Surface dose measurement by optically stimulated luminescent dosimeter: A phantom study

被引:0
|
作者
Sanklaa, K. [1 ]
Pitikiattikul, C. [1 ]
Rittiwong, M. [1 ]
Thiamthan, N. [1 ]
Turathong, S. [2 ]
Sanghangthum, T. [3 ]
机构
[1] Thammasat Univ, Fac Allied Hlth Sci, Dept Radiol Technol, Bangkok, Thailand
[2] Thammasat Univ Hosp, Dept Radiol, Bangkok, Thailand
[3] Chulalongkorn Univ, Fac Med, Dept Radiol, Bangkok, Thailand
关键词
OSLD; Breast cancer; Radiotherapy; CIRS thorax phantom; Surface dose; Treatment planning system; RADIOTHERAPY;
D O I
10.1016/j.radi.2024.06.011
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: Radiotherapy is the standard treatment for breast cancer patients after surgery. However, radiotherapy can cause side effects such as dry and moist desquamation of the patient's skin. The dose calculation from a treatment planning system (TPS) might also be inaccurate. The purpose of this study is to measure the surface dose on the CIRS thorax phantom by an optically stimulated luminescent dosimeter (OSLD). Methods: The characteristics of OSLD were studied in terms of dose linearity, reproducibility, and angulation dependence on the solid water phantom. To determine the surface dose, OSLD (Landauer lnc., USA) was placed on 5 positions at the CIRS phantom (Tissue Simulation and Phantom Technology, USA). The five positions were at the tip, medial, lateral, tip-medial, and tip-lateral. Then, the doses from OSLD and TPS were compared. Results: The dosimeter's characteristic test was good. The maximum dose at a depth of 15 mm was 514.46 cGy, which was at 100%. The minimum dose at the surface was 174.91 cGy, which was at 34%. The results revealed that the surface dose from TPS was less than the measurement. The percent dose difference was -2.17 +/- 6.34, -12.08 +/- 3.85, and -48.71 +/- 1.29 at the tip, medial, and lateral positions, respectively. The surface dose from TPS at tip-medial and tip-lateral was higher than the measurement, which was 12.56 +/- 5.55 and 10.45 +/- 1.76 percent dose different, respectively. Conclusion: The percent dose difference is within the acceptable limit, except for the lateral position because of the body curvature. However, OSLD is convenient to assess the radiation dose, and further study is to measure in vivo. Implication for practice: The OSL NanoDot dosimeter can be used for dose validation with a constant setup location. The measurement dose is higher than the dose from TPS, except for some tilt angles.
引用
收藏
页码:1405 / 1410
页数:6
相关论文
共 50 条
  • [1] On the Use of Optically Stimulated Luminescent Dosimeter for Surface Dose Measurement during Radiotherapy
    Yusof, Fasihah Hanum
    Ung, Ngie Min
    Wong, Jeannie Hsiu Ding
    Jong, Wei Loong
    Ath, Vannyat
    Phua, Vincent Chee Ee
    Heng, Siew Ping
    Ng, Kwan Hoong
    [J]. PLOS ONE, 2015, 10 (06):
  • [2] Evaluation of Optically Stimulated Luminescent Dosimeters as An In-Vivo Dosimeter for Breast Surface Dose Measurement
    Harrop, S.
    Cao, F.
    Kohli, K.
    Harvey, E.
    Dubrowski, P.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [3] Optically Stimulated Luminescent Dosimeter Performance in High Dose Rate Brachytherapy
    Tien, C.
    Hiatt, J.
    Curran, B.
    Sternick, E.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3775 - 3775
  • [4] An affordable optically stimulated luminescent dosimeter reader
    West, WG
    Kearfott, KJ
    Kalchik, AF
    [J]. HEALTH PHYSICS, 2005, 89 (01): : S7 - S8
  • [5] Changes in optically stimulated luminescent dosimeter (OSLD) dosimetric characteristics with accumulated dose
    Jursinic, Paul A.
    [J]. MEDICAL PHYSICS, 2010, 37 (01) : 132 - 140
  • [6] Phantom Measurement of Surface and Organ Dose Using Optically Stimulated Luminescence Detectors in Diagnostic Radiology
    Rodriguez, M.
    Yukihara, E.
    Ahmad, S.
    Ruan, C.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [7] Ultra-Violet Annealing of Optically Stimulated Luminescent Dosimeter
    Lam, K.
    Lee, C.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3721 - 3721
  • [8] Pacemaker Dose Estimate Using Optically Stimulated Luminescent Dosimeter for Left Breast Intraoperative Radiation Therapy
    Chen, Y.
    Goenka, A.
    Sharma, A.
    Wang, L.
    Cao, Y.
    Jamshidi, A.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3614 - 3614
  • [9] Preliminary study for dose evaluation depending on dose range with optically stimulated luminescence dosimeter considering individual dosimeter sensitivity
    Han, Su Chul
    [J]. PLOS ONE, 2022, 17 (03):
  • [10] Chromatic bleaching and fractionation effects on optically stimulated luminescent dosimeter reuse
    Scott, Hayden
    Alvarez, Paola E.
    Howell, Rebecca M.
    Riegel, Adam
    Sun, Ryan
    Liu, Kevin
    Kry, Stephen F.
    [J]. MEDICAL PHYSICS, 2024,