Computed Tomography of the Head and Neck Region for Tumor Staging-Comparison of Dual-Source, Dual-Energy and Low-Kilovolt, Single-Energy Acquisitions

被引:23
|
作者
May, Matthias Stefan [1 ,2 ]
Bruegel, Joscha [1 ]
Brand, Michael [1 ]
Wiesmueller, Marco [1 ]
Krauss, Bernhard [3 ]
Allmendinger, Thomas [3 ]
Uder, Michael [1 ,2 ]
Wuest, Wolfgang [1 ,2 ]
机构
[1] Univ Erlangen Nurnberg, Dept Radiol, Erlangen, Germany
[2] Imaging Sci Inst, Erlangen, Germany
[3] Siemens Healthcare GmbH, Forchheim, Germany
关键词
CT; dual energy; monoenergetic; low kilovolt single energy; neck; TUBE VOLTAGE ADAPTATION; SQUAMOUS-CELL CARCINOMA; IMAGE-QUALITY; METAL-ARTIFACTS; ORAL-CANCER; CT; ANGIOGRAPHY; REDUCTION; COMBINATION; AGREEMENT;
D O I
10.1097/RLI.0000000000000377
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The aim of this study was to intra-individually compare the image quality obtained by dual-source, dual-energy (DSDE) computed tomography (CT) examinations and different virtual monoenergetic reconstructions to a low single-energy (SE) scan. Materials and Methods: Third-generation DSDE-CT was performed in 49 patients with histologically proven malignant disease of the head and neck region. Weighted average images (WAIs) and virtual monoenergetic images (VMIs) for low (40 and 60 keV) and high (120 and 190 keV) energies were reconstructed. A second scan aligned to the jaw, covering the oral cavity, was performed for every patient to reduce artifacts caused by dental hardware using a SE-CT protocol with 70-kV tube voltages and matching radiation dose settings. Objective image quality was evaluated by calculating contrast-to-noise ratios. Subjective image quality was evaluated by experienced radiologists. Results: Highest contrast-to-noise ratios for vessel and tumor attenuation were obtained in 40-keV VMI (all P < 0.05). Comparable objective results were found in 60-keV VMI, WAI, and the 70-kV SE examinations. Overall subjective image quality was also highest for 40-keV, but differences to 60-keV VMI, WAI, and 70-kV SE were nonsignificant (all P > 0.05). High kiloelectron volt VMIs reduce metal artifacts with only limited diagnostic impact because of insufficiency in case of severe dental hardware. CTDIvol did not differ significantly between both examination protocols (DSDE: 18.6 mGy; 70-kV SE: 19.4 mGy; P = 0.10). Conclusions: High overall image quality for tumor delineation in head and neck imaging were obtained with 40-keV VMI. However, 70-kV SE examinations are an alternative and modified projections aligned to the jaw are recommended in case of severe artifacts caused by dental hardware.
引用
收藏
页码:522 / 528
页数:7
相关论文
共 50 条
  • [21] Should Dual-Energy Computed Tomography Pulmonary Angiography Replace Single-Energy Computed Tomography Pulmonary Angiography in Pregnant and Postpartum Patients?
    McDermott, Shaunagh
    Otrakji, Alexi
    Flores, Efren J.
    Kalra, Mannudeep K.
    Shepard, Jo-Anne O.
    Digumarthy, Subba R.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2018, 42 (01) : 25 - 32
  • [22] Dual-Energy Computed Tomography Angiography of the Head and Neck With Single-Source Computed Tomography: A New Technical (Split Filter) Approach for Bone Removal
    Kaemmerer, Nadine
    Brand, Michael
    Hammon, Matthias
    May, Matthias
    Wuest, Wolfgang
    Krauss, Bernhard
    Uder, Michael
    Lell, Michael M.
    INVESTIGATIVE RADIOLOGY, 2016, 51 (10) : 618 - 623
  • [23] Urinary stone differentiation in patients with large body size using dual-energy dual-source computed tomography
    Qu, Mingliang
    Jaramillo-Alvarez, Giselle
    Ramirez-Giraldo, Juan C.
    Liu, Yu
    Duan, Xinhui
    Wang, Jia
    Vrtiska, Terri J.
    Krambeck, Amy E.
    Lieske, John
    McCollough, Cynthia H.
    EUROPEAN RADIOLOGY, 2013, 23 (05) : 1408 - 1414
  • [24] Deep learning based spectral extrapolation for dual-source, dual-energy x-ray computed tomography
    Clark, Darin P.
    Schwartz, Fides R.
    Marin, Daniele
    Ramirez-Giraldo, Juan C.
    Badea, Cristian T.
    MEDICAL PHYSICS, 2020, 47 (09) : 4150 - 4163
  • [25] Urinary stone differentiation in patients with large body size using dual-energy dual-source computed tomography
    Mingliang Qu
    Giselle Jaramillo-Alvarez
    Juan C. Ramirez-Giraldo
    Yu Liu
    Xinhui Duan
    Jia Wang
    Terri J. Vrtiska
    Amy E. Krambeck
    John Lieske
    Cynthia H. McCollough
    European Radiology, 2013, 23 : 1408 - 1414
  • [26] Low-dose dual-energy CT for stone characterization: a systematic comparison of two generations of split-filter single-source and dual-source dual-energy CT
    Nakhostin, Dominik
    Sartoretti, Thomas
    Eberhard, Matthias
    Krauss, Bernhard
    Mueller, Daniel
    Alkadhi, Hatem
    Euler, Andre
    ABDOMINAL RADIOLOGY, 2021, 46 (05) : 2079 - 2089
  • [27] Low-dose dual-energy CT for stone characterization: a systematic comparison of two generations of split-filter single-source and dual-source dual-energy CT
    Dominik Nakhostin
    Thomas Sartoretti
    Matthias Eberhard
    Bernhard Krauss
    Daniel Müller
    Hatem Alkadhi
    André Euler
    Abdominal Radiology, 2021, 46 : 2079 - 2089
  • [28] Myocardial late enhancement and extracellular volume with single-energy, dual-energy, and photon-counting computed tomography
    Oyama-Manabe, Noriko
    Oda, Seitaro
    Ohta, Yasutoshi
    Takagi, Hidenobu
    Kitagawa, Kakuya
    Jinzaki, Masahiro
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2024, 18 (01) : 3 - 10
  • [29] Applications of Dual-Energy Computed Tomography for the Evaluation of Head and Neck Squamous Cell Carcinoma
    Forghani, Reza
    Kelly, Hillary R.
    Curtin, Hugh D.
    NEUROIMAGING CLINICS OF NORTH AMERICA, 2017, 27 (03) : 445 - +
  • [30] Head and neck automatic multi-organ segmentation on Dual-Energy Computed Tomography
    Le, Anh Thu
    Sambourg, Killian
    Sun, Roger
    Deny, Nicolas
    Cifliku, Vjona
    Rouhi, Rahimeh
    Deutsch, Eric
    Fournier-Bidoz, Nathalie
    Robert, Charlotte
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2024, 32