Computational modeling of a single-element transcranial focused ultrasound transducer for subthalamic nucleus stimulation

被引:25
|
作者
Samoudi, Mohammed A. [1 ]
Van Renterghem, Timothy [1 ]
Botteldooren, Dick [1 ]
机构
[1] Univ Ghent, WAVES Res Grp, Dept Informat Technol, Ghent, Belgium
关键词
ultrasonic neuromodulation; deep brain stimulation; full-wave numerical simulations; subthalamic nucleus stimulation; transcranial focused ultrasound; BRAIN-STIMULATION; SKULL; NEUROSTIMULATION; SIMULATIONS; PROPAGATION; THALAMOTOMY; THERAPY;
D O I
10.1088/1741-2552/aafa38
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. While transcranial focused ultrasound is a very promising neuromodulation technique for its non-invasiveness and high spatial resolution, its application to the human deep brain regions such as the subthalamic nucleus (STN) is relatively new. The objective of this study is to design a simple ultrasound transducer and study the transcranial wave propagation through a highly realistic human head model. The effects of skull morphology and skull and brain tissue properties on the focusing performance and energy deposition must therefore be known. Approach. A full-wave finite-difference time-domain simulation platform was used to design and simulate ultrasound radiation from a single-element focused transducer (SEFT) to the STN. Simulations were performed using the state-of-the-art Multimodal Imaging-based and highly Detailed Anatomical (MIDA) head model. In addition, the impact of changes in sound speed, density, and tissue attenuation coefficients were assessed through a sensitivity analysis. Main results. A SEFT model was designed to deliver an intensity of around 100 W m(-2) to the STN region; 20% of the STN volume was sonicated with at least half of the maximum of the peak intensity and it was predicted that 61.5% of the volume of the beam (above half of the peak intensity) falls inside the STN region. The sensitivity analysis showed that the skull's sound speed is the most influential acoustic parameter, which must be known with less than 1.2% error to obtain an acceptable accuracy in intracranial fields and focusing (for less than 5% error). Significance. Ultrasound intensity delivery at the STN by a simple single element transducer is possible and could be a promising alternative to complex multi-element phased arrays, or more general, to invasive or less focused (non-acoustic) neuromodulation techniques. Accurate acoustic skull and brain parameters, including detailed skull geometry, are needed to ensure proper targeting in the deep brain region.
引用
收藏
页数:12
相关论文
共 50 条
  • [21] Development of a high frequency single-element ultrasound needle transducer for anesthesia delivery
    Ameri, Golafsoun
    Son, Jungik
    Liang, Jingwei
    Foster, F. Stuart
    Ganapathy, Sugantha
    Peters, Terry M.
    MEDICAL IMAGING 2017: ULTRASONIC IMAGING AND TOMOGRAPHY, 2017, 10139
  • [22] A head template for computational dose modelling for transcranial focused ultrasound stimulation
    Hosseini, Seyedsina
    Puonti, Oula
    Treeby, Bradley
    Hanson, Lars G.
    Thielscher, Axel
    NEUROIMAGE, 2023, 277
  • [23] Design and Fabrication of Single-Element CMUTs for Forming a Transcranial Array for Focused Beam Applications
    Ibn Minhaj, Tamzid
    Adelegan, Oluwafemi J.
    Biliroglu, Ali Onder
    Annayev, Muhammetgeldi
    Coutant, Zachary A.
    Yamaner, Feysel Yalcin
    Oralkan, Omer
    INTERNATIONAL ULTRASONICS SYMPOSIUM (IEEE IUS 2021), 2021,
  • [24] Calculation of dissipation resistances in a single-element transducer
    Loyau, Vincent
    Feuillard, Guy
    Hue, Louis-Pascal Tran Huu
    Lethiecq, Marc
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2007, 54 (11) : 2385 - 2393
  • [25] Numerical Study and Optimisation of a Novel Single-Element Dual-Frequency Ultrasound Transducer
    Sun, Changhe
    Jiang, Senlin
    Liu, Yufei
    SENSORS, 2018, 18 (03):
  • [26] Muscle Fatigue Assessment Using One-channel Single-element Ultrasound Transducer
    Sun, Xueli
    Li, Yuefeng
    Liu, Honghai
    2017 8TH INTERNATIONAL IEEE/EMBS CONFERENCE ON NEURAL ENGINEERING (NER), 2017, : 122 - 125
  • [27] Single-element ultrasonic transducer modeling using a hybrid FD-PSTD method
    Filoux, Erwan
    Levassort, Franck
    Calle, Samuel
    Certon, Dominique
    Lethiecq, Marc
    ULTRASONICS, 2009, 49 (08) : 611 - 614
  • [28] Photoacoustic Imaging of Nanoparticle-Containing Cells Using Single-Element Focused Transducer: A Simulation Study
    Karmakar, Subhajit
    Roy, Madhusudan
    Saha, Ratan K.
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2015, 62 (03) : 463 - 474
  • [29] Ultrasound Transducer and Amplifier Design in Transcranial Ultrasound Stimulation System
    Shi, Xun
    Zhang, Shuai
    Wang, Zhuo
    Xu, Guizhi
    2016 ASIA-PACIFIC INTERNATIONAL SYMPOSIUM ON ELECTROMAGNETIC COMPATIBILITY (APEMC), 2016, : 828 - 830
  • [30] Computational optimization of transcranial focused ultrasound stimulation: Toward noninvasive, selective stimulation of deep brain structures
    De Angelis, A.
    Leonetti, M.
    Apollonio, F.
    Liberti, M.
    Aglioti, S. M.
    Ruocco, G.
    APPLIED PHYSICS LETTERS, 2021, 118 (23)