Comparison of Monaco treatment planning system algorithms and Monte Carlo simulation for small fields in anthropomorphic RANDO phantom: The esophagus case

被引:3
|
作者
Tugrul, Taylan [1 ]
机构
[1] Yuzuncu Yil Univ, Dept Radiat Oncol, Fac Med, Van, Turkey
关键词
Anthropomorphic RANDO phantom; BEAMnrc; DOSXYZnrc; Monaco treatment planning system; Monte Carlo simulation; INHOMOGENEITY CORRECTION ALGORITHMS; COLLAPSED CONE CONVOLUTION; PHOTON DOSE CALCULATION; PENCIL-BEAM; DOSIMETRIC VERIFICATION; QUANTITATIVE ASSESSMENT; OPTIMIZATION; PARAMETERS; ACCURACY; DENSITY;
D O I
10.4103/jcrt.JCRT_1143_20
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: In this study, the dose distributions obtained by the algorithms used in Monaco treatment planning system (TPS) and Monte Carlo (MC) simulation were compared for small fields in the anthropomorphic RANDO phantom, and then, the results were analyzed using the gamma analysis method. Materials and Methods: In the study, dose distributions obtained from the collapse cone algorithm, MC algorithm, and MC simulation were examined. The EGSnrc was utilized for MC simulation. Results: In radiation fields smaller than 3 cm x 3 cm, the doses calculated by the CC algorithm are particularly high in the region of lung/soft-tissue interfaces. In the region of soft-tissue/vertebral interfaces, the doses calculated by the CC algorithm and the MC algorithm are compatible with the MC simulation. For each algorithm, the main reason for the non-overlapping dose curves in small fields compared to MC simulation is that the lateral electronic equilibrium loss is not taken into account by the algorithms. Conclusion: The doses calculated by the algorithms used in TPS may differ, especially in environments where density changes are sharp. Even if the radiation dose from different angles is calculated similarly in the target area by the algorithms, the calculated doses in the tissues in each radiation field path may be different. Therefore, to increase the quality of radiotherapy and to protect critical organs more accurately, the accuracy of the algorithms in TPS should be checked before treatment, especially in multi-field treatments such as stereotactic body radiation therapy and intensity-modulated radiotherapy for tumors in the abdominal region.
引用
收藏
页码:1370 / 1375
页数:6
相关论文
共 50 条
  • [1] The effect of algorithms on dose distribution in inhomogeneous phantom: Monaco treatment planning system versus monte carlo simulation
    Tugrul, Taylan
    [J]. JOURNAL OF MEDICAL PHYSICS, 2021, 46 (02) : 111 - 115
  • [2] Comparison of treatment planning dose calculations with measurements and Monte Carlo calculations in a RANDO phantom
    Wyatt, M
    Corredor, C
    Tamimi, M
    Miller, LF
    [J]. RADIATION PROTECTION DOSIMETRY, 2005, 116 (1-4) : 461 - 465
  • [3] Commissioning of Monaco Monte Carlo IMRT Treatment Planning System
    Li, J.
    Doemer, A.
    Cao, J.
    Podder, T.
    Harrison, A.
    Yu, Y.
    Galvin, J.
    Xiao, Y.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [4] Monte Carlo evaluation of a treatment planning system for helical tomotherapy in an anthropomorphic heterogeneous phantom and for clinical treatment plans
    Zhao, Ying-Li
    Mackenzie, M.
    Kirkby, C.
    Fallone, B. G.
    [J]. MEDICAL PHYSICS, 2008, 35 (12) : 5366 - 5374
  • [5] DOSIMETRIC COMMISSIONING OF A MONTE CARLO MODEL IN THE ELEKTA MONACO TREATMENT PLANNING SYSTEM
    Hodgson, D.
    Convery, D.
    Greener, A.
    Misson-Yates, S.
    Thomas, C.
    Footman, M.
    Applegate, K.
    Harrison, L.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S522 - S522
  • [6] Dose calculation using Monte Carlo algorithm of Monaco treatment planning system
    Li, J.
    Doemer, A.
    Harrison, A. S.
    Galvin, J. M.
    Xiao, Y.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (01): : S674 - S674
  • [7] Treatment planning for a small animal using Monte Carlo simulation
    Chow, James C. L.
    Leung, Michael K. K.
    [J]. MEDICAL PHYSICS, 2007, 34 (12) : 4810 - 4817
  • [8] Comparison of Treatment Planning Algorithms and Monte Carlo Simulations in Oesophageal Radiotherapy
    Johns, D.
    Spezi, E.
    Downes, P.
    Lewis, D.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S776 - S776
  • [9] Patient Dose Distribution Comparison between Monte Carlo Simulation and Treatment Planning System
    Jia, Y.
    Beltran, C.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [10] A Comparison between Pencil Beam and Monte Carlo Algorithms Against Film Measurements in an Anthropomorphic Phantom for Proton Spot Scanning
    Saini, J.
    Maes, D.
    Bowen, S. R.
    Egan, A. J.
    Bloch, C.
    Wong, T. P.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 99 (02): : E717 - E718