Dispersed Hydrogel Actuator for Modiolar Hugging Cochlear Implant Electrode Arrays

被引:4
|
作者
Stieghorst, J. [1 ,2 ]
Doll, T. [1 ,2 ]
机构
[1] Hannover Med Sch, Dept Otolaryngol, D-30625 Hannover, Germany
[2] Hearing4all, Cluster Excellence, D-30625 Hannover, Germany
关键词
Cochlear implants (CIs); electrode carrier; hydrogel actuation; modiolar coiling electrode; modiolar hugging electrode; FILLER; BIOCOMPATIBILITY;
D O I
10.1109/TBME.2016.2524663
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective: Cochlear implants are used in patients with profound deafness due to degeneration of hair cells inside the cochlea. As improvement for commercially available CIs with straight electrode arrays, it is desired that the electrode array hugs toward the nerve cells located at the central axis of the cochlea in order to reduce the stimulation distance. Therefore, we present a hydrogel-based actuated electrode shaft, which should bend itself when exposed to saline solution (simulating the intracochlear liquid perilymph). Methods: In vitro tests with the electrode arrays were performed in a cochlea model. Different quantities and grain sizes of the hydrogel-building polymer (polyacrylamide) were used to study the self-bending effect. Furthermore, material interfaces between the components were evaluated with scanning electron microscopy, fluorescence microscopy, and an adapted tape test according to DIN EN ISO 2409. Results: Self-bending of the electrode array was observed in the cochlea model and a maximum number of 3.1 turns was achieved. No delamination between the components could be observed. Conclusion: After insertion, we expect that the electrode shaft moves into a patient individual perimodiolar position without losing its functionality because of a delamination of the components. Significance: A modiolar hugging electrode design is presented, which may improve hearing restauration with cochlear implants due to an easier insertion technique and a patient individualized hugging to the modiolus.
引用
收藏
页码:2294 / 2300
页数:7
相关论文
共 50 条
  • [1] Hydrogel-Based Actuation for Modiolar Hugging Cochlear Implant Electrode Arrays
    Stieghorsta, J.
    Tran, B. N.
    Hadeler, S.
    Beckmann, D.
    Doll, T.
    PROCEEDINGS OF THE 30TH ANNIVERSARY EUROSENSORS CONFERENCE - EUROSENSORS 2016, 2016, 168 : 1529 - 1532
  • [2] Speech Recognition Outcomes in Adults With Slim Straight and Slim Modiolar Cochlear Implant Electrode Arrays
    MacPhail, Margaret E.
    Connell, Nathan T.
    Totten, Douglas J.
    Gray, Mitchell T.
    Pisoni, David
    Yates, Charles W.
    Nelson, Rick F.
    OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2022, 166 (05) : 943 - 950
  • [3] Simpler and effective radiological evaluations for modiolar proximity of a slim modiolar cochlear implant electrode
    Lee, Sang-Yeon
    Han, Jin Hee
    Carandang, Marge
    Bae, Yun Jung
    Choi, Byung Yoon
    SCIENTIFIC REPORTS, 2020, 10 (01)
  • [4] Simpler and effective radiological evaluations for modiolar proximity of a slim modiolar cochlear implant electrode
    Sang-Yeon Lee
    Jin Hee Han
    Marge Carandang
    Yun Jung Bae
    Byung Yoon Choi
    Scientific Reports, 10
  • [5] Medical-Grade Silicone Rubber-Hydrogel-Composites for Modiolar Hugging Cochlear Implants
    Yilmaz-Bayraktar, Suheda
    Foremny, Katharina
    Kreienmeyer, Michaela
    Warnecke, Athanasia
    Doll, Theodor
    POLYMERS, 2022, 14 (09)
  • [6] Early Outcomes With a Slim, Modiolar Cochlear Implant Electrode Array
    McJunkin, Jonathan L.
    Durakovic, Nedim
    Herzog, Jacques
    Buchman, Craig A.
    OTOLOGY & NEUROTOLOGY, 2018, 39 (01) : E28 - E33
  • [7] Psychophysics of a prototype peri-modiolar cochlear implant electrode array
    Cohen, LT
    Saunders, E
    Clark, GM
    HEARING RESEARCH, 2001, 155 (1-2) : 63 - 81
  • [8] Slim, Modiolar Cochlear Implant Electrode: Melbourne Experience and Comparison With the Contour Perimodiolar Electrode
    Shaul, Chanan
    Weder, Stefan
    Tari, Sylvia
    Gerard, Jean-Marc
    O'Leary, Stephen J.
    Briggs, Robert J.
    OTOLOGY & NEUROTOLOGY, 2020, 41 (05) : 639 - 643
  • [9] The association between electrode impedance and short-term outcomes in cochlear implant recipients of slim modiolar and slim straight electrode arrays
    Mussoi, Bruna S.
    Meibos, Alex
    Woodson, Erika
    Sydlowski, Sarah
    COCHLEAR IMPLANTS INTERNATIONAL, 2024, 25 (01) : 59 - 68
  • [10] Preservation of residual hearing after cochlear implant surgery with slim modiolar electrode
    Matti Iso-Mustajärvi
    Sini Sipari
    Heikki Löppönen
    Aarno Dietz
    European Archives of Oto-Rhino-Laryngology, 2020, 277 : 367 - 375