tomoya
    Ieda, Masaki
    CORONARY ARTERY DISEASE, 2022, 31 (01) : E5 - E6
  • [38] Real-time monitoring of high-intensity focused ultrasound ablations with photoacoustic technique: An in vitro study
    Cui, Huizhong
    Yang, Xinmai
    MEDICAL PHYSICS, 2011, 38 (10) : 5345 - 5350
  • [39] High Resolution Phase-Sensitive Magnetomotive Optical Coherence Microscopy for Tracking Magnetic Microbeads and Cellular Mechanics
    Crecea, Vasilica
    Graf, Benedikt W.
    Kim, Taewoo
    Popescu, Gabriel
    Boppart, Stephen A.
    IEEE JOURNAL OF SELECTED TOPICS IN QUANTUM ELECTRONICS, 2014, 20 (02)
  • [40] Real-time monitoring of focused ultrasound therapy using intelligence-based thermography: A feasibility study
    Sadeghi-Goughari, Moslem
    Han, Sang-Wook
    Kwon, Hyock-Ju
    ULTRASONICS, 2023, 134