Computer-assisted diagnostic procedure for multislice computed tomography for assessment of the pulmonary arterial circulation

被引:0
|
作者
Malich, A. [1 ]
Hentrich, D.
Hansch, A. [2 ]
Boettcher, J.
Teichgraeber, U. [2 ]
机构
[1] Sudharz Krankenhaus Nordhausen, Inst Radiol, D-99734 Nordhausen, Germany
[2] Univ Jena, Zentrum Radiol, D-6900 Jena, Germany
来源
RADIOLOGE | 2012年 / 52卷 / 04期
关键词
Computer-assisted detection; Pulmonary embolism; Multislice computed tomography; Interobserver reliability; 3D analysis; AIDED DETECTION; CT ANGIOGRAPHY; RISK-FACTORS; EMBOLISM; HYPERTENSION; IMPACT;
D O I
10.1007/s00117-012-2304-5
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A recently developed CAD software which highlights intravascular thrombotic structures from multislice computed tomography (MSCT) data was tested regarding feasibility, interobserver reliability and effect on radiology reports. The CAD system ImageCheckerA (R) CT-Lung was tested in a randomized double-blinded study on 160 MSCT datasets (standardized technical conditions) performed for suspected pulmonary embolism (PE). The CAD data and images were analyzed by three radiologists in an independent and blinded fashion. The data from all 160 cases could be analyzed and 604 CAD prompts were set. Using the CAD analysis significantly more PEs were found in the peripheral pulmonary arterial circulation than described in the initial report. In 38 cases the 3 radiologists in consensus scored the images with the CAD adjunct as PE positive in peripheral vessels, which were initially reported as negative. Despite differences in the evaluation between two radiologists the amended assessment of the imaging data using the CAD softwear was reliable. There was a significant correlation between D-dimer values and the number of embolic structures detected by the CAD analysis. The recently developed CAD system is a useful adjunct as second reader to detect subtle emboli in peripheral vessels of MSCT datasets.
引用
收藏
页码:366 / 372
页数:7
相关论文
共 50 条
  • [21] COMPUTER-ASSISTED TOMOGRAPHY IN ADRENOLEUKODYSTROPHY
    EIBEN, RM
    DICHIRO, G
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1977, 1 (03) : 308 - 314
  • [22] PULMONARY CIRCULATION IN RELATION WITH VENTILATION USING COMPUTER-ASSISTED SCINTILLATION CAMERA
    ISHII, Y
    ITO, H
    NOMURA, S
    TORIZUKA, K
    HARA, A
    [J]. JAPANESE CIRCULATION JOURNAL-ENGLISH EDITION, 1973, 37 (06): : 597 - 597
  • [23] COMPUTER-ASSISTED TOMOGRAPHY OF BRAIN
    GRIFFITH, H
    THOMSON, JLG
    [J]. LANCET, 1974, 2 (7890): : 1205 - 1206
  • [24] COMPUTER-ASSISTED TOMOGRAPHY IN CHILDREN
    MORRISON, IS
    GLANCY, JJ
    CALA, LA
    [J]. AUSTRALIAN PAEDIATRIC JOURNAL, 1977, 13 (01): : 39 - 43
  • [25] COMPUTER-ASSISTED TOMOGRAPHY OF THE CHEST
    DEBAETS, M
    SCHNYDER, P
    [J]. SCHWEIZERISCHE MEDIZINISCHE WOCHENSCHRIFT, 1979, 109 (29) : 1089 - 1093
  • [26] COMPUTER-ASSISTED TOMOGRAPHY OF THE KIDNEY
    ONESON, SR
    HAAGA, JR
    [J]. WORLD JOURNAL OF UROLOGY, 1992, 10 (03) : 142 - 148
  • [27] COMPUTER-ASSISTED TOMOGRAPHY OF KNEE
    PAVLOV, H
    FREIBERGER, RH
    DECK, MF
    MARSHALL, JL
    MORRISSEY, JK
    [J]. INVESTIGATIVE RADIOLOGY, 1978, 13 (01) : 57 - 62
  • [28] CRANIOCERVICAL COMPUTER-ASSISTED TOMOGRAPHY
    SADAMOTO, K
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1977, 1 (02) : 253 - 253
  • [29] COMPUTER-ASSISTED TOMOGRAPHY OF BRAIN
    BLIGH, AS
    FRAZER, AK
    GRAHAM, JG
    THOMPSON, RJ
    WEEKS, RD
    WELLS, CEC
    [J]. LANCET, 1974, 2 (7891): : 1260 - 1260
  • [30] COMPUTER-ASSISTED TOMOGRAPHY IN SYRINGOMYELIA
    RINALDI, I
    KOPP, JE
    HARRIS, WO
    REAGAN, TJ
    MURPHY, D
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1978, 2 (05) : 633 - 635