Fabrication of a polystyrene microfluidic chip coupled to electrospray ionization mass spectrometry for protein analysis

被引:17
|
作者
Hu, Xianqiao [1 ,2 ,3 ]
Dong, Yuanyuan [2 ]
He, Qiaohong [2 ]
Chen, Hengwu [2 ]
Zhu, Zhiwei [1 ,3 ]
机构
[1] China Natl Rice Res Inst, Hangzhou 310006, Zhejiang, Peoples R China
[2] Zhejiang Univ, Dept Chem, Inst Microanalyt Syst, Hangzhou 310058, Zhejiang, Peoples R China
[3] Minist Agr, Lab Qual & Safety Risk Assessment Rice Hangzhou, Hangzhou, Zhejiang, Peoples R China
关键词
Microfluidic chip; Electrospray ionization-mass spectrometry; Polystyrene; Immobilized enzymatic reactor; Gold film electric contact; Interface; CAPILLARY-ELECTROPHORESIS; EMITTER; SEPARATION; PROTEOMICS; MICROCHIP; DIGESTION; TIP; MS;
D O I
10.1016/j.jchromb.2015.03.013
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A highly integrated polystyrene (PS) microfluidic chip coupled to electrospray ionization mass spectrometry for on-chip protein digestion and online analysis was developed. The immobilized enzymatic microreactor for on-chip protein digestion was integrated onto microchip via the novel method of region-selective UV-modification combined with glutaraldehyde-based immobilization. The micro film electric contact for applying high voltage was prepared on chips by using UV-directed electroless plating technique. A micro-tip was machined at the end of main channel, serving as the interface between microchip and mass spectrometric detector. On-chip digestion and online detection of protein was carried out by coupling the microchip with mass spectrometry (MS). The influences of methanol flow rate in side channel on the stability of spray and intensity of signals were investigated systematically. Also the influence of sample flow rate on the performance of immobilized enzymatic reactor were investigated. Stable spray was obtained at the spray voltage of 2.8-3.0 kV and the methanol flow rate of 500-700 nL min(-1) with the relative standard deviation (RSD) of total ion current (TIC) less than 10%. The influence of sample flow rate on the performance of immobilized enzymatic reactor was also studied. The sequence coverage of protein identification decreased with the increase of flow rate of the sample solution. A sequence coverage of 96% was obtained with immobilized enzymatic reactor at the sample flow rate of 100 nL min(-1) with the reaction time of 8.4 min. It could detect cytochrome c as low as 10 mu g mL(-1) with the developed system. No obvious decrease in protein digestion efficiency was observed after the chip continuously performed for 4 h and stored for 15 d. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:96 / 103
页数:8
相关论文
共 50 条
  • [1] Design, fabrication and test of a microfluidic nebulizer chip for desorption electrospray ionization mass spectrometry
    Sen, A. K.
    Darabi, J.
    Knapp, D. R.
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2009, 137 (02) : 789 - 796
  • [2] Chip-based microfluidic devices coupled with electrospray ionization-mass spectrometry
    Sung, WC
    Makamba, H
    Chen, SH
    [J]. ELECTROPHORESIS, 2005, 26 (09) : 1783 - 1791
  • [3] A multifunctional microfluidic droplet-array chip for analysis by electrospray ionization mass spectrometry
    Su, Yuan
    Zhu, Ying
    Fang, Qun
    [J]. LAB ON A CHIP, 2013, 13 (10) : 1876 - 1882
  • [4] A decade of microfluidic analysis coupled with electrospray mass spectrometry: An overview
    Koster, Sander
    Verpoorte, Elisabeth
    [J]. LAB ON A CHIP, 2007, 7 (11) : 1394 - 1412
  • [5] Protein analysis by desorption electrospray ionization mass spectrometry
    Venter, Andre R.
    [J]. MASS SPECTROMETRY REVIEWS, 2024,
  • [7] An enhanced microfluidic chip coupled to an electrospray Qstar mass spectrometer for protein identification
    Pinto, DM
    Ning, YB
    Figeys, D
    [J]. ELECTROPHORESIS, 2000, 21 (01) : 181 - 190
  • [8] On-chip microfluidics for proteomic analysis by ElectroSpray Ionization/Mass Spectrometry
    Tabourier, P
    Druon, C
    Rolando, C
    Lefebvre, P
    [J]. 1ST ANNUAL INTERNATIONAL IEEE-EMBS SPECIAL TOPIC CONFERENCE ON MICROTECHNOLOGIES IN MEDICINE & BIOLOGY, PROCEEDINGS, 2000, : 439 - 444
  • [9] Direct Surface and Droplet Microsampling for Electrospray Ionization Mass Spectrometry Analysis with an Integrated Dual-Probe Microfluidic Chip
    Huang, Cong-Min
    Zhu, Ying
    Jin, Di-Qiong
    Kelly, Ryan T.
    Fang, Qun
    [J]. ANALYTICAL CHEMISTRY, 2017, 89 (17) : 9009 - 9016
  • [10] Electrospray ionization mass spectrometry: Protein structure
    Farmer, TB
    Caprioli, RM
    [J]. MASS SPECTROMETRY IN BIOMOLECULAR SCIENCES, 1996, 475 : 61 - 88