A customizable aluminum compensator system for total body irradiation

被引:0
|
作者
Naessig, Madison [1 ,2 ]
Hernandez, Soleil [2 ]
Astorga, Nestor Rodrigo [1 ]
McCulloch, James [1 ]
Saenz, Daniel [1 ]
Myers, Pamela [1 ]
Rasmussen, Karl [1 ]
Stathakis, Sotirios [1 ]
Ha, Chul S. [1 ]
Papanikolaou, Niko [1 ]
Ford, John [2 ]
Kirby, Neil [1 ]
机构
[1] Univ Texas Hlth Sci Ctr San Antonio, Dept Radiat Oncol, 7979 Wurzbach Rd,MC 7889, San Antonio, TX 78229 USA
[2] Texas A&M Univ, Dept Nucl Engn, College Stn, TX 77843 USA
来源
关键词
compensator; dose homogeneity; total body irradiation; TOTAL MARROW IRRADIATION; INTERSTITIAL PNEUMONITIS; THERAPY;
D O I
10.1002/acm2.13393
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To develop a simplified aluminum compensator system for total body irradiation (TBI) that is easy to assemble and modify in a short period of time for customized patient treatments. Methods The compensator is composed of a combination of 0.3 cm thick aluminum bars, two aluminum T-tracks, spacers, and metal bolts. The system is mounted onto a plexiglass block tray. The design consists of 11 fixed sectors spanning from the patient's head to feet. The outermost sectors utilize 7.6 cm wide aluminum bars, while the remaining sectors use 2.5 cm wide aluminum bars. There is a magnification factor of 5 from the compensator to the patient treatment plane. Each bar of aluminum is interconnected at each adjacent sector with a tongue and groove arrangement and fastened to the T-track using a metal washer, bolt, and nut. Inter-bar leakage of the compensator was tested using a water tank and diode. End-to-end measurements were performed with an ion chamber in a solid water phantom and also with a RANDO phantom using internal and external optically stimulated luminescent detectors (OSLDs). In-vivo patient measurements from the first 20 patients treated with this aluminum compensator were compared to those from 20 patients treated with our previously used lead compensator system. Results The compensator assembly time was reduced to 20-30 min compared to the 2-4 h it would take with the previous lead design. All end-to-end measurements were within 10% of that expected. The median absolute in-vivo error for the aluminum compensator was 3.7%, with 93.8% of measurements being within 10% of that expected. The median error for the lead compensator system was 5.3%, with 85.1% being within 10% of that expected. Conclusion This design has become the standard compensator at our clinic. It allows for quick assembly and customization along with meeting the Task Group 29 recommendations for dose uniformity.
引用
收藏
页码:36 / 44
页数:9
相关论文
共 50 条
  • [41] Preparing the patient for total body irradiation
    Balsdon, H
    Treeby, J
    Routis, D
    BONE MARROW TRANSPLANTATION, 2001, 27 : S379 - S379
  • [42] TOTAL-BODY IRRADIATION DOSIMETRY
    AUKETT, RJ
    PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (03): : 312 - 312
  • [43] THE RESPONSE OF TISSUE TO TOTAL BODY IRRADIATION
    TULLIS, JL
    AMERICAN JOURNAL OF PATHOLOGY, 1949, 25 (05): : 829 - 851
  • [44] A TOTAL-BODY IRRADIATION STAND
    GLASGOW, GP
    WANG, S
    STANTON, J
    MEDICAL PHYSICS, 1987, 14 (03) : 485 - 485
  • [45] Dosimetric Considerations of Total Body Irradiation
    Maghsoudi, K.
    Morin, O.
    Pouliot, J.
    Chang, J.
    Johnson, J.
    Polishchuk, A.
    Fogh, S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S746 - S746
  • [46] Total body irradiation: role and indications
    Heinzelmann, F
    Ottinger, H
    Mueller, C
    Allgaier, S
    Bamberg, M
    Belka, C
    RADIOTHERAPY AND ONCOLOGY, 2004, 73 : S350 - S350
  • [47] Total body irradiation - With review of cases
    Medinger, FG
    Craver, LF
    AMERICAN JOURNAL OF ROENTGENOLOGY AND RADIUM THERAPY, 1942, 48 : 651 - 671
  • [48] SUSCEPTIBILITY TO TOTAL-BODY IRRADIATION
    VRIESENDORP, HM
    VANBEKKUM, DW
    EXPERIMENTAL HEMATOLOGY, 1978, 6 : 19 - 19
  • [49] The radiobiological basis of total body irradiation
    Wheldon, TE
    BRITISH JOURNAL OF RADIOLOGY, 1997, 70 (840): : 1204 - 1207
  • [50] A Century of Total Body Irradiation (TBI)
    Barker, C. A.
    Rimner, A.
    Yahalom, J.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S434 - S435