Development of a new HPLC method for rapid histamine quantification in fish and fishery products without sample clean-up

被引:11
|
作者
Kounnoun, Ayoub [1 ]
Louajri, Adnane [1 ]
Cacciola, Francesco [2 ]
El Baaboua, Aicha [1 ]
Mondello, Luigi [3 ,4 ,5 ]
Bougtaib, Hajar [1 ]
Alahlah, Naoual [6 ]
Stitou, Meriem [6 ]
El Maadoudi, Mohamed [6 ]
机构
[1] Abd Al Malek Essaadi Univ, Fac Sci Tetouan, Dept Biol, Lab Appl Biol & Pathol, Tetouan, Morocco
[2] Univ Messina, Dept Biomed Dent Morphol & Funct Imaging Sci, Messina, Italy
[3] Univ Messina, Dept Chem Biol Pharmaceut & Environm Sci, Messina, Italy
[4] Univ Messina, Chromaleont Srl, Dept Chem Biol Pharmaceut & Environm Sci, Messina, Italy
[5] Univ Campus Biomed Rome, Dept Sci & Technol Human & Environm, Rome, Italy
[6] Natl Off Food Safety ONSSA, Reg Anal & Res Lab, Tangier, Morocco
关键词
Histamine; Fish; HPLC-FLD; Pre-column derivatization; Optimization; Validation; ORTHO-PHTHALALDEHYDE; LIQUID-CHROMATOGRAPHY; FLUORESCENCE REACTION; BIOGENIC-AMINES; DERIVATIZATION; DERIVATIVES; VALIDATION; STABILITY; FRESH; ACIDS;
D O I
10.1007/s00217-022-03995-z
中图分类号
TS2 [食品工业];
学科分类号
0832 ;
摘要
HPLC methods are among the most studied and validated techniques for the quantification of histamine in fish and fishery products; nonetheless, most of these methods require laborious sample extraction and purification. For this purpose, in the present study, an HPLC-FLD method without any sample clean-up procedure was developed, validated and applied for histamine quantification in various fish samples. Histamine was extracted from fish and fishery products samples with perchloric acid; then a simple precolumn derivatization with o-phthalaldehyde was performed in the presence of the reducing agent 2-mercaptoethanol. Once the derivatization was carried out, the separation was performed with a reverse phase HPLC system. In terms of validation, good linearity and specificity over a concentration range from 12.5 to 200 mg/kg were achieved. Also, convenable precision with relative standard deviations smaller than 5.92%, accuracy ranged from 0.40 to 1.84%, and satisfactory recoveries ranged from 100.44 to 101.84%, were obtained. The detection and the quantification limits were calculated at 2.8 and 4.3 mg/kg, respectively. In addition, the effectiveness of the method was evaluated by the analysis of reference materials, proficiency tests and various type of samples. Moreover, the developed method was found to be simple, rapid, and reliable.
引用
收藏
页码:1679 / 1689
页数:11
相关论文
共 50 条
  • [31] Development and validation of an HPLC-FLD method for rapid determination of histamine in skipjack tuna fish (Katsuwonus pelamis)
    Tahmouzi, Saeed
    Khaksar, Ramin
    Ghasemlou, Mehran
    FOOD CHEMISTRY, 2011, 126 (02) : 756 - 761
  • [32] Development and evaluation of a semi-automatic single-step clean-up apparatus for rapid analysis of 18 antibiotics in fish samples
    Mu, Shuhe
    Teng, Tengwang
    Zhou, Ruidong
    Liu, Huan
    Sun, Huiwu
    Li, Jincheng
    JOURNAL OF FOOD COMPOSITION AND ANALYSIS, 2022, 113
  • [33] HPLC-based method using sample pre-column clean-up for the determination of methanethiol and ethanethiol in parenteral amino acid solutions
    do Nascimento, PC
    Bohrer, D
    Rohlfes, ALB
    de Carvalho, LM
    Ramirez, A
    FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2001, 370 (01): : 92 - 96
  • [34] HPLC-based method using sample pre-column clean-up for the determination of methanethiol and ethanethiol in parenteral amino acid solutions
    P. C. do Nascimento
    Denise Bohrer
    A. L. B. Rohlfes
    Leandro M. de Carvalho
    Adrain Ramirez
    Fresenius' Journal of Analytical Chemistry, 2001, 370 : 92 - 96
  • [35] Validation of standard method EN ISO 19343 for the detection and quantification of histamine in fish and fishery products using high-performance liquid chromatography
    Duflos, Guillaume
    Inglebert, Gaelle
    Himber, Charlotte
    Degremont, Sylvie
    Lombard, Bertrand
    Brisabois, Anne
    INTERNATIONAL JOURNAL OF FOOD MICROBIOLOGY, 2019, 288 : 97 - 101
  • [36] A rapid and sensitive high performance liquid chromatographic assay of the new antimalarial compound 80/53 in serum with a novel sample clean-up method and its pharmacokinetics in rabbits
    Paliwal, JK
    Gupta, RC
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1998, 17 (4-5) : 775 - 783
  • [37] Development and validation of quantitative thin layer chromatographic technique for determination of total aflatoxins in poultry feed and food grains without sample clean-up
    Salisu, Bahauddeen
    Anua, Siti Marwanis
    Ishak, Wan Rosli Wan
    Mazlan, Nurzafirah
    JOURNAL OF ADVANCED VETERINARY AND ANIMAL RESEARCH, 2021, 8 (04) : 656 - 670
  • [38] Development and validation of an UHPLC-qTOF-MS method for the quantification of cyclic polyesters oligomers in pasta by applying a modified QuEChERS clean-up
    Tsochatzis, Emmanouil D.
    Lopes, Joao Alberto
    Gika, Helen
    Dalsgaard, Trine Kastrup
    Theodoridis, Georgios
    FOOD CHEMISTRY, 2021, 347
  • [39] Determining aflatoxins in raw peanuts using immunoaffinity column as sample clean-up method followed by normal-phase HPLC-FLD analysis
    Shen, Ming-Hsun
    Singh, Rakesh K.
    FOOD CONTROL, 2022, 139
  • [40] Development and validation of an HPLC-DAD method for Rapid quantification of vasicine in Adhatoda vasica leaves and commercial products
    Velichkova, S.
    Theunis, M.
    Naessens, T.
    Pieters, L.
    Foubert, K.