Computational modeling of a single-element transcranial focused ultrasound transducer for subthalamic nucleus stimulation

被引:25
|
作者
Samoudi, Mohammed A. [1 ]
Van Renterghem, Timothy [1 ]
Botteldooren, Dick [1 ]
机构
[1] Univ Ghent, WAVES Res Grp, Dept Informat Technol, Ghent, Belgium
关键词
ultrasonic neuromodulation; deep brain stimulation; full-wave numerical simulations; subthalamic nucleus stimulation; transcranial focused ultrasound; BRAIN-STIMULATION; SKULL; NEUROSTIMULATION; SIMULATIONS; PROPAGATION; THALAMOTOMY; THERAPY;
D O I
10.1088/1741-2552/aafa38
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. While transcranial focused ultrasound is a very promising neuromodulation technique for its non-invasiveness and high spatial resolution, its application to the human deep brain regions such as the subthalamic nucleus (STN) is relatively new. The objective of this study is to design a simple ultrasound transducer and study the transcranial wave propagation through a highly realistic human head model. The effects of skull morphology and skull and brain tissue properties on the focusing performance and energy deposition must therefore be known. Approach. A full-wave finite-difference time-domain simulation platform was used to design and simulate ultrasound radiation from a single-element focused transducer (SEFT) to the STN. Simulations were performed using the state-of-the-art Multimodal Imaging-based and highly Detailed Anatomical (MIDA) head model. In addition, the impact of changes in sound speed, density, and tissue attenuation coefficients were assessed through a sensitivity analysis. Main results. A SEFT model was designed to deliver an intensity of around 100 W m(-2) to the STN region; 20% of the STN volume was sonicated with at least half of the maximum of the peak intensity and it was predicted that 61.5% of the volume of the beam (above half of the peak intensity) falls inside the STN region. The sensitivity analysis showed that the skull's sound speed is the most influential acoustic parameter, which must be known with less than 1.2% error to obtain an acceptable accuracy in intracranial fields and focusing (for less than 5% error). Significance. Ultrasound intensity delivery at the STN by a simple single element transducer is possible and could be a promising alternative to complex multi-element phased arrays, or more general, to invasive or less focused (non-acoustic) neuromodulation techniques. Accurate acoustic skull and brain parameters, including detailed skull geometry, are needed to ensure proper targeting in the deep brain region.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Method to optimize the placement of a single-element transducer for transcranial focused ultrasound
    Park, Tae Young
    Pahk, Ki Joo
    Kim, Hyungmin
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2019, 179
  • [2] Neuromodulation with single-element transcranial focused ultrasound in human thalamus
    Legon, Wynn
    Ai, Leo
    Bansal, Priya
    Mueller, Jerel K.
    HUMAN BRAIN MAPPING, 2018, 39 (05) : 1995 - 2006
  • [3] Differential evolution method to find optimal location of a single-element transducer for transcranial focused ultrasound therapy
    Park, Tae Young
    Kim, Hyo-Jin
    Park, So Hui
    Chang, Won Seok
    Kim, Hyungmin
    Yoon, Kyungho
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2022, 219
  • [4] Single-element focused ultrasound transducer method for harmonic motion imaging
    Maleke, Caroline
    Pernot, Mathieu
    Konofagou, Elisa E.
    ULTRASONIC IMAGING, 2006, 28 (03) : 144 - 158
  • [5] Image-guided navigation of single-element focused ultrasound transducer
    Kim, Hyungmin
    Chiu, Alan
    Park, Shinsuk
    Yoo, Seung-Schik
    INTERNATIONAL JOURNAL OF IMAGING SYSTEMS AND TECHNOLOGY, 2012, 22 (03) : 177 - 184
  • [6] Deep Neural Network for Navigation of a Single-Element Transducer During Transcranial Focused Ultrasound Therapy: Proof of Concept
    Choi, Minwook
    Jang, Minyeong
    Yoo, Seung-Schik
    Noh, Gunwoo
    Yoon, Kyungho
    IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2022, 26 (11) : 5653 - 5664
  • [7] Transducer modeling for accurate acoustic simulations of transcranial focused ultrasound stimulation
    Pasquinelli, Cristina
    Montanaro, Hazael
    Lee, Hyunjoo J.
    Hanson, Lars G.
    Kim, Hyungkook
    Kuster, Niels
    Siebner, Hartwig R.
    Neufeld, Esra
    Thielscher, Axel
    JOURNAL OF NEURAL ENGINEERING, 2020, 17 (04)
  • [8] Effects of Lamb Waves in a Single-Element High Intensity Focused Ultrasound Transducer
    Otsu, Kenji
    Kaneshima, Yasuhiro
    Yoshizawa, Shin
    Umemura, Shin-ichiro
    10TH INTERNATIONAL SYMPOSIUM ON THERAPEUTIC ULTRASOUND (ISTU 2010), 2011, 1359 : 211 - +
  • [9] Computational Modeling of Ultrasonic Subthalamic Nucleus Stimulation
    Tarnaud, Thomas
    Joseph, Wout
    Martens, Luc
    Tanghe, Emmeric
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2019, 66 (04) : 1155 - 1164
  • [10] Focused ultrasound heating in brain tissue/skull phantoms with 1 MHz single-element transducer
    Antoniou, Anastasia
    Evripidou, Nikolas
    Damianou, Christakis
    JOURNAL OF ULTRASOUND, 2024, 27 (02) : 263 - 274