Real-time monitoring of high-intensity focused ultrasound ablations with photoacoustic technique: An in vitro study

被引:24
|
作者
Cui, Huizhong [1 ]
Yang, Xinmai [1 ]
机构
[1] Univ Kansas, Dept Mech Engn, KU Bioengn Res Ctr, Lawrence, KS 66045 USA
关键词
high intensity focused ultrasound (HIFU); photoacoustic imaging (PAI); real time treatment monitoring; thermal dose; temperature; PROSTATE-CANCER; TEMPERATURE ESTIMATION; DIAGNOSTIC ULTRASOUND; COMPUTED-TOMOGRAPHY; BIOLOGICAL TISSUES; LOCAL RECURRENCE; LIVER-TUMORS; THERAPY; SURGERY; HYPERTHERMIA;
D O I
10.1118/1.3638126
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study is to apply photoacoustic imaging (PAI) to monitor high intensity focused ultrasound (HIFU) process in real-time for accurate evaluation of thermal ablation outcomes. Methods: A photoacoustic (PA) imaging system was used to monitor HIFU ablation process in this study. Single-element, spherically focused ultrasonic transducers with center frequencies of 5 and 10 MHz were used to generate HIFU lesions and detect PA signals in beef kidney during HIFU treatments, respectively. A 25-mu m diameter, T-type thermocouple was used to measure the temperature rise during the treatment as well. Thermal dose, which was used to indicate the coagulation of soft tissue, was calculated with the temperature measured by the thermocouple. Detected PA signals were then related to the coagulation of soft tissue through thermal dose calculations. In addition, PA signals from beef kidney coagulating under a constant temperature was obtained to show the changes of PA signals under a constant temperature during soft tissue coagulation. Results: PA amplitude increases during the HIFU process and tends to have a saturation stage after the soft tissue is completely coagulated, which is indicated by a 240 or more TD43 minutes. Also, a linear relation between the PA signal amplitude changes and temperature was observed when monitoring the sole effect of temperature. In addition, PA amplitude changes finally reached a constant value during soft tissue coagulation under a constant temperature. Conclusions: Through measurement of the photoacoustic signal, the authors demonstrated that monitoring thermal dose may be a more appropriate method in HIFU treatment than monitoring temperature. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3638126]
引用
收藏
页码:5345 / 5350
页数:6
相关论文
共 50 条
  • [1] Real-Time Photoacoustic Thermometry Combined With Clinical Ultrasound Imaging and High-Intensity Focused Ultrasound
    Kim, Jeesu
    Choi, Wonseok
    Park, Eun-Yeong
    Kang, Youngnam
    Lee, Ki Jong
    Kim, Hyung Ham
    Kim, Won Jong
    Kim, Chulhong
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2019, 66 (12) : 3330 - 3338
  • [2] Fluorescence imaging for real-time monitoring of high-intensity focused ultrasound cardiac ablation
    Deng, CRX
    Qu, FJ
    Nikolski, VP
    Zhou, Y
    Efimov, IR
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2005, 33 (10) : 1352 - 1359
  • [3] Fluorescence Imaging for Real-Time Monitoring of High-Intensity Focused Ultrasound Cardiac Ablation
    Cheri X. Deng
    Fujian Qu
    Vladimir P. Nikolski
    Yun Zhou
    Igor R. Efimov
    [J]. Annals of Biomedical Engineering, 2005, 33 : 1352 - 1359
  • [4] Array-based high-intensity focused ultrasound therapy system integrated with real-time ultrasound and photoacoustic imaging
    Zhang, Yachao
    Wang, Lidai
    [J]. BIOMEDICAL OPTICS EXPRESS, 2023, 14 (03) : 1137 - 1145
  • [5] Real-time visualization of high-intensity focused ultrasound treatment using ultrasound imaging
    Vaezy, S
    Shi, XG
    Martin, RW
    Chi, E
    Nelson, PI
    Bailey, MR
    Crum, LA
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (01): : 33 - 42
  • [6] REAL-TIME MONITORING OF HIGH-INTENSITY FOCUSED ULTRASOUND THERMAL THERAPY USING THE MANIFOLD LEARNING METHOD
    Rangraz, Parisa
    Behnam, Hamid
    Sobhebidari, Pooya
    Tavakkoli, Jahan
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2014, 40 (12): : 2841 - 2850
  • [7] An experimental study of real-time monitoring of acoustic output in high-intensity focused ultrasound using a PVDF piezoelectric sensor
    Ye, Fangwei
    Lin, Tao
    Fu, Yinshu
    Qi, Jiajun
    Ma, Yongsong
    Xu, Guihua
    Wang, Zhilong
    Wu, Feng
    Chen, Wenzhi
    Wang, Zhibiao
    [J]. THERAPEUTIC ULTRASOUND, 2006, 829 : 604 - +
  • [8] Integration of photoacoustic imaging and high-intensity focused ultrasound
    Cui, Huizhong
    Staley, Jacob
    Yang, Xinmai
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2010, 15 (02)
  • [9] REAL-TIME MONITORING OF HIGH-INTENSITY FOCUSED ULTRASOUND LESION FORMATION USING ACOUSTO-OPTIC SENSING
    Lai, Puxiang
    McLaughlan, James R.
    Draudt, Andrew B.
    Murray, Todd W.
    Cleveland, Robin O.
    Roy, Ronald A.
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2011, 37 (02): : 239 - 252
  • [10] Optimized Schlieren imaging for real-time visualization of high-intensity focused ultrasound waves
    Nawrath, Helena
    Holl, Andreas
    Bedarf, Lennart
    Mai, Denis
    Alfke, Heiko
    Groebner, Jens
    [J]. BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2024,