On the Use of Optically Stimulated Luminescent Dosimeter for Surface Dose Measurement during Radiotherapy

被引:37
|
作者
Yusof, Fasihah Hanum [1 ]
Ung, Ngie Min [2 ]
Wong, Jeannie Hsiu Ding [1 ,3 ]
Jong, Wei Loong [1 ]
Ath, Vannyat [1 ]
Phua, Vincent Chee Ee [2 ]
Heng, Siew Ping [4 ]
Ng, Kwan Hoong [1 ,3 ]
机构
[1] Univ Malaya, Fac Med, Dept Biomed Imaging, Kuala Lumpur, Malaysia
[2] Univ Malaya, Fac Med, Clin Oncol Unit, Kuala Lumpur, Malaysia
[3] Univ Malaya, Fac Med, Res Imaging Ctr, Kuala Lumpur, Malaysia
[4] Pantai Hosp Kuala Lumpur, Inst Canc, Kuala Lumpur, Malaysia
来源
PLOS ONE | 2015年 / 10卷 / 06期
关键词
BREAST-CANCER; CHEST-WALL; CONTRALATERAL BREAST; 2ND PRIMARY; SKIN; PHOTON; HEAD; FILM; OSL; ASSOCIATION;
D O I
10.1371/journal.pone.0128544
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
This study was carried out to investigate the suitability of using the optically stimulated luminescence dosimeter (OSLD) in measuring surface dose during radiotherapy. The water equivalent depth (WED) of the OSLD was first determined by comparing the surface dose measured using the OSLD with the percentage depth dose at the buildup region measured using a Markus ionization chamber. Surface doses were measured on a solid water phantom using the OSLD and compared against the Markus ionization chamber and Gafchromic EBT3 film measurements. The effect of incident beam angles on surface dose was also studied. The OSLD was subsequently used to measure surface dose during tangential breast radiotherapy treatments in a phantom study and in the clinical measurement of 10 patients. Surface dose to the treated breast or chest wall, and on the contralateral breast were measured. The WED of the OSLD was found to be at 0.4 mm. For surface dose measurement on a solid water phantom, the Markus ionization chamber measured 15.95% for 6 MV photon beam and 12.64% for 10 MV photon beam followed by EBT3 film (23.79% and 17.14%) and OSLD (37.77% and 25.38%). Surface dose increased with the increase of the incident beam angle. For phantom and patient breast surface dose measurement, the response of the OSLD was higher than EBT3 film. The in-vivo measurements were also compared with the treatment planning system predicted dose. The OSLD measured higher dose values compared to dose at the surface (Hp(0.0)) by a factor of 2.37 for 6 MV and 2.01 for 10 MV photon beams, respectively. The measurement of absorbed dose at the skin depth of 0.4 mm by the OSLD can still be a useful tool to assess radiation effects on the skin dermis layer. This knowledge can be used to prevent and manage potential acute skin reaction and late skin toxicity from radiotherapy treatments.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Surface dose measurement by optically stimulated luminescent dosimeter: A phantom study
    Sanklaa, K.
    Pitikiattikul, C.
    Rittiwong, M.
    Thiamthan, N.
    Turathong, S.
    Sanghangthum, T.
    RADIOGRAPHY, 2024, 30 (05) : 1405 - 1410
  • [2] Evaluation of Optically Stimulated Luminescent Dosimeters as An In-Vivo Dosimeter for Breast Surface Dose Measurement
    Harrop, S.
    Cao, F.
    Kohli, K.
    Harvey, E.
    Dubrowski, P.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [3] Optically Stimulated Luminescent Dosimeter Performance in High Dose Rate Brachytherapy
    Tien, C.
    Hiatt, J.
    Curran, B.
    Sternick, E.
    MEDICAL PHYSICS, 2012, 39 (06) : 3775 - 3775
  • [4] An affordable optically stimulated luminescent dosimeter reader
    West, WG
    Kearfott, KJ
    Kalchik, AF
    HEALTH PHYSICS, 2005, 89 (01): : S7 - S8
  • [5] Changes in optically stimulated luminescent dosimeter (OSLD) dosimetric characteristics with accumulated dose
    Jursinic, Paul A.
    MEDICAL PHYSICS, 2010, 37 (01) : 132 - 140
  • [6] Dosimetric measurement of testicular dose for colorectal cancer using optically-stimulated luminescent dosimeters in radiotherapy
    Lu, C. C.
    Wang, F. N.
    Lin, H. H.
    Hsu, C. H.
    Lin, J. P.
    Lai, L. H.
    RADIATION PHYSICS AND CHEMISTRY, 2020, 172
  • [7] Ultra-Violet Annealing of Optically Stimulated Luminescent Dosimeter
    Lam, K.
    Lee, C.
    MEDICAL PHYSICS, 2012, 39 (06) : 3721 - 3721
  • [8] Development and Evaluation of Flexible Dosimeter for Surface Dose Measurement in Radiotherapy
    Kim, J.
    Han, M.
    Moon, Y. M.
    Park, S. K.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S694 - S694
  • [9] Use of Commercially Available Optically Stimulated Luminescence Dosimeter As Extremity Dose Estimator
    Georgeson, David L.
    Christiansen, Byron H.
    HEALTH PHYSICS, 2021, 120 (01): : 86 - 93
  • [10] Multi-channel fibre optic dosimeter based on Optically Stimulated Luminescence for dose verification during radiotherapy treatments
    magne, Sylvain
    Auger, Laureline
    Isambert, Aurelie
    Bridier, Andre
    Ferdinand, Pierre
    Barthe, Jean
    THIRD EUROPEAN WORKSHOP ON OPTICAL FIBRE SENSORS, 2007, 6619