Investigation on Effect of Image Lag in Fluoroscopic Images Obtained with a Dynamic Flat-panel Detector (FPD) on Accuracy of Target Tracking in Radiotherapy

被引:12
|
作者
Tanaka, Rie [1 ]
Ichikawa, Katsuhiro [1 ]
Mori, Shinichiro [2 ]
Dobashi, Suguru [2 ]
Kumagai, Motoki [2 ]
Kawashima, Hiroki [3 ]
Minohara, Shinichi [2 ]
Sanada, Sigeru [1 ]
机构
[1] Kanazawa Univ, Coll Med Pharmaceut & Hlth Sci, Sch Hlth Sci, Dept Radiol Technol, Kanazawa, Ishikawa 9200942, Japan
[2] Natl Inst Radiol Sci, Res Ctr Charged Particle Therapy, Inage Ku, Chiba 2638555, Japan
[3] Kanazawa Univ Hosp, Dept Radiol, Kanazawa, Ishikawa 9208641, Japan
关键词
Target tracking; Image lag; Fluoroscopic image; Flat-panel detector (FPD); Radiotherapy; GUIDED RADIOTHERAPY; SYSTEM;
D O I
10.1269/jrr.10059
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Real-time tumor tracking in external radiotherapy can be achieved by diagnostic (kV) X-ray imaging with a dynamic flat-panel detector (FPD). The purpose of this study was to address image lag in target tracking and its influence on the accuracy of tumor tracking. Fluoroscopic images were obtained using a direct type of dynamic FPD. Image lag properties were measured without test devices according to IEC 62220-1. Modulation transfer function (MTF) and profile curves were measured on the edges of a moving tungsten plate at movement rate of 10 and 20 mm/s, covering lung tumor movement of normal breathing. A lung tumor and metal sphere with blurred edge due to image lag was simulated using the results and then superimposed on breathing chest radiographs of a patient. The moving target with and without image lag was traced using a template-matching technique. In the results, the image lag for the first frame after X-ray cutoff was 2.0% and decreased to less than 0.1% in the fifth frame. In the measurement of profile curves on the edges of static and moving tungsten material plates, the effect of image lag was seen as blurred edges of the plate. The blurred edges of a moving target were indicated as reduction of MTF. However, the target could be traced within an error of +/- 5 mm. The results indicated that there was no effect of image lag on target tracking in usual breathing speed in a radiotherapy situation.
引用
收藏
页码:723 / 731
页数:9
相关论文
共 26 条
  • [21] Effect of X-ray incident direction and scintillator layer design on image quality of indirect-conversion flat-panel detector with GOS phosphor
    Sato, K.
    Nariyuki, F.
    Nomura, H.
    Takasu, A.
    Fukui, S.
    Nakatsu, M.
    Okada, Y.
    Nabeta, T.
    Hosoi, Y.
    [J]. MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [22] Projected lung area using dynamic X-ray (DXR) with a flat-panel detector system and automated tracking in patients with chronic obstructive pulmonary disease (COPD)
    Hino, Takuya
    Tsunomori, Akinori
    Fukumoto, Takenori
    Hata, Akinori
    Hida, Tomoyuki
    Yamada, Yoshitake
    Ueyama, Masako
    Kamitani, Takeshi
    Nishino, Mizuki
    Kurosaki, Atsuko
    Jinzaki, Masahiro
    Ishigami, Kousei
    Honda, Hiroshi
    Yoneyama, Tsutomu
    Nagatsuka, Sumiya
    Kudoh, Shoji
    Hatabu, Hiroto
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2022, 157
  • [23] Investigation of the effect of anode/filter materials on the dose and image quality of a digital mammography system based on an amorphous selenium flat panel detector
    Baldelli, P.
    Phelan, N.
    Egan, G.
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2010, 83 (988): : 290 - 295
  • [24] Investigation of image lag in a high-speed flat panel detector-based cone beam CT imaging system - art. no. 61420Z
    Zhang, Yan
    Ning, Ruola
    Conover, David
    [J]. Medical Imaging 2006: Physics of Medical Imaging, Pts 1-3, 2006, 6142 : Z1420 - Z1420
  • [25] Digital subtraction angiography during transjugular intrahepatic portosystemic shunt creation or revision: data on radiation exposure and image quality obtained using a standard and a low-dose acquisition protocol in a flat-panel detector-based system
    Roberto Miraglia
    Luigi Maruzzelli
    Kelvin Cortis
    Corrado Tafaro
    Roberta Gerasia
    Carmelo Parisi
    Angelo Luca
    [J]. Abdominal Imaging, 2015, 40 : 1808 - 1812
  • [26] Digital subtraction angiography during transjugular intrahepatic portosystemic shunt creation or revision: data on radiation exposure and image quality obtained using a standard and a low-dose acquisition protocol in a flat-panel detector-based system
    Miraglia, Roberto
    Maruzzelli, Luigi
    Cortis, Kelvin
    Tafaro, Corrado
    Gerasia, Roberta
    Parisi, Carmelo
    Luca, Angelo
    [J]. ABDOMINAL IMAGING, 2015, 40 (06): : 1808 - 1812